Skip to content

TiBrid Stand Alone Intervertebral Body Fusion Device, SA… recallZ-2556-2025 · Omnia Medical · initiated Jul 15, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Omnia Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Failure of fusion system instruments in the field.

FDA root cause category: Device Design

What the company is doing

Consignees were called on or around 8/1/2024 to notify them of this recall of instruments. Affected instruments were to be sent back to Omnia Medical and replaced with instruments that underwent rework in the form of design changes aimed at increasing their durability.

Recall details

Recall number
Z-2556-2025
Status
Open, Classified
Initiated by the company
Jul 15, 2024
Posted by the FDA
Sep 11, 2025
Recalling firm
Omnia Medical
Quantity
15 units
Product code
OVD
FDA submissions
K203207
Distribution
US Nationwide distribution in the states of AZ, CA, NJ, NV, OK & WA.

Product description

TiBrid Stand Alone Intervertebral Body Fusion Device, SA Angled Punch Awl, Sizes: 30 degree, 45 degree and 60 degree.

Code information (models, lots, serial numbers)

Model No. 17-5A-AWL3, 17-5A-AWL4, 17-5A-AWL5; UDI: 00843511113056, 00843511122287, 00843511122294.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K203207 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97402).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page