XERFRecalls & reported problems across all models
Lutronic Corporation · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one XERF model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 1
We found 2 models of XERF by Lutronic Corporation in the FDA's device registry (GUDID). Together they are named in 18 adverse event reports in the FDA's MAUDE database, received between Sep 11, 2025 and May 7, 2026 — classified as 2 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (10 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
XERF recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 4, 2025Open, Classified
Manufacturing defects in devices manufactured from 4/1/25 to 8/20/25 can on rare occasions result in adverse events for patients.
Z-0146-2026 · Root cause: Process change control · Recall details · FDA record ↗
Most reported XERF problems
Device problems
- Insufficient Device Problem Information10
- Defective Component4
- Gas/Air Leak4
- Improper or Incorrect Procedure or Method2
- Material Deformation2
Patient problems
- Burn(s)12
- Full thickness (Third Degree) Burn4
- No Clinical Signs, Symptoms or Conditions2
XERF models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- XERF XERF EFFECTOR 60(900)Electrosurgical, Cutting & Coagulation & Accessories1 recall0 reports
- XERFElectrosurgical, Cutting & Coagulation & Accessories18 reports
Recent XERF adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT SPARKING OCCURRED ON A PATIENT FACE AND PATIENT FELT "POPPING SENSATION" DURING A TREATMENT USING XERF 2.0. TREATMENT WAS IMMEDIATELY STOPPED AND EFFECTOR TIP WAS REPLACED. SETTINGS USED FOR TREATMENT INCLUDED DEEP MODE, CRYO 1, LEVEL 6.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25116707 ↗MalfunctionMay 7, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT SPARKING OCCURRED ON A PATIENT FACE AND PATIENT FELT "POPPING SENSATION" DURING A TREATMENT USING XERF 2.0. TREATMENT WAS IMMEDIATELY STOPPED AND EFFECTOR TIP WAS REPLACED. SETTINGS USED FOR TREATMENT INCLUDED DEEP MODE, CRYO 1, LEVEL 6. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM REACHED OUT TO THE SITE TO INVESTIGATE THE EVENT. DESPITE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION, THE SITE DID NOT PROVIDE ANY ADDITIONAL INFORMATION. DUE TO LACK OF COMMUNICATION WITH THE SITE, CYNOSURE LUTRONIC CLINICAL TEAM COULD NOT PROVIDE CLINICAL SUPPORT. XERF TIP HAS NOT YET BEEN RETURNED FOR A DEVICE EVALUATION, BUT A PHOTO WAS PROVIDED DURING INITIAL COMMUNICATION. SURFACE DAMAGE WAS OBSERVED ON THE TIP. ONCE TIP HAS BEEN RETURNED AND EVALUATED, A CYNOSURE FIELD SERVICE ENGINEER (FSE) WILL BE ABLE TO IDENTIFY ROOT CAUSE. THE EVENT IS REPORTABLE PER FDA MEDICAL DEVICE REPORTING TITLE 21 CFR PART 803 SINCE A MALFUNCTION OCCURRED AND COULD LIKELY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF THE MALFUNCTION WERE TO RECUR. WE ARE REPORTING THIS AS AN INITIAL MDR AND WILL SUBMIT A FINAL MDR ONCE INVESTIGATION HAS BEEN COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25114265 ↗MalfunctionApr 10, 2026Gas/Air Leak
IT WAS REPORTED THAT A PATIENT WAS BURNED DURING A TEST TREATMENT WITH XERF. XERF HANDPIECE WAS ALSO LEAKING ICD OUT OF THE TIP AND CANNISTERS WERE REGISTERING EMPTY EVEN WHEN HALF FULL.
Patient problems: Burn(s)
MAUDE report 24844572 ↗MalfunctionApr 10, 2026Gas/Air Leak
IT WAS REPORTED THAT A TREATMENT PROVIDER BURNED THEMSELVES WHILE TESTING WITH XERF. XERF HANDPIECE WAS ALSO LEAKING ICD OUT OF THE TIP AND CANNISTERS WERE REGISTERING EMPTY EVEN WHEN HALF FULL.
Patient problems: Burn(s)
MAUDE report 24844319 ↗MalfunctionApr 10, 2026Gas/Air Leak
IT WAS REPORTED THAT A PATIENT WAS BURNED DURING A TEST TREATMENT WITH XERF. XERF HANDPIECE WAS ALSO LEAKING ICD OUT OF THE TIP AND CANNISTERS WERE REGISTERING EMPTY EVEN WHEN HALF FULL. ON (B)(6) 2026, A TRAINED CYNOSURE LUTRONIC FIELD SERVICE ENGINEER (FSE) WENT TO THE SITE AND PERFORMED A DEVICE EVALUATION. FSE OBSERVED DEVICE WAS CONNECTED TO EXTENDED OUTLET ADAPTER WITH MULTIPLE DEVICES CONNECTED. ADVISED SITE TO MOVE DEVICE FROM ADAPTER TO A DIRECT OUTLET CONNECTION. FSE THEN INSPECTED THE HANDPIECE AND DID NOT FIND ANY DAMAGE VISIBLY OR INTERNALLY. FSE INSERTED CRYO INTO THE DEVICE AND SAW IT IMMEDIATELY LEAKING FROM THE HANDPIECE. TO RESOLVE THE ISSUE, FSE REPLACED THE HANDPIECE. FSE NOTES THAT THE CLIENT WAS NOT FULLY TRAINED. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM CONTACTED THE SITE TO FURTHER INVESTIGATE THE EVENT. SITE SHARED THAT THE INITIAL TRAINING SCHEDULED FOR (B)(6) 2026 WAS NOT COMPLETED DUE TO TECHNICAL ISSUES. A RETRAINING SESSION WAS COMPLETED ON (B)(6) 2026. ALTHOUGH ADDITIONAL INFORMATION WAS REQUESTED, THERE WAS NO ADDITIONAL INFORMATION MADE AVAILABLE FROM THE CUSTOMER SITE DESPITE MULTIPLE REQUESTS FROM CYNOSURE LUTRONIC CLINICAL. AS A RESULT, EX
Patient problems: Burn(s)
MAUDE report 24844316 ↗MalfunctionApr 10, 2026Gas/Air Leak
IT WAS REPORTED THAT A TREATMENT PROVIDER BURNED THEMSELVES WHILE TESTING WITH XERF. XERF HANDPIECE WAS ALSO LEAKING ICD OUT OF THE TIP AND CANNISTERS WERE REGISTERING EMPTY EVEN WHEN HALF FULL. ON 17 MARCH 2026, A TRAINED CYNOSURE LUTRONIC FIELD SERVICE ENGINEER (FSE) WENT TO THE SITE AND PERFORMED A DEVICE EVALUATION. FSE OBSERVED DEVICE WAS CONNECTED TO EXTENDED OUTLET ADAPTER WITH MULTIPLE DEVICES CONNECTED. ADVISED SITE TO MOVE DEVICE FROM ADAPTER TO A DIRECT OUTLET CONNECTION. FSE THEN INSPECTED THE HANDPIECE AND DID NOT FIND ANY DAMAGE VISIBLY OR INTERNALLY. FSE INSERTED CRYO INTO THE DEVICE AND SAW IT IMMEDIATELY LEAKING FROM THE HANDPIECE. TO RESOLVE THE ISSUE, FSE REPLACED THE HANDPIECE. FSE NOTES THAT THE CLIENT WAS NOT FULLY TRAINED. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM CONTACTED THE SITE TO FURTHER INVESTIGATE THE EVENT. SITE SHARED THAT THE INITIAL TRAINING SCHEDULED FOR FEBRUARY 19, 2026 WAS NOT COMPLETED DUE TO TECHNICAL ISSUES. A RETRAINING SESSION WAS COMPLETED ON MARCH 19, 2026. ALTHOUGH ADDITIONAL INFORMATION WAS REQUESTED, THERE WAS NO ADDITIONAL INFORMATION MADE AVAILABLE FROM THE CUSTOMER SITE DESPITE MULTIPLE REQUESTS FROM CYNOSURE LUTRONIC CLINIC
Patient problems: Burn(s)
MAUDE report 24844308 ↗MalfunctionApr 3, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT EXPERIENCED A BURN ON THE FACE DURING TREATMENT PERFORMED ON (B)(6) 2026 WITH A XERF DEVICE. TIME THE DEVICE WAS SUBJECTED TO FULL FUNCTIONALITY TESTING, VERIFIED DEVICE SET UP AND CONFIRMED POWER MEASUREMENT. THE DEVICE WAS FOUND TO BE WORKING WITHIN PUBLISHED SPECIFICATIONS. VISUAL INSPECTION OF THE EFFECTOR TIP USED DURING TREATMENT OBSERVED A SMALL BLACK DOT ON THE TIPS SURFACE, WHICH WAS REPORTED TO NOT BE PRESENT PRIOR TO TREATMENT. AN IMAGE WAS TAKEN OF THIS DAMAGE, AND ITS LOCATION SUGGESTS CARBONIZATION OCCURRED AT THE NTC LINE TIP LOT AND MANUFACTURING DATE WAS CHECKED AND IT WAS CONFIRMED TO BE PRODUCED BEFORE THE IMPLEMENTATION OF THE RESISTANCE INSPECTION THAT WAS COMPLETED IN (B)(6) 2025. ONCE THE TIP HAS BEEN RETURNED AND EVALUATED; ROOT CAUSE CAN BE BETTER VERIFIED. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM CONTACTED THE TREATMENT CLINIC FOR ADDITIONAL INFORMATION. IT WAS DETERMINED THAT THE TREATMENT PARAMETERS USED WERE WITHIN THE RECOMMENDED CLINICAL GUIDELINES OF THE QUICKSTART GUIDE (QSG). IT IS UNDETERMINED WHAT CAUSED THE REPORTED BURNING, IT IS POSSIBLE THAT THE EFFECTOR TIP MAY HAVE NOT HAD FULL CONTACT WITH THE TREATMEN
Patient problems: Burn(s)
MAUDE report 24783599 ↗MalfunctionApr 3, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT EXPERIENCED A BURN ON THE FACE DURING TREATMENT PERFORMED ON (B)(6) 2026 WITH A XERF DEVICE.
Patient problems: Burn(s)
MAUDE report 24781378 ↗MalfunctionFeb 13, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT DURING TREATMENT, AFTER APPROXIMATELY 450 PULSES THE DEVICE RELEASED AN UNEXPECTED SOUND AND SPARK FROM THE EFFECTOR. THE TREATMENT WAS INTERRUPTED AND A BURN MARK ON THE EDGE OF THE EFFECTOR TIP WAS OBSERVED WHERE THE SPARK WAS REPORTED TO HAVE OCCURRED CAUSING A BURN ON THE PATIENTS SKIN. THE EFFECTOR WAS REPLACED AND TREATMENT CONTINUED. A TRAINED CYNOSURE LUTRONIC FIELD SERVICE ENGINEER (FSE) WENT TO THE TREATMENT PROVIDER SITE FOR A DEVICE EVALUATION. DURING THIS TIME, THE DEVICE CALIBRATION WAS TESTED AND FOUND TO BE WORKING WITHIN PUBLISHED SPECIFICATIONS. HOWEVER, CUSTOMER PROVIDED IMAGES OF THE FIRST EFFECTOR TIP THAT WAS USED OBSERVED VISUAL DAMAGE. IT IS UNDETERMINED WHAT CAUSED THIS DAMAGE. THE TIP WAS REQUESTED FOR ADDITIONAL EVALUATION, HOWEVER, HAS NOT BEEN RETURNED. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM MADE CONTACT WITH THE TREATMENT PROVIDER AND DETERMINED THAT THE TREATMENT PARAMETERS USED FOR BOTH EFFECTOR TIPS WERE WITHIN THE NORMAL RANGES OUTLINED IN THE QUICKSTART GUIDE (QSG). THE PATIENT WAS EVALUATED BY THE TREATMENT CLINIC'S MEDICAL DIRECTOR AND RECEIVED PREVENTATIVE TREATMENT; IT WAS CONFIRMED THAT THE PATIENT HAS RECOVERED
Patient problems: Burn(s)
MAUDE report 24336638 ↗MalfunctionFeb 13, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT DURING TREATMENT, AFTER APPROXIMATELY 450 PULSES THE DEVICE RELEASED AN UNEXPECTED SOUND AND SPARK FROM THE EFFECTOR TIP. THE TREATMENT WAS INTERRUPTED AND A BURN MARK ON THE EDGE OF THE EFFECTOR TIP WAS OBSERVED WHERE THE SPARK WAS REPORTED TO HAVE OCCURRED CAUSING A REPORTED BURN ON THE PATIENTS SKIN. THE EFFECTOR WAS REPLACED AND TREATMENT CONTINUED.
Patient problems: Burn(s)
MAUDE report 24332414 ↗InjuryJan 22, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT FOLLOWING MONOPOLAR RADIOFREQUENCY TREATMENT ON THE FACE WITH A XERF DEVICE, THE TREATMENT PROVIDER OBSERVED A BLISTER ON THE UPPER LATERAL SIDE OF THE NEM PAD AREA.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 24146038 ↗InjuryJan 22, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT FOLLOWING MONOPOLAR RADIOFREQUENCY TREATMENT ON THE FACE WITH A XERF DEVICE, THE TREATMENT PROVIDER OBSERVED A BLISTER ON THE UPPER LATERAL SIDE OF THE NEM PAD AREA. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM MADE CONTACT WITH THE TREATMENT PROVIDER WHO REPORTED THAT THE PATIENT REPORTED A WARM SENSATION UNDER THE NEM PAD, HOWEVER, IT WAS TOLERABLE AND DID NOT INTERRUPT THE TREATMENT SESSION. THE BLISTERING WAS OBSERVED NEAR THE APPLIED DISPOSABLE PATCH AND BECAME VISIBLE NEAR THE END OF THE TREATMENT SESSION. THE CLINICAL DETAILS WERE REQUESTED ON MULTIPLE OCCASIONS, DESPITE THESE REQUESTS, THE TREATMENT PROVIDER DID NOT DISCLOSE ANY INFORMATION TO THE CLINICAL TEAM. THE OPERATORS MANUAL (OM) STATES THAT THAT "PATIENT RETURN PAD MUST BE THOROUGHLY ADHERED TO THE PROPER AREA OF THE PATIENT'S BODY AS SPECIFIED BY THE MANUFACTURER. AVOID BONY-PROMINENCES AND HAIR WHEN ADHERING THE PATIENT RETURN PAD TO THE PATIENT TO DECREASE THE RISK OF ADVERSE EFFECTS OR BURNS. PATIENT RETURN PAD CABLES SHOULD BE POSITIONED IN SUCH A WAY THAT THEY DO NOT COME INTO CONTACT WITH THE PATIENT OR OTHER CABLES." IMAGES OF THE PATIENT BURNS WERE PROVIDED ON 1/14/2026 AND WERE FOR
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 24145263 ↗MalfunctionDec 31, 2025Insufficient Device Problem Information; Defective Component; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT EXPERIENCED SKIN BURNING AND SCABBING AFTER XERF TREATMENT ON THE FACE.
Patient problems: Burn(s)
MAUDE report 23933926 ↗MalfunctionDec 31, 2025Insufficient Device Problem Information; Defective Component; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT EXPERIENCED SKIN BURNING AND SCABBING AFTER XERF TREATMENT ON THE FACE. A MEMBER OF THE CYNOSURE LUTRONIC CLINICAL TEAM REVIEWED THE TREATMENT PARAMETERS USED AND CONFIRMED THAT THEY WERE WITHIN THE NORMAL APPROVED CLINICAL RANGE. THE PATIENT REPORTED THAT THE BURNING AND SCABBING WAS EXPERIENCED TWO DAYS POST TREATMENT AND LASTED FOR SEVEN DAYS. IT WAS REPORTED THAT THE PATIENT HAS FULLY RECOVERED. REVIEW OF THE IMAGES PROVIDED OF THE DEVICE TIP DID OBSERVE A SMALL BURN MARK ON THE EFFECTOR SURFACE. THE TIP HAS BEEN RETURNED TO THE MANUFACTURER FOR FURTHER EVALUATION. AT THIS TIME, IT IS UNKNOWN WHAT CONTRIBUTED TO THE REPORTED MALFUNCTION AND BURNING OBSERVED DAMAGE TO THE TIP OBSERVED IN THE PROVIDED IMAGE. IT IS UNKNOWN WHAT CAUSED THIS OBSERVED DEFECT TO THE TIP. THE DEFECT WAS OBSERVED BUT UNABLE TO BE VERIFIED IF IT CONTRIBUTED TO THE REPORTED BURNING. A FINAL MDR WILL BE SUBMITTED ONCE INVESTIGATION HAS BEEN COMPLETED. SINCE A REPORTED MALFUNCTION AND POTENTIAL FOR SERIOUS INJURY OCCURRED, WE ARE REPORTING THIS AS AN INITIAL MDR.
Patient problems: Burn(s)
MAUDE report 23933675 ↗MalfunctionSep 11, 2025Material Deformation; Insufficient Device Problem Information; Material Deformation; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING THE CONTROVERSIES EVENT IN (B)(6), A PATIENT RECEIVED BURNS DURING ELECTROCOAGULATION TREATMENT ON THE FACE WITH A XERF DEVICE. REVIEW OF THE TREATMENT PARAMETERS USED WERE FOUND TO BE WITHIN THE RECOMMENDED CLINICAL GUIDELINES. REVIEW OF THE SUPPLIED IMAGE OF THE TIP OBSERVED WHAT APPEARED TO BE A BURN MARK. IT IS UNKNOWN WHAT CAUSED THIS VISUAL DISCREPANCY. IMAGES OF THE PATIENT HAVE BEEN REQUESTED TO REVIEW WITH THE CYNOSURE LUTRONIC MEDICAL DIRECTOR, HOWEVER, DESPITE MULTIPLE ATTEMPTS TO OBTAIN THE REQUESTED IMAGES, NONE HAVE NOT BEEN SUPPLIED AT THIS TIME. AT THIS TIME, IT IS UNKNOWN WHAT CONTRIBUTED TO THE REPORTED MALFUNCTION AND BURNING OBSERVED DAMAGE TO THE TIP OBSERVED IN THE PROVIDED IMAGE. IT IS UNKNOWN WHAT CAUSED THIS OBSERVED DEFECT TO THE TIP. THE DEFECT WAS OBSERVED BUT UNABLE TO BE VERIFIED IF IT CONTRIBUTED TO THE REPORTED BURNING. A FINAL MDR WILL BE SUBMITTED ONCE INVESTIGATION HAS BEEN COMPLETED. SINCE A REPORTED MALFUNCTION AND SERIOUS INJURY OCCURRED, WE ARE REPORTING THIS AS AN INITIAL MDR.
Patient problems: Burn(s)
MAUDE report 23027970 ↗
Other Lutronic Corporation products
All Lutronic Corporation devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.