Model: Incisive CT for Brazil SKD recallZ-0377-2026 · Philips Medical Systems · initiated Sep 25, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
FDA root cause category: Under Investigation by firm
What the company is doing
On September 25, 2025, ELECTRONIC PRODUCT RADIATION CORRECTION/URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken. - Follow the instructions within the letter on the use of the systems with updated recommendations. Complete and return the provided response form. Circulate the notice to all users of the device, retain the letter and post it in a visible place likely to be seen/viewed. Actions by recalling firm: Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve the issue (reference FCO72800841). Philips will, without charge bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. On September 29, 2025 updated ELECTRONIC PRODUCT RADIATIONCORRECTION/URGENT MEDICAL DEVICE CORRECTION letters were sent to customers to notify of incorrect UDI numbers provided in the initial letter.
Recall details
- Recall number
- Z-0377-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 25, 2025
- Posted by the FDA
- Oct 27, 2025
- Recalling firm
- Philips Medical Systems
- Quantity
- 1 unit (OUS only)
- Product code
- JAK
- FDA submissions
- K180015, K212441, K232491
- Distribution
- Domestic: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT, VA; International: Argentina, Australia, Austria, Brazil, Czeck Republic, Denmark, Ecuador, Finland, France, Germany, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kenya, Latvia, Lebanon, Libya, Netherlands, Nicaragua, Norway, Panama, Philippines, Poland, Portugal, Reunion, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, U.A.E., Yemen;
Product description
Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;
Code information (models, lots, serial numbers)
Product Code (REF): 728146; UDI-DI: None; Serial Number: 540000
Other devices under the same FDA submission
1 device cleared under K180015, K212441, K232491 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97699).
- Z-0375-2026 · Open, ClassifiedModel: Incisive CT; Product Code (REF): (1) 728143, (2) 728144; Software Versions: 5.1.0.X & 5.1.1.X;
- Z-0376-2026 · Open, ClassifiedModel: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.