JuggerKnot Soft Anchors recallZ-2619-2025 · Biomet, Inc. · initiated Sep 9, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to [email protected]. This form must be returned even if you do not have affected products available to return in your territory. b. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. c. M

Recall details

Recall number
Z-2619-2025
Status
Open, Classified
Initiated by the company
Sep 9, 2025
Posted by the FDA
Sep 23, 2025
Recalling firm
Biomet, Inc.
Quantity
40
Product code
MBI
FDA submissions
K110879
Distribution
US Nationwide distribution.

Product description

Juggerknot Mini Soft Anchors, Model/Catalog Number: 912080

Code information (models, lots, serial numbers)

UDI-DI: 00880304520950; Lot Number 0002713620

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.