Neuroguard IEP System recallZ-0131-2026 · Contego Medical LLC · initiated Sep 4, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Contego Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.
FDA root cause category: Device Design
What the company is doing
On 9/4/2025, communication and training to the field representatives started with in person letter distribution following immediately after. Customers were asked to do the following: Action Required 1) Inventory Check: Medtronic and Contego sales representatives will visit your facility to conduct an inventory count of devices from the affected lots. 2) Quarantine Affected Devices: Any identified devices will be removed from use and quarantined by the Medtronic or Contego team. 3) Return: Medtronic and Contego will remove affected devices. If replacements are desired, they can be provided during the same visit. 4) Confirmation: Sign and return the enclosed acknowledgment form to confirm receipt of this advisory and the completion of actions performed by Medtronic or Contego. 5) Please email the completed form to Contego Medical [email protected] as soon as possible and retain a copy of the form to provide to the representative at the time of their visit. If you have questions, contact the firm's Recall Support Team at [email protected] or call 919-459-7250 Extension 4.
Recall details
- Recall number
- Z-0131-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 4, 2025
- Posted by the FDA
- Oct 10, 2025
- Recalling firm
- Contego Medical LLC
- Quantity
- 568
- Product code
- NIM
- FDA submissions
- P240009
- Distribution
- US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.
Product description
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40
Code information (models, lots, serial numbers)
REF/UDI-DI/Lot(Expiration): NG-0740-140-2/00851616007613/Z2460707D(Z2460707D), Z2460733A(03/05/2027), Z2460766A(04/07/2027), Z2460774B(04/17/2027), Z2460780A(04/18/2027); NG-0640-140-2/00851616007590/Z2460746C(03/20/2027), Z2460774A(04/18/2027); NG-NV-7-40/00851616007620/Z2460776D(04/14/2027); NG-NV-6-40/00851616007651/Z2460776B(04/16/2027)
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Neuroguard IEP® System NG-0740-140-2Contego Medical, Inc. · Stent, Carotid2 recalls32 reports
- Neuroguard IEP® System NG-NV-6-40Contego Medical, Inc. · Stent, Carotid2 recalls9 reports
- Neuroguard IEP® System NG-NV-7-40Contego Medical, Inc. · Stent, Carotid2 recalls8 reports
- Neuroguard IEP® System NG-0640-140-2Contego Medical, Inc. · Stent, Carotid2 recalls7 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under P240009 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Neuroguard IEP® System NG-0730-140-2Contego Medical, Inc. · Stent, Carotid1 recall5 reports
- Neuroguard IEP® System NG-0630-140-2Contego Medical, Inc. · Stent, Carotid1 recall3 reports
- Neuroguard IEP® System NG-NV-6-30Contego Medical, Inc. · Stent, Carotid1 recall2 reports
- Neuroguard IEP® System NG-NV-7-30Contego Medical, Inc. · Stent, Carotid1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page