Skip to content

Extractor Pro RX recallZ-2618-2025 · Boston Scientific Corporation · initiated Aug 28, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.

FDA root cause category: Process control

What the company is doing

On August 28, 2025 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Instructions: 1. Immediately cease further distribution or use of any remaining affected devices. Segregate the units in a secure place until they can be returned. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Return affected devices.

Recall details

Recall number
Z-2618-2025
Status
Open, Classified
Initiated by the company
Aug 28, 2025
Posted by the FDA
Sep 23, 2025
Recalling firm
Boston Scientific Corporation
Quantity
156 units
Product code
FGE
FDA submissions
K102082
Distribution
US Nationwide distribution in the states of CA, CO, GA, ID, IL, IN, MA, ME, MI, MS, NC, OH, OK, PA, SD, TN, TX, VA, WA.

Product description

Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);

Code information (models, lots, serial numbers)

Instructions for Use: 51878205; Pre-Printed Ouch: 51368417; 1. UPN (REF): M00547030 (9-12mm Below); UDI-DI: 08714729790280; Lot Numbers: 36398057; 2. UPN (REF): (1) M00547000(9-12mm Above); UDI-DI: 08714729790266; Lot Numbers: 36426495;

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page