Why it was recalled

Vital sign monitoring instrument may trigger an error message during blood pressures measurements, Specifically, when the initial inflation pressure is set in the high-pressure range.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On August 8, 2025, Schiller AG issued an Field Safety Notice to affected consignees via E-Mail. On September 6, 2025, Schiller issued a "Urgent Medical Device Recall Expansion their recall to additional products. Schiller asked consignees to take the following actions: 1. Read the Field Safety Notice (FSN) provided by the system manufacturer carefully 2. Wait until your system manufacturer contacts you for device testing or to arrange the remediation of your device(s). 3. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (As appropriate) 4. Please transfer this notice to other organizations on which this action has an impact. (As appropriate) 5. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents

Recall details

Recall number
Z-2672-2025
Status
Open, Classified
Initiated by the company
Aug 8, 2025
Posted by the FDA
Sep 29, 2025
Recalling firm
Schiller, Ag
Quantity
266 units
Product code
MHX
FDA submissions
K221056
Distribution
US Nationwide distribution in the states of MT, WI, IL.

Product description

Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006

Code information (models, lots, serial numbers)

REF: 1A.701307: UDI- 07613365003024 REF: 3.921002, UDI: 07613365001914 REF: 3.921030, UDI: 07613365003468 REF: 3.921031: UDI: 07613365003475 REF: 3.921006, UDI: 07613365000115 Affected Serial Numbers: 7010.000160 and lower, 7012.000356 and lower, 7013.000256 and lower, 7014.000152 and lower

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K221056 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page