Skip to content

GE HealthCare OEC 9900 Elite C-arm Systems recallZ-2342-2025 · GE Medical Systems, LLC · initiated Jul 2, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

GE HealthCare will send a customer letter providing safety instructions as an immediate mitigation, to communicate the risk of additional radiation and that the equipment should only be used by personnel properly trained and authorized in the use of ionizing radiation. A user should observe proper safety practices during equipment operation as required by the operator manual. GE HealthCare will correct all affected OEC 9800 and OEC 9900 Elite systems (28 units) by visually inspecting for the presence of lead shield tape on the collimator cover and replacing the collimator cover if it is found non-compliant, at no cost to the customer.

Recall details

Recall number
Z-2342-2025
Status
Open, Classified
Initiated by the company
Jul 2, 2025
Posted by the FDA
Aug 22, 2025
Recalling firm
GE Medical Systems, LLC
Product code
OWB
FDA submissions
K073543, K122234, K082781, K120613
Distribution
US Nationwide distribution.

Product description

GE HealthCare OEC 9900 Elite C-arm Systems.

Code information (models, lots, serial numbers)

GTIN: 00840682114349

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K073543, K122234, K082781, K120613 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97454).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page