TactiFlex Sensor Enabled recallZ-0112-2026 · Abbott · initiated Sep 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
FDA root cause category: Package design/selection
What the company is doing
Abbott issued a Medical Device Advisory to its consignees on 09/10/2025 by letter delivered via Federal Express. The notice explained the issue, impact and associated risk to the patient and requested the following: "Action Requested/Recommendations/Next Steps . Continue following the product IFU by inspecting the packaging and catheter for damage and/or kinking prior to use. . Review the guidance in Appendix A information on proper removal technique of the TactiFlex" Ablation Catheter, Sensor Enabled" from the packaging tray, ensuring to inspect the electrodes and catheter tip carefully for integrity and overall condition prior to use. . Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal from the packaging. The deployment of the new packaging has begun. Please complete the provided Acknowledgement Form with the assistance of your Abbott representative. This form ensures you are aware of the advisory notice and have been given an opportunity to ask questions. Please forward this notice to anyone in your organization who may need to be notified and maintain this notice for your records to ensure effectiveness of this communication. Should you have questions about this issue, please contact your local Abbott representative or Abbott Support at 1-855-478-5833 (U.S), 8:30 a.m. - 5:30 p.m. Central Standard Time, Monday thru Friday."
Recall details
- Recall number
- Z-0112-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 10, 2025
- Posted by the FDA
- Oct 10, 2025
- Recalling firm
- Abbott
- Quantity
- 11810 units
- Product code
- OAE
- FDA submissions
- P220013
- Distribution
- Worldwide
Product description
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
GTIN 05415067034618, Batch Numbers: 10093213, 10093588, 10095366, 10096269, 10118676, 10131389, 10151333, 10180969, 10286877, 10315661, 10317649, 10329929, 10332297, 10344177, 10350400, 10360030, 10374171, 10375844, 10418954, 10431765, 10433617, 10450462, 10492093, 10494298, 10537000, 10537747, 10538321, 10557710, 10557878, 10558450, 10559655, 10569703, 10570507, 10571902, 10571907, 10574230, 10575746, 10575752, 10577567, 10578953, 10579724, 10580699, 10581612, 10581681, 10583203, 10583793, 10587033, 10587043, 10587653, 10589068, 10589921, 10591178, 10591238, 10592227, 10606628, 10606720, 10608438, 10608708, 10611469, 10611683, 10612686, 10614973, 10616329, 10618038, 10619154, 10621180, 10622172, 10622962, 10622967, 10625101, 10630682, 10630687, 10632744, 10633145, 10634830, 10636059, 10637023, 10639373, 10639374, 10639376, 10639377, 10641739, 10641741, 10642345, 10643887, 10644681, 10645838, 10646429, 10648576, 10662844, 10665077, 10666137, 10667210, 10667217, 10668065, 10668185, 10670353, 10670360, 10671604, 10671663, 10672934, 10674544, 10675329, 10675541, 10677719, 10677722, 10678784, 10680052, 10680962, 10684449, 10691502, 10691504, 10691507, 10692805, 10696082, 10696083, 10696084, 10705825, 10708749, 10708800, 10709735, 10711022, 10711803, 10712639, 10712641, 10714607, 10714611, 10715850, 10717049, 10717050, 10719520, 10721238, 10721239, 10725441, 10725442, 10747196, 10749817, 10749820, 10749823, 10751484, 10762186, 10762464, 10764517, 10764523, 10765657, 10766897, 10767820, 10769643, 10770889, 10772623, 10772625, 10773775, 10773811, 10776251, 10777548, 10778157, 10779451, 10808389, 10808390, 10808393, 10811066, 10811067, 10811262, 10821034, 10827717, 10829276, 10829990, 10830744, 10832015, 10832019, 10832727, 10834411, 10834412, 10835187, 10836468, 10836482, 10836911, 10837558, 10837559, 10839959, 10839960, 10841679, 10841690, 10842962, 10845198, 10845199, 10846600, 10847688, 10847695, 10850757, 10850761, 10856004, 10856065, 10858260, 10867490, 10867495, 10868706, 10873737, 10873924, 10875772, 10875773, 10877194, 10897765, 10899500, 10899504, 10902557, 10902569, 10905178, 10905907, 10906368, 10907259, 10909995, 10910050, 10911332, 10911333, 10913977, 10915227, 10915230, 10920111, 10922315, 10922528, 10924643, 10924802, 10927218, 10954504, 10954507, 10955716, 10956520, 10956521, 10958150, 10959364, 10959373, 10960184, 10962699, 10965714, 10970014, 10970054, 10985513, 10989401, 10990017, 10990759, 10992074, 10992078, 10987781.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TactiFlex™ Sensor Enabled™ A-TFSE-FJSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation2 recalls2 deaths37 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under P220013 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- TactiFlex™ Sensor Enabled™ A-TFSE-DFSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall6 deaths241 reports
- TactiFlex™ Sensor Enabled™ A-TFSE-FSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall18 reports
- TactiFlex™ Sensor Enabled™ A-TFSE-DDSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall14 reports
- TactiFlex™ Sensor Enabled™ A-TFSE-DSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall12 reports
- TactiFlex™ Sensor Enabled™ A-TFSE-JSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 report
- TactiSys™ Quartz Equipment TSQ-RF-TFSE-CBLSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 report
- TactiFlex™ Sensor Enabled™ A-TFSE-DF-XSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 report
- TactiFlex™ Sensor Enabled™ A-TFSE-FJ-XSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 97595).
- Z-0109-2026 · Open, ClassifiedTactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter
- Z-0110-2026 · Open, ClassifiedTactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DD, (BiD Curve D-D), REF A-TFSE-DD; cardiac catheter
- Z-0111-2026 · Open, ClassifiedTactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DF, (BiD Curve D-F), REF A-TFSE-DF; cardiac catheter
- Z-0113-2026 · Open, ClassifiedTactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, F, (UniD Curve F), REF A-TFSE-F; cardiac catheter
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page