Sterile Medline Convenience Kits recallZ-2592-2025 · Medline Industries, LP · initiated Aug 8, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Convenience kits labeled as sterile have not gone through the sterilization process.
FDA root cause category: Process control
What the company is doing
Customers were alerted via email and telephone beginning 8/8/25 of a potential issue with the products. An URGENT MEDICAL DEVICE RECALL notification letter dated 8/22/25 was sent to customers. This communication is a follow-up to the notification you received from your Medline Sales Representative and to acknowledge the subsequent return of impacted product to Medline. This communication is to provide formal documentation of the recall for your records. No additional action is required to be taken. Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Sterile Medline Convenience Kits. It has been identified that specific convenience kits labeled as sterile have not gone through the sterilization process. As a result, these kits are non-sterile. If the non-sterile kits are assumed sterile upon use, this could result in potential contamination of the sterile-field and an increased risk of patient infection that may require the use of intravenous or oral antibiotics. Refer to the enclosed list for all affected item number(s) and affected lot number(s). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax and to Medline Industries, LP. If you have any questions regarding this notification, please contact the Director, Regulatory Compliance at ([email protected]). or Product Recall Specialist at [email protected].
Recall details
- Recall number
- Z-2592-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 8, 2025
- Posted by the FDA
- Sep 26, 2025
- Recalling firm
- Medline Industries, LP
- Quantity
- 22 units
- Product code
- LRO
- FDA submissions
- K213481
- Distribution
- US Distribution to States of: FL, MD, NJ, TN
Product description
Sterile Medline Convenience Kits: 1) MAJOR PACK L-F, Model Number: DYNJ0382730O; 2) LB BASIC CUSTOM PACK, Model Number: DYNJ61038B
Code information (models, lots, serial numbers)
1) DYNJ0382730O, UDI-DI: 10198459117992(each), 40198459117993(case), Lot Number: 25GBC583; 2) DYNJ61038B, UDI-DI: 10195327596316(each), 40195327596317(case), Lot Number: 25GBF720;
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ0382730OMedline Industries, Inc. · General surgery tray1 recall2 reports
- Medline Industries, Inc. DYNJ61038BMedline Industries, Inc. · General surgery tray1 recall0 reports
Other devices under the same FDA submission
321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Centurion CVI4505Centurion Medical Products Corporation · General Surgery Tray3 recalls0 reports
- MEDLINE ECVC6355Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE ECVC1045Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ART255Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- Centurion CVI3880Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- MEDLINE ECVC1375Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE DYNDV2520Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ST11670ACenturion Medical Products Corporation2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC6840Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE ECVC6265Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97495).
- Z-2593-2025 · Open, ClassifiedSterile Medline Convenience Kits: 1) EXTREMITY PACK, Model Number: DYNJ45701B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.