MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217 recallZ-2662-2025 · Siemens Medical Solutions USA, Inc · initiated Aug 28, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthcare notified consignees on about 08/28/2025 via letter. Consignees were informed that Siemens Customer Service Organization will conduct an inspection of affected units to address the issue. If customers identify issues, they are cautioned to not reboot the system, to call their local Siemens Healthineers service organization, and grant access to the system. Siemens will inspect devices and perform a correction, if needed. Consignees were also instructed to ensure all users or customers are notified of the issue and to maintain the Customer Safety Advisory Notice. Consignees were also requested to confirm receipt and acknowledgement of the notification. Siemens expanded the recall and notified consignees on about 10/01/2025 via letter. These notifications included the same information as initially sent to consignees. A follow up communication was sent to consignees on about 10/15/2025 via letter. This informed users of further details and additional necessary safety actions. If a specific "Magnent Supervision" error message appears on the host computer, no further scans can be performed, and you are requested to: not reboot the system, remove the patient from the system, lock/secure all access to the scanning room with the system remaining energized at field, post a DO NOT USE notice on or near the device, and call the Siemens Healthineers service organization. In addition, Siemens reminded users that the Magnet Stop button is only intended for emergencies. DO NOT initiate a manual quench of the magnet unless there is an emergency. Ensure all designated emergency exit routes remain unobstructed and are marked. Consignees were also instructed to ensure all users or customers are notified of the issue, to maintain the Customer Safety Advisory Notice is retained in product related records, to confirm receipt and acknowledgement of the notification, and if the affected system is no longer in the consignee's possession, notify the new owner. ***Updat
Recall details
- Recall number
- Z-2662-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 28, 2025
- Posted by the FDA
- Sep 30, 2025
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 39 units
- Product code
- LNH
- FDA submissions
- K220589, K232494, K250443
- Distribution
- Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Product description
MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Model Number: 11516217. UDI Numbers: (01)04056869299273(21)212087, (01)04056869299273(21)212098, (01)04056869299273(21)212088, (01)04056869299273(21)212068, (01)04056869299273(21)212081, (01)04056869299273(21)212018, (01)04056869299273(21)212041, (01)04056869299273(21)212038, (01)04056869299273(21)212047, (01)04056869299273(21)212019, (01)04056869299273(21)212037, (01)04056869299273(21)212032, (01)04056869299273(21)212049, (01)04056869299273(21)212017, (01)04056869299273(21)212064, (01)04056869299273(21)212096, (01)04056869299273(21)212065, (01)04056869299273(21)212016, (01)04056869299273(21)212023, (01)04056869299273(21)212056, (01)04056869299273(21)212054, (01)04056869299273(21)212063, (01)04056869299273(21)212051, (01)04056869299273(21)212055, (01)04056869299273(21)212021, (01)04056869299273(21)212060, (01)04056869299273(21)212011, (01)04056869299273(21)212066, (01)04056869299273(21)212074, (01)04056869299273(21)212090, (01)04056869299273(21)212034, (01)04056869299273(21)212029, (01)04056869299273(21)212097, (01)04056869299273(21)212099, (01)04056869299273(21)212039, (01)04056869299273(21)212059, (01)04056869299273(21)212030, (01)04056869299273(21)212035, (01)04056869299273(21)212036. Serial Numbers: 212087, 212098, 212088, 212068, 212081, 212018, 212041, 212038, 212047, 212019, 212037, 212032, 212049, 212017, 212064, 212096, 212065, 212016, 212023, 212056, 212054, 212063, 212051, 212055, 212021, 212060, 212011, 212066, 212074, 212090, 212034, 212029, 212097, 212099, 212039, 212059, 212030, 212035, 212036.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MAGNETOM Skyra Fit 11516217Siemens Healthcare GmbH · System, nuclear magnetic resonance spectroscopic2 recalls1 report
Other devices under the same FDA submission
5 devices cleared under K220589, K232494, K250443 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAGNETOM Skyra 10849580Siemens Healthcare GmbH · Coil, magnetic resonance, specialty2 recalls2 reports
- MAGNETOM Avanto 10849579Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging1 recall1 report
- MAGNETOM Avanto 10849578Siemens Healthcare GmbH · System, nuclear magnetic resonance spectroscopic1 recall0 reports
- MAGNETOM Avanto Fit 11516216Siemens Healthcare GmbH · System, nuclear magnetic resonance spectroscopic1 recall0 reports
- MAGNETOM Viato.Mobile 11689652Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97530).
- Z-2654-2025 · Open, ClassifiedBiograph mMR. Model Number: 10433372.
- Z-2655-2025 · Open, ClassifiedBIOGRAPH One (DE). Model Number: 11689172.
- Z-2656-2025 · Open, ClassifiedMAGNETOM Cima.X (DE). Model Number: 11647158.
- Z-2658-2025 · Open, ClassifiedMAGNETOM Lumina (DE). Model Number: 11344916.
- Z-2659-2025 · Open, ClassifiedMAGNETOM Prisma. Model Number: 10849582.
- Z-2660-2025 · Open, ClassifiedMAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.
- Z-2661-2025 · Open, ClassifiedMAGNETOM Skyra fit. Model Number: 10849580.
- Z-2663-2025 · Open, ClassifiedMAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.
- Z-2664-2025 · Open, ClassifiedMAGNETOM Verio. Model Number: 10276755.
- Z-2665-2025 · Open, ClassifiedMAGNETOM Verio Dot. Model Number: 10684333.
- Z-2666-2025 · Open, ClassifiedMAGNETOM Verio Dot Upgrade. Model Number: 10684334.
- Z-2667-2025 · Open, ClassifiedMAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.
- Z-2668-2025 · Open, ClassifiedMAGNETOM Vida Fit. Model Number: 11410481.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page