Skip to content

XERF recallZ-0146-2026 · Lutronic Corporation · initiated Sep 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Lutronic Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Manufacturing defects in devices manufactured from 4/1/25 to 8/20/25 can on rare occasions result in adverse events for patients.

FDA root cause category: Process change control

What the company is doing

An URGENT-Field Safety Corrective Action - XERF Tips notification dated 9/4/25 was mailed to consignees notifying them of this recall. Consignees are instructed to immediately stop use of affected devices and to return all units utilizing the provided shipping label. Consignees with any questions can contact Customer Service at 1-800-886-2966 or by email at [email protected].

Recall details

Recall number
Z-0146-2026
Status
Open, Classified
Initiated by the company
Sep 4, 2025
Posted by the FDA
Oct 14, 2025
Recalling firm
Lutronic Corporation
Quantity
7,490 units
Product code
GEI
FDA submissions
K251327
Distribution
Worldwide - US Nationwide distribution in the states of CA, IL, MA, NJ, NY, PA, TX, and UT. The countries of Canada, South Korea, Hong Kong, Japan, Singapore, Vietnam, Myanmar, Philippines, Taiwan, and Thailand.

Product description

XERF EFFECTOR 60. Electrosurgical unit.

Code information (models, lots, serial numbers)

Model/Catalog Number: 1101210000, 1101210100, 1101210200; UDI: 08809447657752, SERIAL/LOT: MP60J25C02, MP60J25G05, MP60J25G06, MP60J25H03, and MP60J25H01. Devices manufactured from 04/01/25 to 08/20/25.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page