Skip to content

Equinoxe recallZ-0535-2026 · Exactech, Inc. · initiated Sep 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.

FDA root cause category: Release of Material/Component prior to receiving test results

What the company is doing

On 09/05/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing that, during routine inventory review, it was discovered that the a number of parts/products, which were pending design verification testing were incorrectly placed in active internal inventory rather than designated quarantine locations. Several part/products were distributed to customers. Customers are instructed to: "Carefully review this notification to ensure that they fully understand the issue and all actions required. "Immediately identify, quarantine, and return any of the subject devices in their inventory and/or customer s inventory. "Stop and cease use of any affected devices. "Notify all their accounts that may have received these products and extend this recall notice, including the Description of Issue and Potential Clinical Impact, to them. "Complete and Return the attached Recall Inventory Response Form to Exactech via email at [email protected]. For questions or assistance - contact Recall Support Team at [email protected]

Recall details

Recall number
Z-0535-2026
Status
Open, Classified
Initiated by the company
Sep 5, 2025
Posted by the FDA
Nov 17, 2025
Recalling firm
Exactech, Inc.
Quantity
1
Product code
PHX
FDA submissions
K042021
Distribution
U.S.: SC O.U.S.: N/A

Product description

Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 42mm Model/Catalog Number: 321-59-42 Software Version: N/A Product Description: rTSA Reamer, 42mm Component: Not a component

Code information (models, lots, serial numbers)

Lot Code: Part Number: 321-59-42, UDI-DI: 10885862593009, Lot Number: 410398-013

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

72 devices cleared under K042021 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97588).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page