Skip to content

Omnipod 5Recalls & reported problems across all models

Insulet Corporation · 8 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one Omnipod 5 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
136,063
Deaths reported
100
Injuries reported
14,724
Recalls
5

We found 8 models of Omnipod 5 by Insulet Corporation in the FDA's device registry (GUDID). Together they are named in 136,063 adverse event reports in the FDA's MAUDE database, received between Mar 8, 2022 and Sep 30, 2026 — classified as 100 deaths, 14,724 injuries, 121,236 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (64,011 reports).

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 2,037 reportsDec 2023: 1,694 reportsJan 2024: 2,415 reportsFeb 2024: 730 reportsMar 2024: 2,595 reportsApr 2024: 3,348 reportsMay 2024: 2,901 reportsJun 2024: 3,050 reportsJul 2024: 3,174 reportsAug 2024: 3,199 reportsSep 2024: 3,451 reportsOct 2024: 3,485 reportsNov 2024: 2,849 reportsDec 2024: 1,621 reportsJan 2025: 3,338 reportsFeb 2025: 2,483 reportsMar 2025: 3,678 reportsApr 2025: 2,541 reportsMay 2025: 3,530 reportsJun 2025: 4,361 reportsJul 2025: 4,184 reportsAug 2025: 3,807 reportsSep 2025: 3,538 reportsOct 2025: 4,324 reportsNov 2025: 5,411 reportsDec 2025: 4,847 reportsJan 2026: 3,942 reportsFeb 2026: 5,058 reportsMar 2026: 5,085 reportsApr 2026: 6,229 reportsMay 2026: 5,321 reportsJun 2026: 4,682 reportsJul 2026: 1,945 reportsAug 2026: 1,344 reportsSep 2026: 1,110 reportsOct 2026: 0 reports2024202520266,229
Last 36 months. The most recent month may be incomplete.

Omnipod 5 recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 20, 2026Open, Classified

    External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

    Z-2603-2026 · Root cause: Process design · names 3 of these models · Recall details · FDA record ↗

  2. Mar 12, 2026Open, Classified

    Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

    Z-1797-2026 · Root cause: Process design · Recall details · FDA record ↗

  3. Sep 11, 2025Open, Classified

    The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

    Z-0132-2026 · Root cause: Software design · Recall details · FDA record ↗

  4. May 23, 2025Open, Classified

    Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.

    Z-2016-2025 · Root cause: Software design · Recall details · FDA record ↗

  5. Feb 28, 2023Open, Classified

    An error message was received when using the Omnipod 5 App on compatible smartphones that prevented phone control users from controlling the Omnipod 5 Automated Insulin Delivery System from their Omnipod 5 App which may cause a delay in therapy. To reduce the volume of inquiries Insulet was receiving from customers, an email was sent to all phone control users and the cause of the error message was resolved by reverting to prior certificates on the Insulet cloud to secure communications between the Omnipod 5 App sotware and the Insulet Cloud. This issue was limited to only certain compatible Android smartphones and did not affect the Omnipod 5 Pods or the dedicated locked-down Controller provided by Insulet. All other Omnipod 5 users were still able to manage their insulin with the device without use of the smartphone app and affected users were able to switch to the locked-down control provided to all users when they initiated the device during the time the App was unavailable to them. Further, the Omnipod itself continued to deliver insulin as per its pre-programmed settings so although affected users who did not have the locked-down controller readily available may not have been able to command insulin boluses during this time, they continued to receive basal insulin without interruption. Lastly, users of the Omnipod 5 were advised during training to always have backup supplies ready in the event of a device malfunction which would include alternative means to deliver insulin subcutaneously until the issue is resolved to minimize disruption to their insulin regimen. There were 2,168 complaints received out of 7,838 users with the Android compatible smartphones affected and 9 MDRs were submitted, however, there were no serious injuries reported or reports of erroneous results as the primary issue was a delay in therapy. This defect cannot cause false results that could negatively impact patients.

    Z-2618-2023 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported Omnipod 5 problems

Device problems

  • Unintended Movement64,011
  • Loss of or Failure to Bond57,949
  • Material Twisted/Bent23,939
  • Insufficient Device Problem Information14,167
  • Fluid/Blood Leak13,651
  • Failure to Fire13,619
  • Difficult to Insert10,224
  • Activation, Positioning or Separation Problem6,044

Patient problems

  • Hyperglycemia104,994
  • No Clinical Signs, Symptoms or Conditions24,802
  • Elevated Ketones/ Diabetic Ketoacidosis4,644
  • Erythema3,395
  • Vomiting3,285
  • Pain2,865
  • Hemorrhage/Blood Loss/Bleeding2,675
  • Hypoglycemia2,404

Omnipod 5 models

8 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Omnipod 5 adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Insulet Corporation products

All Insulet Corporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page