The failure occurs if a user on the Omnipod 5 iOS… recallZ-0132-2026 · Insulet Corporation · initiated Sep 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Insulet Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

FDA root cause category: Software design

What the company is doing

Insulet notified both healthcare professionals and users on about 09/11/2025 via URGENT: Medical Device Correction letter. Instructions included that an update to the application will be pushed to all devices and should be implemented by users. Users were also requested to confirm receipt and acknowledgement of the notification but selecting the "acknowledgement button" in the email.

Recall details

Recall number
Z-0132-2026
Status
Open, Classified
Initiated by the company
Sep 11, 2025
Posted by the FDA
Oct 10, 2025
Recalling firm
Insulet Corporation
Quantity
174,013 units
Product code
QRX
FDA submissions
K231826
Distribution
US Nationwide distribution.

Product description

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Code information (models, lots, serial numbers)

Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X