MEDLINE convenience kits labeled as recallZ-0524-2026 · Medline Industries, LP · initiated Sep 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, please complete and submit the form. 3. Upon receipt of the submitted response form, the customer account will receive over-labels to place on affected inventory, with instructions for the customer staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. For distributors or those who have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0524-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2025
- Posted by the FDA
- Nov 14, 2025
- Recalling firm
- Medline Industries, LP
- Quantity
- 2844 kits
- Product code
- OHD
- Distribution
- US Nationwide distribution.
Product description
MEDLINE convenience kits labeled as: 1) GYN LAPAROSCOPY CDS #34-RF, REF CDS920099S; 2) GYN LAPAROSCOPY CDS-LF, REF CDS983764J; 3) ROBOTIC, REF CDS984262N; 4) ROBOTIC, REF CDS984262P; 5) ROBOTIC UROLOGY-RF, REF DYNJ38844N; 6) DA VINCI PROSTATE/COLPOPEXY, REF DYNJ44864S; 7) LAP-GYN PACK, REF DYNJ44876M; 8) ROBOT PACK, REF DYNJ45184F; 9) GYN LAP PACK, REF DYNJ52524D; 10) GYN LAP PACK, REF DYNJ64234C; 11) GYN DAVINCI PACK, REF DYNJ65373A; 12) ROBOT PACK, REF DYNJ67190C; 13) ROBOTIC PACK, REF DYNJ86154; 14) GYN LAPAROTOMY, REF DYNJ900296K; 15) ROBOT THOR, REF DYNJ900300K; 16) ROBOT URO, REF DYNJ900301K; 17) SM DAVINCI, REF DYNJ903539G; 18) ROBOTIC PACK-LF, REF DYNJ904250I; 19) ROBOTIC-LF, REF DYNJ907910B; 20) GP-ROBOTIC PACK-LF, REF PHS504647004G.
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
REF CDS920099S: UDI/DI 10195327140656 (each) 40195327140657 (case), Lot Numbers: 23LBV707; REF CDS920099S: UDI/DI 10195327140656 (each) 40195327140657 (case), Lot Numbers: 24CBD429; REF CDS920099S: UDI/DI 10195327140656 (each) 40195327140657 (case), Lot Numbers: 24CBD498; REF CDS920099S: UDI/DI 10195327140656 (each) 40195327140657 (case), Lot Numbers: 24DBK836; REF CDS920099S: UDI/DI 10195327140656 (each) 40195327140657 (case), Lot Numbers: 24HBE577; REF CDS983764J: UDI/DI 10195327451219 (each) 40195327451210 (case), Lot Numbers: 23HBO463; REF CDS983764J: UDI/DI 10195327451219 (each) 40195327451210 (case), Lot Numbers: 23JBF467; REF CDS984262N: UDI/DI 10195327241735 (each) 40195327241736 (case), Lot Numbers: 24ABP835; REF CDS984262P: UDI/DI 10195327647704 (each) 40195327647705 (case), Lot Numbers: 24CBK422; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 23GBT512; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 23HBI619; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 23LBN095; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 24BBR322; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 24DBA027; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 24DBT103; REF DYNJ38844N: UDI/DI 10195327243166 (each) 40195327243167 (case), Lot Numbers: 24EBG556; REF DYNJ44864S: UDI/DI 10195327127626 (each) 40195327127627 (case), Lot Numbers: 23GBO806; REF DYNJ44864S: UDI/DI 10195327127626 (each) 40195327127627 (case), Lot Numbers: 23HBJ772; REF DYNJ44876M: UDI/DI 10195327298418 (each) 40195327298419 (case), Lot Numbers: 23EDB052; REF DYNJ44876M: UDI/DI 10195327298418 (each) 40195327298419 (case), Lot Numbers: 23GDB708; REF DYNJ44876M: UDI/DI 10195327298418 (each) 40195327298419 (case), Lot Numbers: 23GDB759; REF DYNJ45184F: UDI/DI 10195327191474 (each) 40195327191475 (case), Lot Numbers: 23DBJ363; REF DYNJ45184F: UDI/DI 10195327191474 (each) 40195327191475 (case), Lot Numbers: 23EBU559; REF DYNJ45184F: UDI/DI 10195327191474 (each) 40195327191475 (case), Lot Numbers: 23HBL687; REF DYNJ45184F: UDI/DI 10195327191474 (each) 40195327191475 (case), Lot Numbers: 23JBD927; REF DYNJ52524D: UDI/DI 10195327561956 (each) 40195327561957 (case), Lot Numbers: 23LBG811; REF DYNJ52524D: UDI/DI 10195327561956 (each) 40195327561957 (case), Lot Numbers: 24BBQ970; REF DYNJ64234C: UDI/DI 10195327486921 (each) 40195327486922 (case), Lot Numbers: 23JMA672; REF DYNJ65373A: UDI/DI 10195327417987 (each) 40195327417988 (case), Lot Numbers: 23EMH332; REF DYNJ65373A: UDI/DI 10195327417987 (each) 40195327417988 (case), Lot Numbers: 23FMD177; REF DYNJ65373A: UDI/DI 10195327417987 (each) 40195327417988 (case), Lot Numbers: 23HMB194; REF DYNJ67190C: UDI/DI 10195327304201 (each) 40195327304202 (case), Lot Numbers: 23DBJ145; REF DYNJ67190C: UDI/DI 10195327304201 (each) 40195327304202 (case), Lot Numbers: 23DBK217; REF DYNJ67190C: UDI/DI 10195327304201 (each) 40195327304202 (case), Lot Numbers: 23FBS005; REF DYNJ67190C: UDI/DI 10195327304201 (each) 40195327304202 (case), Lot Numbers: 23GBP632; REF DYNJ67190C: UDI/DI 10195327304201 (each) 40195327304202 (case), Lot Numbers: 23IBP551; REF DYNJ67190C: UDI/DI 10195327304201 (each) 40195327304202 (case), Lot Numbers: 23JBH964; REF DYNJ86154: UDI/DI 10195327598112 (each) 40195327598113 (case), Lot Numbers: 24CBA421; REF DYNJ900296K: UDI/DI 10195327554088 (each) 40195327554089 (case), Lot Numbers: 23LBG402; REF DYNJ900300K: UDI/DI 10195327554194 (each) 40195327554195 (case), Lot Numbers: 23LBF280; REF DYNJ900300K: UDI/DI 10195327554194 (each) 40195327554195 (case), Lot Numbers: 23LBO223; REF DYNJ900300K: UDI/DI 10195327554194 (each) 40195327554195 (case), Lot Numbers: 24ABV771; REF DYNJ900300K: UDI/DI 10195327554194 (each) 40195327554195 (case), Lot Numbers: 24CBC856; REF DYNJ900301K: UDI/DI 10195327554200 (each) 40195327554201 (case), Lot Numbers: 24ABS777; REF DYNJ900301K: UDI/DI 10195327554200 (each) 40195327554201 (case), Lot Numbers: 24BBB142; REF DYNJ900301K: UDI/DI 10195327554200 (each) 40195327554201 (case), Lot Numbers: 24CBI705; REF DYNJ900301K: UDI/DI 10195327554200 (each) 40195327554201 (case), Lot Numbers: 24EBN319; REF DYNJ903539G: UDI/DI 10195327425524 (each) 40195327425525 (case), Lot Numbers: 23JMC169; REF DYNJ903539G: UDI/DI 10195327425524 (each) 40195327425525 (case), Lot Numbers: 23JMC442; REF DYNJ903539G: UDI/DI 10195327425524 (each) 40195327425525 (case), Lot Numbers: 23JMG251; REF DYNJ903539G: UDI/DI 10195327425524 (each) 40195327425525 (case), Lot Numbers: 23KMC642; REF DYNJ903539G: UDI/DI 10195327425524 (each) 40195327425525 (case), Lot Numbers: 23KME178; REF DYNJ903539G: UDI/DI 10195327425524 (each) 40195327425525 (case), Lot Numbers: 23LMC669; REF DYNJ904250I: UDI/DI 10195327420826 (each) 40195327420827 (case), Lot Numbers: 24ABT614; REF DYNJ904250I: UDI/DI 10195327420826 (each) 40195327420827 (case), Lot Numbers: 24DBM183; REF DYNJ904250I: UDI/DI 10195327420826 (each) 40195327420827 (case), Lot Numbers: 24EBJ705; REF DYNJ907910B: UDI/DI 10195327445096 (each) 40195327445097 (case), Lot Numbers: 24ABC736; REF DYNJ907910B: UDI/DI 10195327445096 (each) 40195327445097 (case), Lot Numbers: 24DBD770; REF DYNJ907910B: UDI/DI 10195327445096 (each) 40195327445097 (case), Lot Numbers: 24EBH118; REF PHS504647004G: UDI/DI 10195327454142 (each) 40195327454143 (case), Lot Numbers: 23HDB026.
Devices named in this recall
20 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ44864SMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. DYNJ38844NMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. CDS920099SMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ45184FMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. CDS984262NMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ44876MMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ67190CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ65373AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ903539GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ904250IMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907910BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS983764JMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. PHS504647004GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ64234CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ900296KMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ900300KMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ900301KMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ52524DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ86154Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS984262PMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97846).
- Z-0511-2026 · Open, ClassifiedMEDLINE convenience kits labeled as LOCAL BASIC, REF DYNJ34418M
- Z-0512-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: CATH LAB PACEMAKER PACK, REF DYNJ59818A
- Z-0513-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GENERAL AAA #11-RF, REF CDS840261AB; 2) CAROTID ENDARTERECTOMY CDS-LF, REF CDS983637I; 3) PERIPHERAL VASCULAR CDS, REF CDS983723J; 4) A…
- Z-0514-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) CENTRAL LINE PACK, REF DYNJ00281P; 2) FREE FLAP PACK, REF DYNJ68213C.
- Z-0515-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) BREAST BIOPSY PACK-LF, REF DYNJ0160598C; 2) BIOPSY BREAST PACK, REF DYNJ65361A; 3) BREAST BIOPSY/MASTECTOMY, REF DYNJ908839A.
- Z-0516-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) HEAD AND NECK, REF CDS984259O; 2) HEAD AND NECK, REF CDS984259P; 3) MINOR ENT HARPER PACK-LF, REF DYNJ0160700F; 4) MAJOR ENT HARPER PAC…
- Z-0517-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MAJOR LAPAROTOMY CDS, REF CDS860015W; 2) MAJOR LAPAROTOMY CDS, REF CDS860015X; 3) MINOR LAPAROTOMY CDS, REF CDS860016R; 4) MINOR LAPARO…
- Z-0518-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597
- Z-0519-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) ACH PLASTICS BREAST-LF, REF CDS983639K; 2) BREAST ABD, REF CDS984267J; 3) PLASTIC PACK, REF DYNJ00264M; 4) BREAST ABDOMINOPLASTY MZ PK-…
- Z-0520-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MINOR SINGLE BASIN PACK, REF DYNJ37698L; 2) BASIN PACK, REF DYNJ50123D; 3) MAJOR BASIN PACK, REF DYNJ59816A; 4) OSC BASIN PACK, REF DYN…
- Z-0521-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GENERAL LAPAROSCOPY, REF CDS930027W; 2) GENERAL MAJOR CDS, REF CDS980245X; 3) BASIC GENERAL CDS, REF CDS980566F; 4) TOTAL KNEE CDS, REF…
- Z-0522-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, RE…
- Z-0523-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MAJOR VAGINAL HARPER PACK-LF, REF DYNJ0161768G; 2) ACH MAJOR VAGINAL PACK, REF DYNJ67757B.
- Z-0525-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: CIRCUMCISION PACK, REF DYNJ16826O
- Z-0526-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) C-SECTION OR 30 CDS, REF CDS820075W; 2) C-SECTION, REF CDS984261J; 3) C-SECTION CDS, REF CDS985282F; 4) C-SECTION PACK, REF DYNJ02577D;…
- Z-0527-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F; 2) MAJOR LITHOTOMY PACK-LF, REF DYNJ0425778O; 3) MINOR VAGINAL #76-RF, REF DYNJ27434R; …
- Z-0528-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: GP-LEEP PACK-LF, REF PHS396969004F
- Z-0529-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MINOR EYE TRAY #81, REF DYNJ27466S; 2) MINOR EYE TRAY #81, REF DYNJ27466T.
- Z-0530-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, REF CDS780119V; 2) BASIC BACK CDS, REF CDS780147N; 3) INSTRUMENTED BACK CDS, REF CDS780148R; 4) NEURO-LAMI…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page