ICEfx Cryoablation System, portable helium-free desktop… recallZ-2550-2025 · Boston Scientific Corporation · initiated Aug 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.

FDA root cause category: Process control

What the company is doing

An Urgent Medical Device Correction dated 8/18/25 was sent to customers. Customers should cease use of the systems listed in Table 1 and segregate them until a Boston Scientific representative has corrected this issue. A Boston Scientific Field Service Engineer will contact your facility to schedule the correction of your system. For patients in whom an impacted device has already been used, there are no specific recommendations beyond the local standard post-procedure care. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative. The Acknowledgement Form enclosed with this product correction must be completed and returned even if you do not have any affected units. An Urgent Medical Device Removal notification letter dated 8/18/25 was sent to a customer regarding the PM Kit ICEFX. Instructions: 1. Further distribution or use of any devices (Table 1) affected by this removal should cease immediately. Do NOT use affected devices and remove those devices from your facility s inventory. Segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page three. 5. Return affected devices. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.

Recall details

Recall number
Z-2550-2025
Status
Open, Classified
Initiated by the company
Aug 18, 2025
Posted by the FDA
Sep 11, 2025
Recalling firm
Boston Scientific Corporation
Quantity
18 units
Product code
GEH
FDA submissions
K234002
Distribution
Worldwide - US Nationwide and the countries of Canada, France, Germany and Italy. (Preventive Maintenance Kit distributed to Canada only.)

Product description

ICEfx Cryoablation System, portable helium-free desktop cryoablation unit, Material Numbers (UPN) FPRCH8000, FPRCH8000-02, H749396118000

Code information (models, lots, serial numbers)

FPRCH8000 UDI-DI: 00859019006370, Serial Numbers: ICP0305, ICP0369, ICP0338; FPRCH8000-02 UDI-DI: 00859019006417, Serial Numbers: 86643865, 108415728, 88589050, 108470986, 89392275, 108612524, 107799735, ICP0299, 107887794, 3779, 108254011, IF0096 108312221, 104733618; H749396118000 UDI-DI: 00191506014995, Serial Numbers: 106066356, 107851990

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.