LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/… recallZ-0228-2026 · Synthes (USA) Products LLC · initiated Sep 26, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
FDA root cause category: Process control
What the company is doing
DePuy Synthes notified consignees via email and mailed letter on 09/26/2025. Consignees were instructed to examine inventory and quarantine affected units, arrange for the return of any affected units, complete and return the Business Response Form, and place the notification in a visible area for awareness. If units were further distributed within your facility or to customers, please notify those customers and personnel.
Recall details
- Recall number
- Z-0228-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 26, 2025
- Posted by the FDA
- Oct 21, 2025
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 120 units
- Product code
- JDS
- FDA submissions
- K201336, K201346
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
Product description
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Code information (models, lots, serial numbers)
Product Code: 04.045.346TS. GTIN: 10886982301413. Lot Number: 31756P7. Expiration Date: 08/31/2025
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA 04.045.346TSSynthes GmbH · Screw, fixation, bone1 recall0 reports
Other devices under the same FDA submission
166 devices cleared under K201336, K201346 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE234 reports
- NA 03.043.025Synthes GmbH · Screw, fixation, bone142 reports
- NA 03.233.006Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES71 reports
- NA 04.045.342TSSynthes GmbH · Screw, fixation, bone1 recall1 report
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- NA 04.043.240SSynthes GmbH · NAIL, FIXATION, BONE24 reports
- NA 03.233.005Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES23 reports
- NA 04.043.245SSynthes GmbH · NAIL, FIXATION, BONE19 reports
- NA 03.043.021Synthes GmbH · Screw, fixation, bone18 reports
- NA 04.045.080Synthes GmbH · NAIL, FIXATION, BONE17 reports
- NA 04.043.235SSynthes GmbH · Screw, fixation, bone15 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97693).
- Z-0229-2026 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.