10Fore recallZ-2553-2025 · Merit Medical Systems, Inc. · initiated Jul 25, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a manufacturing issue involving the silicone quad ring where the ring is deformed or otherwise damaged and may cause a foreign body in the fluid pathway.
FDA root cause category: Employee error
What the company is doing
On 07/23/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing them that due to a manufacturing defect. Specifically, the o-ring between the y-body and the rotator may be damaged during assembly and a portion of the o-ring may remain loose or detach. Use of the affected product may result in a foreign body in the fluid path. Customer are instructed to: Immediately stop using or distributing the affected lots and return to Merit. 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan and email the completed Customer Response Form to Customer Service at [email protected] within 10 business days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. Any questions concerning this communication, contact your Merit Sales Representative or Merit Customer Service via email at [email protected] or via phone at +1 800 356 3748 | Hours: 6 am to 6 pm MST | Mon-Fri.
Recall details
- Recall number
- Z-2553-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 25, 2025
- Posted by the FDA
- Sep 11, 2025
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 447
- Product code
- DTL
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, FL, LA, MT, NY, SD, and WI. The countries of Canada and Korea
Product description
10Fore Hemostasis Valve Catalog Number / UDI-DI code: MAPTEN4 / 00884450821305 MAPTEN42 / 00884450832509 MAPTEN452 / 00884450821329 The 10Fore Hemostasis Valve is intended to maintain hemostasis during the introduction/withdrawal and use of diagnostic and interventional devices.
Code information (models, lots, serial numbers)
Catalog Number / UDI-DI code: MAPTEN4 / 00884450821305 Lot Numbers: H3098831 H3111155 H3152503 H3155611 H3186901 MAPTEN42 / 00884450832509 Lot Numbers: H3076189 H3111156 H3152504 H3155614 H3173674 MAPTEN452 / 00884450821329 Lot Number: H3111162
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- 10Fore™ Initial ReleaseMerit Medical Systems, Inc. · Angiography/Angioplasty Kit1 recall1 report
- 10Fore™ Initial ReleaseMerit Medical Systems, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page