NOxBOXi Nitric Oxide Delivery System recallZ-2544-2025 · Noxbox Ltd · initiated Jul 19, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Noxbox Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Replacement of the internal Sample Pump due to high failure rates.

FDA root cause category: Component design/selection

What the company is doing

An email with the subject "New Internal Sampling Pump" was sent to consignees on 7/19/2023 notifying them that the internal Sampling Pump of NOxBOXi Nitric Oxide Delivery System devices was to be replaced. The replacement does not change the use of the device but does result in a quieter device functionality. This replacement will be conducted during annual servicing by Level 3 Service Engineers at authorized NOXBOX Service Centers. Consignees with any questions are to email [email protected] or their account manager.

Recall details

Recall number
Z-2544-2025
Status
Open, Classified
Initiated by the company
Jul 19, 2023
Posted by the FDA
Sep 9, 2025
Recalling firm
Noxbox Ltd
Quantity
1,667 units
Product code
MRN
FDA submissions
K171696
Distribution
US Nationwide distribution in the state of TN.

Product description

NOxBOXi Nitric Oxide Delivery System

Code information (models, lots, serial numbers)

Show the full code information (17k characters)

Model No. NOXBOXI; UDI: 05060541640009; Serial No. NI102738, NI103021, NI103146, NI101560, NI101059, NI102922, NI101034, NI102152, NI102854, NI101289, NI102957, NI102642, NI101960, NI102163, NI101717, NI102928, NI101146, NI101209, NI102128, NI101751, NI101743, NI102266, NI101813, NI102037, NI102079, NI102850, NI100850, NI101221, NI101708, NI101744, NI101879, NI101905, NI101948, NI102410, NI101208, NI101354, NI101940, NI102117, NI102338, NI101002, NI101729, NI102094, NI102310, NI102676, NI102888, NI102683, NI102933, NI102537, NI101594, NI101958, NI101989, NI102104, NI102206, NI102620, NI102639, NI102696, NI102771, NI102814, NI102852, NI101147, NI102431, NI100983, NI101257, NI101353, NI101581, NI101987, NI102111, NI102308, NI102816, NI102908, NI102698, NI102725, NI102883, NI102932, NI102936, NI103084, NI101703, NI101993, NI102422, NI102485, NI102879, NI102939, NI101234, NI101821, NI102947, NI101029, NI101679, NI102031, NI102538, NI102624, NI102794, NI102839, NI102848, NI102865, NI102942, NI102999, NI101077, NI101600, NI102121, NI102424, NI102545, NI102812, NI102838, NI102849, NI102861, NI102877, NI102885, NI102944, NI101136, NI101340, NI101416, NI101916, NI101937, NI102060, NI102815, NI102437, NI102513, NI102541, NI102762, NI102828, NI102836, NI101380, NI101712, NI102062, NI102123, NI102278, NI102343, NI102492, NI102853, NI102963, NI102540, NI102727, NI102350, NI102382, NI102507, NI102604, NI102625, NI102658, NI102871, NI101095, NI102584, NI102726, NI102823, NI102832, NI102855, NI102998, NI103022, NI101081, NI101151, NI101211, NI101212, NI101726, NI102614, NI103012, NI101278, NI101638, NI101710, NI101775, NI102052, NI102168, NI102271, NI102274, NI102476, NI102659, NI102724, NI102825, NI101675, NI101769, NI101835, NI101901, NI101943, NI102076, NI102290, NI102475, NI102482, NI102562, NI102617, NI101869, NI102300, NI102670, NI102700, NI102913, NI101086, NI101288, NI101680, NI101716, NI101756, NI102053, NI102057, NI102068, NI102887, NI102929, NI102356, NI103011, NI103016, NI101148, NI101812, NI101952, NI102078, NI102110, NI100866, NI101149, NI101430, NI101446, NI101540, NI102219, NI102397, NI102488, NI102744, NI102864, NI103017, NI100895, NI101067, NI101207, NI101639, NI101684, NI101999, NI102240, NI103123, NI103126, NI103136, NI103151, NI103204, NI103208, NI101793, NI102588, NI103104, NI103127, NI103145, NI103203, NI100882, NI101163, NI101964, NI102166, NI103147, NI103148, NI103168, NI103181, NI103183, NI101084, NI101303, NI101782, NI101819, NI101857, NI101912, NI101970, NI102404, NI102561, NI102748, NI102759, NI102931, NI101100, NI101252, NI101467, NI101629, NI101654, NI101700, NI101732, NI101748, NI101794, NI101928, NI102072, NI102423, NI102463, NI102809, NI102973, NI103182, NI103220, NI100880, NI101156, NI101321, NI101896, NI101974, NI101995, NI102043, NI102049, NI102254, NI102286, NI102394, NI101044, NI101349, NI101357, NI101776, NI101814, NI101953, NI102021, NI102063, NI102114, NI102120, NI102435, NI102536, NI102722, NI102866, NI102164, NI102917, NI101028, NI101373, NI101449, NI101691, NI102023, NI102113, NI102141, NI102184, NI102345, NI102630, NI102645, NI102772, NI100953, NI101019, NI102445, NI102605, NI102962, NI101246, NI101356, NI101615, NI101778, NI102017, NI102148, NI102259, NI102616, NI100840, NI100869, NI102131, NI102402, NI102406, NI102415, NI103120, NI103139, NI100855, NI101026, NI101138, NI101223, NI101464, NI101554, NI101761, NI101796, NI101824, NI101906, NI102190, NI102967, NI101116, NI101205, NI101231, NI101379, NI101890, NI101939, NI102048, NI102484, NI100878, NI101013, NI101115, NI101444, NI101578, NI101608, NI101925, NI102241, NI102340, NI102564, NI102881, NI102911, NI100861, NI100976, NI101150, NI101161, NI101176, NI101402, NI101443, NI101478, NI101959, NI102284, NI102512, NI102598, NI102706, NI102834, NI101659, NI101765, NI102165, NI102180, NI102295, NI102506, NI102542, NI102576, NI102688, NI101006, NI101073, NI101153, NI101251, NI101390, NI101405, NI101815, NI101864, NI101867, NI102396, NI102766, NI102986, NI101031, NI101233, NI101244, NI101809, NI101891, NI101982, NI102008, NI102040, NI102081, NI102205, NI102288, NI102619, NI102656, NI101512, NI102096, NI102357, NI102486, NI102699, NI102784, NI103144, NI101107, NI101199, NI101265, NI101299, NI101685, NI101837, NI101996, NI102647, NI102651, NI102691, NI102818, NI102958, NI101219, NI101232, NI101245, NI101313, NI101344, NI101736, NI101979, NI102961, NI101227, NI101298, NI101843, NI101873, NI101926, NI102292, NI102494, NI101110, NI101327, NI101469, NI101572, NI101945, NI102196, NI102713, NI101065, NI101177, NI101317, NI101709, NI101841, NI101846, NI102230, NI102285, NI100887, NI101117, NI101135, NI101180, NI101237, NI101328, NI101651, NI102055, NI102233, NI102381, NI103155, NI101094, NI101612, NI101722, NI101746, NI102091, NI102097, NI102150, NI102395, NI102898, NI100876, NI101061, NI101165, NI101666, NI101682, NI101771, NI102145, NI102554, NI103116, NI103159, NI100999, NI101322, NI101347, NI101479, NI101724, NI102296, NI102842, NI103196, NI100872, NI101217, NI101350, NI102107, NI102138, NI102490, NI102502, NI102660, NI102681, NI102714, NI102843, NI103157, NI101057, NI101145, NI101216, NI101307, NI101384, NI101686, NI101787, NI102019, NI102071, NI102119, NI102172, NI102282, NI102411, NI102483, NI101160, NI101409, NI101460, NI102414, NI102553, NI102669, NI102785, NI101956, NI102362, NI102937, NI102383, NI102481, NI102594, NI103152, NI101101, NI101530, NI101941, NI102332, NI102610, NI101079, NI101431, NI101620, NI101983, NI102090, NI102118, NI102544, NI102557, NI102579, NI102757, NI100932, NI101083, NI101106, NI101159, NI101338, NI101388, NI101396, NI101486, NI101606, NI101665, NI102102, NI102192, NI102200, NI102802, NI102919, NI103083, NI101336, NI101439, NI101492, NI101565, NI102026, NI102247, NI102900, NI100881, NI101414, NI101942, NI102144, NI102452, NI101218, NI101250, NI102012, NI102427, NI102599, NI101155, NI101475, NI101954, NI101985, NI101175, NI101570, NI100975, NI101617, NI101660, NI101994, NI102279, NI101859, NI100993, NI101343, NI101411, NI101461, NI101697, NI102280, NI101309, NI101696, NI101862, NI102149, NI102235, NI102627, NI101090, NI101763, NI101961, NI102304, NI102363, NI102923, NI101263, NI101601, NI102046, NI102213, NI102408, NI101038, NI101521, NI101668, NI102330, NI102389, NI102731, NI101024, NI101694, NI101641, NI102186, NI101092, NI101109, NI101817, NI101910, NI101049, NI102065, NI102093, NI102417, NI101335, NI101392, NI101395, NI101842, NI102384, NI102528, NI100989, NI101133, NI101968, NI102640, NI103023, NI100980, NI101089, NI101505, NI102167, NI102253, NI100868, NI101017, NI101035, NI101142, NI102161, NI100839, NI101202, NI101566, NI101573, NI101884, NI102075, NI102245, NI102256, NI102432, NI101010, NI101282, NI101739, NI101838, NI102074, NI102525, NI101408, NI101489, NI101851, NI101903, NI102044, NI102198, NI100848, NI101113, NI101224, NI101285, NI102133, NI102560, NI100852, NI101014, NI101214, NI101462, NI102735, NI101766, NI101816, NI102275, NI102667, NI101266, NI101508, NI102027, NI102199, NI102216, NI101635, NI101885, NI102125, NI102238, NI103082, NI101105, NI101616, NI102126, NI102146, NI102739, NI100961, NI101907, NI102137, NI102607, NI102897, NI101553, NI101625, NI101662, NI101823, NI101990, NI101607, NI102193, NI102232, NI102298, NI102535, NI101329, NI101655, NI101797, NI101807, NI101946, NI101308, NI102268, NI102270, NI102276, NI102429, NI101337, NI101472, NI101557, NI102613, NI102717, NI101170, NI101377, NI102265, NI102306, NI102358, NI103001, NI101213, NI101429, NI101577, NI101876, NI102418, NI102608, NI101270, NI101844, NI101886, NI102001, NI102140, NI102305, NI102489, NI101645, NI101998, NI102059, NI102339, NI102359, NI102571, NI101134, NI101168, NI101287, NI101622, NI101657, NI102596, NI102920, NI100847, NI101064, NI101108, NI102520, NI102872, NI101254, NI102106, NI102169, NI102666, NI102702, NI101072, NI101342, NI102228, NI102403, NI102574, NI102175, NI100858, NI101636, NI102182, NI102227, NI102447, NI102472, NI000000, NI101015, NI102124, NI102215, NI102550, NI103014, NI101515, NI101977, NI102578, NI102745, NI103170, NI101039, NI101179, NI101992, NI102413, NI101004, NI101011, NI102992, NI101613, NI102025, NI102127, NI102134, NI101610, NI101914, NI101967, NI101976, NI102558, NI102032, NI102038, NI102160, NI102405, NI102692, NI101368, NI101991, NI102056, NI102195, NI101451, NI101783, NI102011, NI102631, NI102795, NI101576, NI102255, NI102970, NI102996, NI100942, NI101387, NI101404, NI102281, NI102747, NI100873, NI101264, NI101596, NI102552, NI101060, NI101585, NI102509, NI102597, NI103024, NI100935, NI101922, NI102018, NI102531, NI102547, NI101016, NI101588, NI101704, NI101711, NI101114, NI101268, NI101324, NI101592, NI101484, NI102311, NI102477, NI102655, NI102893, NI100841, NI101808, NI101840, NI102674, NI101271, NI101323, NI102020, NI100883, NI101220, NI102568, NI102689, NI101509, NI101836, NI102209, NI102497, NI102615, NI100885, NI101294, NI102392, NI102611, NI102649, NI101190, NI102287, NI101192, NI102355, NI102426, NI102566, NI102704, NI102728, NI101720, NI102177, NI102234, NI102511, NI102673, NI102858, NI101406, NI101677, NI101820, NI102737, NI101301, NI102510, NI102695, NI100996, NI101423, NI101584, NI102743, NI100851, NI101902, NI102061, NI102250, NI102643, NI101005, NI102183, NI102188, NI102399, NI101602, NI101849, NI102707, NI102746, NI101676, NI101692, NI102136, NI102587, NI102623, NI100859, NI101658, NI101798, NI101963, NI102729, NI101695, NI101957, NI102277, NI102438, NI101474, NI102585, NI102720, NI102742, NI102756, NI101130, NI101425, NI101522, NI101669, NI102546, NI100833, NI101524, NI101973, NI102092, NI101003, NI101767, NI101825, NI102730, NI101255, NI101437, NI101520, NI101611, NI102331, NI102593, NI102755, NI101412, NI101605, NI101747, NI101848, NI101981, NI102135, NI102249, NI102595, NI102708, NI102711, NI102752, NI102845, NI101491, NI102466, NI102703, NI101456, NI101731, NI101791, NI102299, NI102581, NI101810, NI102591, NI102808, NI101628, NI102083, NI102109, NI102790, NI101715, NI101822, NI102806, NI102977, NI101042, NI101740, NI102391, NI102496, NI102792, NI101375, NI102629, NI102904, NI101144, NI101178, NI101359, NI101532, NI102010, NI102291, NI102675, NI102697, NI101516, NI101690, NI101997, NI102273, NI102377, NI102480, NI100990, NI101719, NI101184, NI101693, NI101781, NI102307, NI102657, NI102662, NI102841, NI101243, NI101428, NI101448, NI102283, NI102425, NI102543, NI103153, NI101640, NI101305, NI101386, NI102297, NI102914, NI101230, NI101473, NI101627, NI102070, NI102191, NI102327, NI102341, NI102380, NI102997, NI101033, NI101483, NI101173, NI101924, NI102602, NI102478, NI101292, NI101598, NI101832, NI100875, NI100987, NI101434, NI101799, NI102387, NI102890, NI101312, NI101468, NI101805, NI101881, NI102570, NI102671, NI100849, NI100888, NI101131, NI101381, NI101920, NI102237, NI102398, NI102668, NI102768, NI101030, NI101099, NI101496, NI101898, NI102246, NI102349, NI102634, NI102801, NI103074, NI103087, NI100994, NI101593, NI101604, NI101644, NI101718, NI102559, NI102565, NI102582, NI102600, NI102618, NI102685, NI102732, NI102905, NI100986, NI101643, NI102170, NI102504, NI102680, NI102817, NI102846, NI103002, NI101351, NI101360, NI102194, NI102781, NI101868, NI101878, NI102069, NI102777, NI101001, NI101071, NI101253, NI101295, NI101314, NI101595, NI102174, NI102433, NI102473, NI102661, NI102799, NI102934, NI101415, NI101426, NI101481, NI102009, NI102047, NI102272, NI102364, NI102444, NI102499, NI102556, NI102590, NI102709, NI102736, NI102830, NI102862, NI103215, NI101330, NI101865, NI101892, NI102115, NI102856, NI102918, NI101236, NI101826, NI102498, NI101158, NI101361, NI101574, NI101919, NI102028, NI102586, NI102622, NI101225, NI101672, NI101750, NI101874, NI102927, NI102244, NI101563, NI101678, NI101880, NI102366, NI102648, NI102892, NI101355, NI101457, NI101614, NI101689, NI101777, NI101671, NI101965, NI102269, NI103189, NI101376, NI101514, NI101759, NI101955, NI101687, NI102024, NI102155, NI102741, NI102878, NI101591, NI102222, NI102884, NI102987, NI101413, NI101895, NI102103, NI102804, NI101784, NI102173, NI102361, NI102569, NI102770, NI102910, NI101441, NI101459, NI102116, NI102189, NI102385, NI102479, NI101656, NI101889, NI101913, NI102518, NI103122, NI101527, NI101988, NI102176, NI102212, NI102347, NI102798, NI102979, NI102035, NI102733, NI102876, NI101526, NI101706, NI101883, NI102142, NI102214, NI102293, NI102555, NI102567, NI102635, NI102638, NI102760, NI102968, NI101590, NI102147, NI102329, NI102551, NI102767, NI102789, NI102891, NI101487, NI101525, NI101667, NI101699, NI101723, NI101737, NI102517, NI102819, NI102882, NI101432, NI101463, NI101533, NI101623, NI102036, NI102519, NI102950, NI101742, NI102171, NI102436, NI102563, NI101713, NI101749, NI102045, NI103094, NI102474, NI102575, NI102763, NI101442, NI101550, NI101861, NI102467, NI102522, NI102539, NI102837, NI102860, NI101433, NI101597, NI101768, NI102221, NI102693, NI101972, NI102208, NI102487, NI102637, NI102653, NI102705, NI103018, NI103020, NI103079, NI102015, NI102393, NI102400, NI102601, NI102764, NI102847, NI102899, NI102907, NI101986, NI102098, NI102734, NI102803, NI102955, NI102684, NI102710, NI102915, NI103019, NI102077, NI102095, NI102530, NI102580, NI102906, NI103003, NI101167, NI101872, NI102990, NI102434, NI102686, NI102894, NI101454, NI101683, NI101811, NI102000, NI102197, NI102889, NI101579, NI102753, NI102793, NI102880, NI101714, NI101394, NI101984, NI102013, NI102428, NI102776, NI101410, NI101609, NI101753, NI101845, NI102628, NI102758, NI101599, NI101664, NI101674, NI101975, NI102239, NI102351, NI102501, NI102503, NI102577, NI102664, NI102740, NI102870, NI102935, NI102309, NI102840, NI102926, NI101626, NI101688, NI101760, NI101854, NI101887, NI102099, NI102495, NI102652, NI102783, NI102805, NI102810, NI102003, NI102139, NI102226, NI102715, NI102791, NI102821, NI102966, NI103089, NI101670, NI101882, NI101904, NI102930, NI103085, NI102991, NI101575, NI101829, NI101393, NI102342, NI102346, NI102663, NI102873, NI101407, NI101707, NI101735, NI101770, NI101962, NI102051, NI102807, NI102868, NI102901, NI101506, NI101529, NI102811, NI102863, NI102916, NI103091, NI101435, NI101899, NI102004, NI102067, NI102592, NI102687, NI102751, NI103093, NI103095, NI102989, NI102824, NI102641, NI102548, NI102774, NI102800, NI101944, NI102672, NI102859, NI101551, NI101733, NI102089, NI102390, NI102621, NI102632, NI102701, NI102833, NI102874, NI102921, NI101624, NI102773, NI101971, NI102902, NI103117, NI100862, NI100978, NI101228, NI101241, NI101256, NI101258, NI101284, NI101297, NI101300, NI101311, NI101721, NI101831, NI101229, NI101828, NI103007, NI101041, NI101204, NI101273, NI101286, NI101490, NI100884, NI101215, NI101222, NI101795, NI101012, NI101018, NI101080, NI101088, NI101091, NI101093, NI100948, NI100991, NI100997, NI101069, NI101125, NI102813, NI102851, NI100879, NI101154, NI101181, NI101182, NI101183, NI101063, NI101482, NI101112, NI101804, NI102006, NI102500, NI101132, NI101164, NI101174, NI101009, NI101372, NI102108, NI102844, NI102007, NI101569, NI101978, NI102105, NI101806, NI102086, NI102263, NI101485, NI101818, NI101333, NI101681, NI102360, NI102769, NI101908, NI101346, NI101909, NI102082, NI103088, NI102002, NI101552, NI102778, NI102765, NI102964, NI102014, NI102129, NI100865, NI101008, NI101966, NI101048, NI101259, NI102606, NI101936, NI102886, NI103264, NI103298, NI103322, NI103287, NI103317, NI100949, NI101947, NI100886, NI101137, NI101162, NI103302, NI103305, NI103307, NI103311, NI103321, NI101728, NI101436, NI101452, NI100951, NI100982, NI100988, NI100992, NI101000, NI101025, NI101056, NI101074, NI101075, NI101076, NI101085, NI101087, NI101098, NI101119, NI101120, NI101129, NI101139, NI101143, NI101152, NI101166, NI101201, NI101203, NI101206, NI101210, NI101226, NI101248, NI101272, NI101283, NI101290, NI101293, NI101302, NI101348, NI101403, NI101417, NI101427, NI101447, NI101465, NI101501, NI101502, NI101507, NI101555, NI101562, NI101580, NI101589, NI101603, NI101618, NI101619, NI101630, NI101631, NI101632, NI101661, NI101663, NI101698, NI101725, NI101734, NI101785, NI101847, NI101863, NI101870, NI101877, NI101921, NI101927, NI101929, NI101950, NI101969, NI102022, NI102030, NI102039, NI102054, NI102058, NI102064, NI102066, NI102073, NI102080, NI102084, NI102112, NI102143, NI102151, NI102187, NI102229, NI102243, NI102294, NI102312, NI102348, NI102388, NI102401, NI102491, NI102493, NI102514, NI102515, NI102573, NI102589, NI102626, NI102633, NI102636, NI102646, NI102650, NI102654, NI102679, NI102682, NI102690, NI102712, NI102750, NI102775, NI102779, NI102780, NI102796, NI102829, NI102835, NI102867, NI102896, NI102909, NI102912, NI102925, NI103064, NI103076, NI103092, NI103132.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page