MEDLINE convenience kits labeled as recallZ-0530-2026 · Medline Industries, LP · initiated Sep 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, please complete and submit the form. 3. Upon receipt of the submitted response form, the customer account will receive over-labels to place on affected inventory, with instructions for the customer staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. For distributors or those who have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0530-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2025
- Posted by the FDA
- Nov 14, 2025
- Recalling firm
- Medline Industries, LP
- Quantity
- 40151 kits
- Product code
- OJH
- Distribution
- US Nationwide distribution.
Product description
MEDLINE convenience kits labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, REF CDS780119V; 2) BASIC BACK CDS, REF CDS780147N; 3) INSTRUMENTED BACK CDS, REF CDS780148R; 4) NEURO-LAMI CDS, REF CDS780208I; 5) TOTAL HIP, REF CDS920027Y; 6) TOTAL KNEE, REF CDS940072Y; 7) TOTAL KNEE REPLACEMENT CDS, REF CDS940823AB; 8) TOTAL HIP CDS, REF CDS940843AA; 9) TOTAL HIP CDS, REF CDS940963Y; 10) SHOULDER CDS, REF CDS941065L; 11) LAMINECTOMY CDS-LF, REF CDS981923X; 12) TOTAL KNEE CDS, REF CDS982823S; 13) LOWER EXTREMITY CDS, REF CDS982943S; 14) LOWER EXTREMITY CDS, REF CDS982943T; 15) TOTAL KNEE, REF CDS982986U; 16) EXTREMITY CDS, REF CDS983031Q; 17) PODIATRY CDS, REF CDS983610L; 18) TOTAL HIP CDS-LF, REF CDS983761N; 19) BASIC ORTHO CDS, REF CDS983769L; 20) UPPER EXTREMITY CDS, REF CDS983907L; 21) LOWER EXTREMITY CDS, REF CDS983909P; 22) GENERAL ORTHO CDS, REF CDS983915M; 23) MAJOR ORTHO, REF CDS984253L; 24) LOWER EXTREMTIY, REF CDS984254M; 25) TOTAL KNEE, REF CDS985566R; 26) ORTHO PACK, REF DYNJ00284M; 27) CRANIOTOMY PACK-LF, REF DYNJ0160354C; 28) CRANIOTOMY PACK-LF, REF DYNJ0161342C; 29) CRANIOTOMY PACK-LF, REF DYNJ0161342D; 30) CRANIOTOMY HARPER PACK-LF, REF DYNJ0162022G; 31) SPINE PACK-LF, REF DYNJ0376635P; 32) GP-TOTAL JOINT -LF, REF DYNJ0396868V; 33) GP-EXTREMITY/SHOULDER PACK-LF, REF DYNJ0396906P; 34) EXTREMITY PACK-LF, REF DYNJ0519273N; 35) TOTAL KNEE PACK-LF, REF DYNJ0530895AO; 36) TOTAL KNEE PACK-LF, REF DYNJ0530895AP; 37) UPPER EXTREMITY PACK-LF, REF DYNJ0532846S; 38) LOWER EXTREMITY PACK-LF, REF DYNJ0532852Q; 39) ACL SSC PACK-LF, REF DYNJ0536892U; 40) TOTAL HIP PACK-LF, REF DYNJ0536903AL; 41) TOTAL HIP PACK-LF, REF DYNJ0536903AM; 42) LOWER EXTREMITY PACK-LF, REF DYNJ0587115R; 43) TOTAL HIP PACK-LF, REF DYNJ0618354V; 44) TOTAL KNEE PACK-LF, REF DYNJ0646272S; 45) SPINE ANT POST ADD A PACK-LF, REF DYNJ0753208J; 46) OPEN SHOULDER SSC PACK, REF DYNJ0840659T; 47) TOTAL JOINT PACK-LF, REF DYNJ0885263F; 48) TOTAL KNEE PACK, REF DYNJ21813J; 49) CRANIO/MAXILOFACIAL #49-RF, REF DYNJ21826R; 50) TOTAL HIP PACK, REF DYNJ27315K; 51) PODIATRY PACK, REF DYNJ33978K; 52) LOWER EXTREMITY, REF DYNJ34420N; 53) LOWER EXTREMITY, REF DYNJ34420O; 54) GENERAL ORTHO, REF DYNJ34428O; 55) ARTHROSCOPY PACK, REF DYNJ35835G; 56) HAND PACK, REF DYNJ37697J; 57) ACH SPINE PACK, REF DYNJ38042K; 58) SHOULDER ARTHROSCOPY PACK, REF DYNJ38224C; 59) LAMINECTOMY PACK, REF DYNJ38227C; 60) EXTREMITY PACK, REF DYNJ38234G; 61) TOTAL JOINT PACK, REF DYNJ40982D; 62) SPINE PACK, REF DYNJ40995B; 63) EXTREMITY PACK, REF DYNJ41416G; 64) TOTAL JOINT PACK, REF DYNJ42793D; 65) CRANIOTOMY PACK-LF, REF DYNJ43208G; 66) HIP I PACK-LF, REF DYNJ43215I; 67) LAMINECTOMY PACK-LF, REF DYNJ43218G; 68) LIMB PACK-LF, REF DYNJ43220G; 69) SPLIT PACK-LF, REF DYNJ43225G; 70) TOTAL KNEE PACK-LF, REF DYNJ43226I; 71) EXTREMITY PACK, REF DYNJ44681J; 72) SHOULDER PACK, REF DYNJ44685L; 73) TOTAL KNEE PACK, REF DYNJ44687L; 74) UPPER EXTREMITY PACK, REF DYNJ44688I; 75) PODIATRY PACK, REF DYNJ44692G; 76) ACL PACK, REF DYNJ44849F; 77) CENTRACARE PLAZA-MINOR ORTHO, REF DYNJ44860I; 78) MAJOR EXTREMITY, REF DYNJ44878D; 79) ORTHO HIP, REF DYNJ44884G; 80) EXTREMITY PACK, REF DYNJ45332G;
Code information (models, lots, serial numbers)
Show the full code information (67k characters)
REF CDS780119V: UDI/DI 10195327114916 (each) 40195327114917 (case), Lot Numbers: 23GBQ904; REF CDS780119V: UDI/DI 10195327114916 (each) 40195327114917 (case), Lot Numbers: 23JBS979; REF CDS780147N: UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers: 23LBD720; REF CDS780147N: UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers: 24BBK043; REF CDS780147N: UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers: 24CBA347; REF CDS780147N: UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers: 24EBM283; REF CDS780147N: UDI/DI 10195327268770 (each) 40195327268771 (case), Lot Numbers: 24GBB624; REF CDS780148R: UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 23GBS348; REF CDS780148R: UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 23HBJ325; REF CDS780148R: UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 24ABK205; REF CDS780148R: UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 24ABW098; REF CDS780148R: UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 24BBN591; REF CDS780148R: UDI/DI 10195327268787 (each) 40195327268788 (case), Lot Numbers: 24CBS634; REF CDS780208I: UDI/DI 10195327525231 (each) 40195327525232 (case), Lot Numbers: 23JBU709; REF CDS920027Y: UDI/DI 10195327298531 (each) 40195327298532 (case), Lot Numbers: 23LBF304; REF CDS920027Y: UDI/DI 10195327298531 (each) 40195327298532 (case), Lot Numbers: 24ABB904; REF CDS920027Y: UDI/DI 10195327298531 (each) 40195327298532 (case), Lot Numbers: 24CBA311; REF CDS940072Y: UDI/DI 10195327300845 (each) 40195327300846 (case), Lot Numbers: 23KBU712; REF CDS940072Y: UDI/DI 10195327300845 (each) 40195327300846 (case), Lot Numbers: 24ABY634; REF CDS940823AB: UDI/DI 10193489883282 (each) 40193489883283 (case), Lot Numbers: 24ABQ717; REF CDS940823AB: UDI/DI 10193489883282 (each) 40193489883283 (case), Lot Numbers: 24ABU281; REF CDS940823AB: UDI/DI 10193489883282 (each) 40193489883283 (case), Lot Numbers: 24CBD739; REF CDS940823AB: UDI/DI 10193489883282 (each) 40193489883283 (case), Lot Numbers: 24DBP226; REF CDS940843AA: UDI/DI 10195327151409 (each) 40195327151400 (case), Lot Numbers: 24DBD229; REF CDS940843AA: UDI/DI 10195327151409 (each) 40195327151400 (case), Lot Numbers: 24EBM262; REF CDS940843AA: UDI/DI 10195327151409 (each) 40195327151400 (case), Lot Numbers: 24EBV396; REF CDS940963Y: UDI/DI 10193489883671 (each) 40193489883672 (case), Lot Numbers: 24BBO653; REF CDS940963Y: UDI/DI 10193489883671 (each) 40193489883672 (case), Lot Numbers: 24BBV236; REF CDS940963Y: UDI/DI 10193489883671 (each) 40193489883672 (case), Lot Numbers: 24CBS637; REF CDS940963Y: UDI/DI 10193489883671 (each) 40193489883672 (case), Lot Numbers: 24IBD594; REF CDS941065L: UDI/DI 10193489445350 (each) 40193489445351 (case), Lot Numbers: 23GBJ674; REF CDS941065L: UDI/DI 10193489445350 (each) 40193489445351 (case), Lot Numbers: 24ABP205; REF CDS941065L: UDI/DI 10193489445350 (each) 40193489445351 (case), Lot Numbers: 24CBD703; REF CDS941065L: UDI/DI 10193489445350 (each) 40193489445351 (case), Lot Numbers: 24EBT843; REF CDS981923X: UDI/DI 10195327434298 (each) 40195327434299 (case), Lot Numbers: 23HBB366; REF CDS981923X: UDI/DI 10195327434298 (each) 40195327434299 (case), Lot Numbers: 23HBS882; REF CDS982823S: UDI/DI 10195327674403 (each) 40195327674404 (case), Lot Numbers: 24EBL031; REF CDS982943S: UDI/DI 10195327432393 (each) 40195327432394 (case), Lot Numbers: 23HBA580; REF CDS982943S: UDI/DI 10195327432393 (each) 40195327432394 (case), Lot Numbers: 23HBN648; REF CDS982943T: UDI/DI 10195327627003 (each) 40195327627004 (case), Lot Numbers: 24DBW083; REF CDS982943T: UDI/DI 10195327627003 (each) 40195327627004 (case), Lot Numbers: 24GBC807; REF CDS982986U: UDI/DI 10195327554095 (each) 40195327554096 (case), Lot Numbers: 23LBJ387; REF CDS982986U: UDI/DI 10195327554095 (each) 40195327554096 (case), Lot Numbers: 23LBQ432; REF CDS982986U: UDI/DI 10195327554095 (each) 40195327554096 (case), Lot Numbers: 24ABP570; REF CDS982986U: UDI/DI 10195327554095 (each) 40195327554096 (case), Lot Numbers: 24CBI682; REF CDS982986U: UDI/DI 10195327554095 (each) 40195327554096 (case), Lot Numbers: 24EBC693; REF CDS983031Q: UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 23IBW098; REF CDS983031Q: UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 24BBB180; REF CDS983031Q: UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 24BBU600; REF CDS983031Q: UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 24CBN838; REF CDS983031Q: UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 24EBT875; REF CDS983031Q: UDI/DI 10193489848359 (each) 40193489848350 (case), Lot Numbers: 24FBS402; REF CDS983610L: UDI/DI 10195327210151 (each) 40195327210152 (case), Lot Numbers: 23HBC818; REF CDS983610L: UDI/DI 10195327210151 (each) 40195327210152 (case), Lot Numbers: 24DBE385; REF CDS983761N: UDI/DI 10195327473662 (each) 40195327473663 (case), Lot Numbers: 23JBC403; REF CDS983761N: UDI/DI 10195327473662 (each) 40195327473663 (case), Lot Numbers: 23JBM156; REF CDS983761N: UDI/DI 10195327473662 (each) 40195327473663 (case), Lot Numbers: 23JBU065; REF CDS983761N: UDI/DI 10195327473662 (each) 40195327473663 (case), Lot Numbers: 23LBD278; REF CDS983769L: UDI/DI 10195327451196 (each) 40195327451197 (case), Lot Numbers: 23HBP622; REF CDS983769L: UDI/DI 10195327451196 (each) 40195327451197 (case), Lot Numbers: 23JBC159; REF CDS983769L: UDI/DI 10195327451196 (each) 40195327451197 (case), Lot Numbers: 23KBP622; REF CDS983907L: UDI/DI 10195327238346 (each) 40195327238347 (case), Lot Numbers: 23IBC640; REF CDS983907L: UDI/DI 10195327238346 (each) 40195327238347 (case), Lot Numbers: 23LBM312; REF CDS983909P: UDI/DI 10195327238384 (each) 40195327238385 (case), Lot Numbers: 23IBW039; REF CDS983909P: UDI/DI 10195327238384 (each) 40195327238385 (case), Lot Numbers: 23JBB094; REF CDS983909P: UDI/DI 10195327238384 (each) 40195327238385 (case), Lot Numbers: 23LBK752; REF CDS983915M: UDI/DI 10195327238353 (each) 40195327238354 (case), Lot Numbers: 24ABG846; REF CDS984253L: UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 24ABD711; REF CDS984253L: UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 24ABP282; REF CDS984253L: UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 24BBI150; REF CDS984253L: UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 24CBS643; REF CDS984253L: UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 24DBK828; REF CDS984253L: UDI/DI 10193489288513 (each) 40193489288514 (case), Lot Numbers: 24EBP547; REF CDS984254M: UDI/DI 10195327415242 (each) 40195327415243 (case), Lot Numbers: 23HBD301; REF CDS984254M: UDI/DI 10195327415242 (each) 40195327415243 (case), Lot Numbers: 24ABT691; REF CDS984254M: UDI/DI 10195327415242 (each) 40195327415243 (case), Lot Numbers: 24BBS781; REF CDS984254M: UDI/DI 10195327415242 (each) 40195327415243 (case), Lot Numbers: 24DBS099; REF CDS984254M: UDI/DI 10195327415242 (each) 40195327415243 (case), Lot Numbers: 24EBF862; REF CDS985566R: UDI/DI 10198459014758 (each) 40198459014759 (case), Lot Numbers: 24GBH839; REF DYNJ00284M: UDI/DI 10195327412036 (each) 40195327412037 (case), Lot Numbers: 23GMA018; REF DYNJ00284M: UDI/DI 10195327412036 (each) 40195327412037 (case), Lot Numbers: 24AMC690; REF DYNJ0160354C: UDI/DI 10195327320072 (each) 40195327320073 (case), Lot Numbers: 23GDA315; REF DYNJ0160354C: UDI/DI 10195327320072 (each) 40195327320073 (case), Lot Numbers: 23GDB867; REF DYNJ0161342C: UDI/DI 10195327357269 (each) 40195327357260 (case), Lot Numbers: 23EDA760; REF DYNJ0161342D: UDI/DI 10195327478506 (each) 40195327478507 (case), Lot Numbers: 23HDB481; REF DYNJ0162022G: UDI/DI 10195327404598 (each) 40195327404599 (case), Lot Numbers: 23FDB286; REF DYNJ0162022G: UDI/DI 10195327404598 (each) 40195327404599 (case), Lot Numbers: 23GDA631; REF DYNJ0162022G: UDI/DI 10195327404598 (each) 40195327404599 (case), Lot Numbers: 23HDA239; REF DYNJ0376635P: UDI/DI 10195327532987 (each) 40195327532988 (case), Lot Numbers: 24AMB890; REF DYNJ0396868V: UDI/DI 10195327454258 (each) 40195327454259 (case), Lot Numbers: 23GDB690; REF DYNJ0396906P: UDI/DI 10195327454197 (each) 40195327454198 (case), Lot Numbers: 23HDB063; REF DYNJ0396906P: UDI/DI 10195327454197 (each) 40195327454198 (case), Lot Numbers: 23JDB511; REF DYNJ0519273N: UDI/DI 10195327083472 (each) 40195327083473 (case), Lot Numbers: 23EMA335; REF DYNJ0530895AO: UDI/DI 10193489782790 (each) 40193489782791 (case), Lot Numbers: 24ABS237; REF DYNJ0530895AP: UDI/DI 10195327628550 (each) 40195327628551 (case), Lot Numbers: 24CBJ626; REF DYNJ0530895AP: UDI/DI 10195327628550 (each) 40195327628551 (case), Lot Numbers: 24DBH846; REF DYNJ0532846S: UDI/DI 10195327356781 (each) 40195327356782 (case), Lot Numbers: 23JME365; REF DYNJ0532846S: UDI/DI 10195327356781 (each) 40195327356782 (case), Lot Numbers: 23KMF703; REF DYNJ0532846S: UDI/DI 10195327356781 (each) 40195327356782 (case), Lot Numbers: 23KMJ025; REF DYNJ0532846S: UDI/DI 10195327356781 (each) 40195327356782 (case), Lot Numbers: 24AMI561; REF DYNJ0532852Q: UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers: 23JMC960; REF DYNJ0532852Q: UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers: 23LMA587; REF DYNJ0532852Q: UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers: 24BMJ693; REF DYNJ0532852Q: UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers: 24BMJ826; REF DYNJ0532852Q: UDI/DI 10195327356705 (each) 40195327356706 (case), Lot Numbers: 24DMK245; REF DYNJ0536892U: UDI/DI 10195327206239 (each) 40195327206230 (case), Lot Numbers: 23HBH153; REF DYNJ0536892U: UDI/DI 10195327206239 (each) 40195327206230 (case), Lot Numbers: 24DBH427; REF DYNJ0536892U: UDI/DI 10195327206239 (each) 40195327206230 (case), Lot Numbers: 24DBI493; REF DYNJ0536903AL: UDI/DI 10193489782783 (each) 40193489782784 (case), Lot Numbers: 24ABK586; REF DYNJ0536903AM: UDI/DI 10195327628567 (each) 40195327628568 (case), Lot Numbers: 24DBA383; REF DYNJ0536903AM: UDI/DI 10195327628567 (each) 40195327628568 (case), Lot Numbers: 24JBS809; REF DYNJ0587115R: UDI/DI 10195327561888 (each) 40195327561889 (case), Lot Numbers: 24BBP114; REF DYNJ0587115R: UDI/DI 10195327561888 (each) 40195327561889 (case), Lot Numbers: 24DBO138; REF DYNJ0618354V: UDI/DI 10889942860760 (each) 40889942860761 (case), Lot Numbers: 23GBG269; REF DYNJ0618354V: UDI/DI 10889942860760 (each) 40889942860761 (case), Lot Numbers: 23GMC905; REF DYNJ0618354V: UDI/DI 10889942860760 (each) 40889942860761 (case), Lot Numbers: 24ABH235; REF DYNJ0618354V: UDI/DI 10889942860760 (each) 40889942860761 (case), Lot Numbers: 24CBR569; REF DYNJ0646272S: UDI/DI 10889942860920 (each) 40889942860921 (case), Lot Numbers: 23KBU390; REF DYNJ0646272S: UDI/DI 10889942860920 (each) 40889942860921 (case), Lot Numbers: 24CBU458; REF DYNJ0753208J: UDI/DI 10195327689896 (each) 40195327689897 (case), Lot Numbers: 24FME095; REF DYNJ0840659T: UDI/DI 10195327206512 (each) 40195327206513 (case), Lot Numbers: 23LBS322; REF DYNJ0885263F: UDI/DI 10195327219635 (each) 40195327219636 (case), Lot Numbers: 23JMG075; REF DYNJ0885263F: UDI/DI 10195327219635 (each) 40195327219636 (case), Lot Numbers: 23KMI286; REF DYNJ0885263F: UDI/DI 10195327219635 (each) 40195327219636 (case), Lot Numbers: 24AMH655; REF DYNJ0885263F: UDI/DI 10195327219635 (each) 40195327219636 (case), Lot Numbers: 24CMC882; REF DYNJ21813J: UDI/DI 10195327515867 (each) 40195327515868 (case), Lot Numbers: 23JBE704; REF DYNJ21813J: UDI/DI 10195327515867 (each) 40195327515868 (case), Lot Numbers: 24ABV081; REF DYNJ21826R: UDI/DI 10195327104412 (each) 40195327104413 (case), Lot Numbers: 23LBA701; REF DYNJ27315K: UDI/DI 10195327515966 (each) 40195327515967 (case), Lot Numbers: 23KBC303; REF DYNJ33978K: UDI/DI 10195327436315 (each) 40195327436316 (case), Lot Numbers: 23GBR447; REF DYNJ33978K: UDI/DI 10195327436315 (each) 40195327436316 (case), Lot Numbers: 23IBR613; REF DYNJ34420N: UDI/DI 10195327384388 (each) 40195327384389 (case), Lot Numbers: 23GBI363; REF DYNJ34420N: UDI/DI 10195327384388 (each) 40195327384389 (case), Lot Numbers: 23IBG432; REF DYNJ34420N: UDI/DI 10195327384388 (each) 40195327384389 (case), Lot Numbers: 23IBR127; REF DYNJ34420N: UDI/DI 10195327384388 (each) 40195327384389 (case), Lot Numbers: 23LBN020; REF DYNJ34420O: UDI/DI 10195327610616 (each) 40195327610617 (case), Lot Numbers: 24ABN426; REF DYNJ34420O: UDI/DI 10195327610616 (each) 40195327610617 (case), Lot Numbers: 24BBV753; REF DYNJ34420O: UDI/DI 10195327610616 (each) 40195327610617 (case), Lot Numbers: 24DBA214; REF DYNJ34420O: UDI/DI 10195327610616 (each) 40195327610617 (case), Lot Numbers: 24DBJ230; REF DYNJ34428O: UDI/DI 10195327384333 (each) 40195327384334 (case), Lot Numbers: 23KMF246; REF DYNJ34428O: UDI/DI 10195327384333 (each) 40195327384334 (case), Lot Numbers: 24AMI130; REF DYNJ34428O: UDI/DI 10195327384333 (each) 40195327384334 (case), Lot Numbers: 24DMF181; REF DYNJ35835G: UDI/DI 10195327412005 (each) 40195327412006 (case), Lot Numbers: 23GMI346; REF DYNJ35835G: UDI/DI 10195327412005 (each) 40195327412006 (case), Lot Numbers: 23IMF703; REF DYNJ37697J: UDI/DI 10195327448325 (each) 40195327448326 (case), Lot Numbers: 23JME360; REF DYNJ37697J: UDI/DI 10195327448325 (each) 40195327448326 (case), Lot Numbers: 23LMF291; REF DYNJ37697J: UDI/DI 10195327448325 (each) 40195327448326 (case), Lot Numbers: 24BME117; REF DYNJ37697J: UDI/DI 10195327448325 (each) 40195327448326 (case), Lot Numbers: 24EMF837; REF DYNJ38042K: UDI/DI 10195327416539 (each) 40195327416530 (case), Lot Numbers: 23GBC482; REF DYNJ38042K: UDI/DI 10195327416539 (each) 40195327416530 (case), Lot Numbers: 23HBU981; REF DYNJ38042K: UDI/DI 10195327416539 (each) 40195327416530 (case), Lot Numbers: 23KBF956; REF DYNJ38224C: UDI/DI 10195327381479 (each) 40195327381470 (case), Lot Numbers: 23DBJ365; REF DYNJ38224C: UDI/DI 10195327381479 (each) 40195327381470 (case), Lot Numbers: 23GBO385; REF DYNJ38224C: UDI/DI 10195327381479 (each) 40195327381470 (case), Lot Numbers: 23GBU542; REF DYNJ38224C: UDI/DI 10195327381479 (each) 40195327381470 (case), Lot Numbers: 23HBL905; REF DYNJ38227C: UDI/DI 10195327381486 (each) 40195327381487 (case), Lot Numbers: 23KBG323; REF DYNJ38234G: UDI/DI 10195327327651 (each) 40195327327652 (case), Lot Numbers: 23DBJ638; REF DYNJ38234G: UDI/DI 10195327327651 (each) 40195327327652 (case), Lot Numbers: 23EBR720; REF DYNJ40982D: UDI/DI 10889942564002 (each) 40889942564003 (case), Lot Numbers: 24ABA449; REF DYNJ40982D: UDI/DI 10889942564002 (each) 40889942564003 (case), Lot Numbers: 24BBG971; REF DYNJ40995B: UDI/DI 10889942431236 (each) 40889942431237 (case), Lot Numbers: 24BBS824; REF DYNJ41416G: UDI/DI 10195327220976 (each) 40195327220977 (case), Lot Numbers: 23GBA790; REF DYNJ41416G: UDI/DI 10195327220976 (each) 40195327220977 (case), Lot Numbers: 23HBE316; REF DYNJ42793D: UDI/DI 10193489427547 (each) 40193489427548 (case), Lot Numbers: 23IME206; REF DYNJ43208G: UDI/DI 10193489953053 (each) 40193489953054 (case), Lot Numbers: 23DMG563; REF DYNJ43208G: UDI/DI 10193489953053 (each) 40193489953054 (case), Lot Numbers: 23EME340; REF DYNJ43215I: UDI/DI 10193489953084 (each) 40193489953085 (case), Lot Numbers: 23JMA398; REF DYNJ43218G: UDI/DI 10195327421977 (each) 40195327421978 (case), Lot Numbers: 23HMC253; REF DYNJ43220G: UDI/DI 10193489953114 (each) 40193489953115 (case), Lot Numbers: 23GME630; REF DYNJ43225G: UDI/DI 10193489958874 (each) 40193489958875 (case), Lot Numbers: 23EMA404; REF DYNJ43226I: UDI/DI 10193489953039 (each) 40193489953030 (case), Lot Numbers: 23GMC914; REF DYNJ43226I: UDI/DI 10193489953039 (each) 40193489953030 (case), Lot Numbers: 23IME870; REF DYNJ43226I: UDI/DI 10193489953039 (each) 40193489953030 (case), Lot Numbers: 23IMF935; REF DYNJ44681J: UDI/DI 10195327130886 (each) 40195327130887 (case), Lot Numbers: 23EDA416; REF DYNJ44685L: UDI/DI 10195327127404 (each) 40195327127405 (case), Lot Numbers: 23EBT781; REF DYNJ44687L: UDI/DI 10195327127411 (each) 40195327127412 (case), Lot Numbers: 23JMA252; REF DYNJ44688I: UDI/DI 10195327127374 (each) 40195327127375 (case), Lot Numbers: 23HBA230; REF DYNJ44688I: UDI/DI 10195327127374 (each) 40195327127375 (case), Lot Numbers: 23JBJ203; REF DYNJ44692G: UDI/DI 10195327277574 (each) 40195327277575 (case), Lot Numbers: 23EDA887; REF DYNJ44692G: UDI/DI 10195327277574 (each) 40195327277575 (case), Lot Numbers: 23FDB510; REF DYNJ44692G: UDI/DI 10195327277574 (each) 40195327277575 (case), Lot Numbers: 23JDB517; REF DYNJ44849F: UDI/DI 10195327127572 (each) 40195327127573 (case), Lot Numbers: 23EDA066; REF DYNJ44849F: UDI/DI 10195327127572 (each) 40195327127573 (case), Lot Numbers: 23GDA271; REF DYNJ44849F: UDI/DI 10195327127572 (each) 40195327127573 (case), Lot Numbers: 23GDC025; REF DYNJ44860I: UDI/DI 10195327127343 (each) 40195327127344 (case), Lot Numbers: 23HDB166; REF DYNJ44878D: UDI/DI 10195327127701 (each) 40195327127702 (case), Lot Numbers: 23EDB339; REF DYNJ44878D: UDI/DI 10195327127701 (each) 40195327127702 (case), Lot Numbers: 23GDB902; REF DYNJ44884G: UDI/DI 10195327298425 (each) 40195327298426 (case), Lot Numbers: 23EDB345; REF DYNJ44884G: UDI/DI 10195327298425 (each) 40195327298426 (case), Lot Numbers: 23FDB539; REF DYNJ45332G: UDI/DI 10195327242534 (each) 40195327242535 (case), Lot Numbers: 23GMI355; REF DYNJ46676F: UDI/DI 10193489887617 (each) 40193489887618 (case), Lot Numbers: 23JDB444; REF DYNJ46676F: UDI/DI 10193489887617 (each) 40193489887618 (case), Lot Numbers: 23JDB935; REF DYNJ46676F: UDI/DI 10193489887617 (each) 40193489887618 (case), Lot Numbers: 23KDA590; REF DYNJ46676F: UDI/DI 10193489887617 (each) 40193489887618 (case), Lot Numbers: 24ADA920; REF DYNJ46698G: UDI/DI 10195327224066 (each) 40195327224067 (case), Lot Numbers: 23DDB946; REF DYNJ46698G: UDI/DI 10195327224066 (each) 40195327224067 (case), Lot Numbers: 23FDA276; REF DYNJ46698G: UDI/DI 10195327224066 (each) 40195327224067 (case), Lot Numbers: 23GDA719; REF DYNJ46931D: UDI/DI 10195327515638 (each) 40195327515639 (case), Lot Numbers: 23JBJ349; REF DYNJ47698P: UDI/DI 10195327437084 (each) 40195327437085 (case), Lot Numbers: 23GME940; REF DYNJ47698Q: UDI/DI 10195327509422 (each) 40195327509423 (case), Lot Numbers: 23JMC147; REF DYNJ47698Q: UDI/DI 10195327509422 (each) 40195327509423 (case), Lot Numbers: 23JME063; REF DYNJ47698Q: UDI/DI 10195327509422 (each) 40195327509423 (case), Lot Numbers: 23KMG557; REF DYNJ47698Q: UDI/DI 10195327509422 (each) 40195327509423 (case), Lot Numbers: 23LMH560; REF DYNJ47698Q: UDI/DI 10195327509422 (each) 40195327509423 (case), Lot Numbers: 24BME277; REF DYNJ47701K: UDI/DI 10195327509118 (each) 40195327509119 (case), Lot Numbers: 23IBU539; REF DYNJ47701K: UDI/DI 10195327509118 (each) 40195327509119 (case), Lot Numbers: 23LBE644; REF DYNJ47701K: UDI/DI 10195327509118 (each) 40195327509119 (case), Lot Numbers: 24CBE391; REF DYNJ47701K: UDI/DI 10195327509118 (each) 40195327509119 (case), Lot Numbers: 24DBU536; REF DYNJ48090I: UDI/DI 10195327500115 (each) 40195327500116 (case), Lot Numbers: 23LMC258; REF DYNJ48090I: UDI/DI 10195327500115 (each) 40195327500116 (case), Lot Numbers: 25BMA374; REF DYNJ48091I: UDI/DI 10195327500122 (each) 40195327500123 (case), Lot Numbers: 23IMG184; REF DYNJ48091I: UDI/DI 10195327500122 (each) 40195327500123 (case), Lot Numbers: 24DMG049; REF DYNJ48130K: UDI/DI 10195327349448 (each) 40195327349449 (case), Lot Numbers: 23FBQ393; REF DYNJ48130K: UDI/DI 10195327349448 (each) 40195327349449 (case), Lot Numbers: 24BBL782; REF DYNJ50127D: UDI/DI 10195327212186 (each) 40195327212187 (case), Lot Numbers: 23JBD975; REF DYNJ50621K: UDI/DI 10195327002756 (each) 40195327002757 (case), Lot Numbers: 23GMI091; REF DYNJ50621L: UDI/DI 10195327501174 (each) 40195327501175 (case), Lot Numbers: 23IMH003; REF DYNJ50621L: UDI/DI 10195327501174 (each) 40195327501175 (case), Lot Numbers: 23KMB960; REF DYNJ50621L: UDI/DI 10195327501174 (each) 40195327501175 (case), Lot Numbers: 23LMF863; REF DYNJ50622K: UDI/DI 10195327501198 (each) 40195327501199 (case), Lot Numbers: 23KMF214; REF DYNJ50622K: UDI/DI 10195327501198 (each) 40195327501199 (case), Lot Numbers: 24AMD236; REF DYNJ50622K: UDI/DI 10195327501198 (each) 40195327501199 (case), Lot Numbers: 24BME594; REF DYNJ51567G: UDI/DI 10195327430153 (each) 40195327430154 (case), Lot Numbers: 23KBR665; REF DYNJ51567G: UDI/DI 10195327430153 (each) 40195327430154 (case), Lot Numbers: 24CBL506; REF DYNJ51567G: UDI/DI 10195327430153 (each) 40195327430154 (case), Lot Numbers: 24DBU743; REF DYNJ51567G: UDI/DI 10195327430153 (each) 40195327430154 (case), Lot Numbers: 24EBL973; REF DYNJ52007C: UDI/DI 10889942796878 (each) 40889942796879 (case), Lot Numbers: 23GMD252; REF DYNJ52007F: UDI/DI 10195327563141 (each) 40195327563142 (case), Lot Numbers: 23LBE651; REF DYNJ52007F: UDI/DI 10195327563141 (each) 40195327563142 (case), Lot Numbers: 24DBB359; REF DYNJ52007F: UDI/DI 10195327563141 (each) 40195327563142 (case), Lot Numbers: 24DBO178; REF DYNJ52331F: UDI/DI 10195327237691 (each) 40195327237692 (case), Lot Numbers: 24ABQ379; REF DYNJ52332G: UDI/DI 10195327237707 (each) 40195327237708 (case), Lot Numbers: 23EBQ263; REF DYNJ52332I: UDI/DI 10195327447731 (each) 40195327447732 (case), Lot Numbers: 23FBV979; REF DYNJ52332I: UDI/DI 10195327447731 (each) 40195327447732 (case), Lot Numbers: 23KBI476; REF DYNJ52332I: UDI/DI 10195327447731 (each) 40195327447732 (case), Lot Numbers: 24ABS792; REF DYNJ52333F: UDI/DI 10195327434083 (each) 40195327434084 (case), Lot Numbers: 23HBE931; REF DYNJ52333F: UDI/DI 10195327434083 (each) 40195327434084 (case), Lot Numbers: 23JBB688; REF DYNJ52333F: UDI/DI 10195327434083 (each) 40195327434084 (case), Lot Numbers: 24ABO902; REF DYNJ53006C: UDI/DI 10193489782745 (each) 40193489782746 (case), Lot Numbers: 23HMC677; REF DYNJ53006C: UDI/DI 10193489782745 (each) 40193489782746 (case), Lot Numbers: 23JMA848; REF DYNJ53006C: UDI/DI 10193489782745 (each) 40193489782746 (case), Lot Numbers: 23LMA051; REF DYNJ53006D: UDI/DI 10195327628390 (each) 40195327628391 (case), Lot Numbers: 24CBP971; REF DYNJ53555F: UDI/DI 10195327410056 (each) 40195327410057 (case), Lot Numbers: 24ABR006; REF DYNJ53555F: UDI/DI 10195327410056 (each) 40195327410057 (case), Lot Numbers: 24DBK404; REF DYNJ53555F: UDI/DI 10195327410056 (each) 40195327410057 (case), Lot Numbers: 24EBA656; REF DYNJ55852C: UDI/DI 10195327267575 (each) 40195327267576 (case), Lot Numbers: 23GMF007; REF DYNJ55852C: UDI/DI 10195327267575 (each) 40195327267576 (case), Lot Numbers: 24CMI123; REF DYNJ57648D: UDI/DI 10195327027391 (each) 40195327027392 (case), Lot Numbers: 23IMG853; REF DYNJ57648D: UDI/DI 10195327027391 (each) 40195327027392 (case), Lot Numbers: 23KMH101; REF DYNJ57648D: UDI/DI 10195327027391 (each) 40195327027392 (case), Lot Numbers: 23LME559; REF DYNJ57648F: UDI/DI 10195327610357 (each) 40195327610358 (case), Lot Numbers: 24BME622; REF DYNJ58557C: UDI/DI 10193489972276 (each) 40193489972277 (case), Lot Numbers: 23KMF224; REF DYNJ58557C: UDI/DI 10193489972276 (each) 40193489972277 (case), Lot Numbers: 23LMG741; REF DYNJ58557D: UDI/DI 10195327615192 (each) 40195327615193 (case), Lot Numbers: 24BME252; REF DYNJ58563D: UDI/DI 10193489972320 (each) 40193489972321 (case), Lot Numbers: 23IMG567; REF DYNJ58564D: UDI/DI 10193489972337 (each) 40193489972338 (case), Lot Numbers: 23KMD387; REF DYNJ58665F: UDI/DI 10195327529741 (each) 40195327529742 (case), Lot Numbers: 23KMH652; REF DYNJ58665F: UDI/DI 10195327529741 (each) 40195327529742 (case), Lot Numbers: 23LMB459; REF DYNJ58665F: UDI/DI 10195327529741 (each) 40195327529742 (case), Lot Numbers: 24CME058; REF DYNJ58665F: UDI/DI 10195327529741 (each) 40195327529742 (case), Lot Numbers: 24DMG301; REF DYNJ59819A: UDI/DI 10195327331597 (each) 40195327331598 (case), Lot Numbers: 23GBF959; REF DYNJ59819A: UDI/DI 10195327331597 (each) 40195327331598 (case), Lot Numbers: 23LBD961; REF DYNJ60049B: UDI/DI 10193489782776 (each) 40193489782777 (case), Lot Numbers: 23HBH184; REF DYNJ60049C: UDI/DI 10195327628413 (each) 40195327628414 (case), Lot Numbers: 24BBU890; REF DYNJ60049C: UDI/DI 10195327628413 (each) 40195327628414 (case), Lot Numbers: 24DBS448; REF DYNJ60194C: UDI/DI 10195327529765 (each) 40195327529766 (case), Lot Numbers: 23LMA120; REF DYNJ60495I: UDI/DI 10195327599645 (each) 40195327599646 (case), Lot Numbers: 24ABB590; REF DYNJ60550B: UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers: 23EBB225; REF DYNJ60550B: UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers: 23EBM228; REF DYNJ60550B: UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers: 23GBW407; REF DYNJ60550B: UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers: 23JBC621; REF DYNJ60550B: UDI/DI 10195327361853 (each) 40195327361854 (case), Lot Numbers: 24ABM138; REF DYNJ60553B: UDI/DI 10195327361822 (each) 40195327361823 (case), Lot Numbers: 23FBD830; REF DYNJ60553B: UDI/DI 10195327361822 (each) 40195327361823 (case), Lot Numbers: 23GBV207; REF DYNJ60553B: UDI/DI 10195327361822 (each) 40195327361823 (case), Lot Numbers: 23HBX302; REF DYNJ60553B: UDI/DI 10195327361822 (each) 40195327361823 (case), Lot Numbers: 23JBJ796; REF DYNJ61340B: UDI/DI 10195327401092 (each) 40195327401093 (case), Lot Numbers: 23EBI869; REF DYNJ61414C: UDI/DI 10195327401016 (each) 40195327401017 (case), Lot Numbers: 23EBI870; REF DYNJ61651J: UDI/DI 10195327388447 (each) 40195327388448 (case), Lot Numbers: 23IMF466; REF DYNJ61651J: UDI/DI 10195327388447 (each) 40195327388448 (case), Lot Numbers: 23JMB216; REF DYNJ61651J: UDI/DI 10195327388447 (each) 40195327388448 (case), Lot Numbers: 23KMI579; REF DYNJ61651J: UDI/DI 10195327388447 (each) 40195327388448 (case), Lot Numbers: 23LMH297; REF DYNJ61651J: UDI/DI 10195327388447 (each) 40195327388448 (case), Lot Numbers: 24CMJ082; REF DYNJ61663A: UDI/DI 10193489929652 (each) 40193489929653 (case), Lot Numbers: 23EBI830; REF DYNJ61663A: UDI/DI 10193489929652 (each) 40193489929653 (case), Lot Numbers: 23IBD773; REF DYNJ61663A: UDI/DI 10193489929652 (each) 40193489929653 (case), Lot Numbers: 23IBS880; REF DYNJ62612D: UDI/DI 10195327528089 (each) 40195327528080 (case), Lot Numbers: 23LMG302; REF DYNJ62612D: UDI/DI 10195327528089 (each) 40195327528080 (case), Lot Numbers: 24BME477; REF DYNJ62612D: UDI/DI 10195327528089 (each) 40195327528080 (case), Lot Numbers: 24CMD895; REF DYNJ62612D: UDI/DI 10195327528089 (each) 40195327528080 (case), Lot Numbers: 24DMD795; REF DYNJ62650B: UDI/DI 10195327121846 (each) 40195327121847 (case), Lot Numbers: 23LBJ206; REF DYNJ62650B: UDI/DI 10195327121846 (each) 40195327121847 (case), Lot Numbers: 24ABU953; REF DYNJ62650B: UDI/DI 10195327121846 (each) 40195327121847 (case), Lot Numbers: 24DBG292; REF DYNJ62650B: UDI/DI 10195327121846 (each) 40195327121847 (case), Lot Numbers: 24DBO663; REF DYNJ63082C: UDI/DI 10195327505615 (each) 40195327505616 (case), Lot Numbers: 23KMC330; REF DYNJ63082C: UDI/DI 10195327505615 (each) 40195327505616 (case), Lot Numbers: 23KMH957; REF DYNJ63082C: UDI/DI 10195327505615 (each) 40195327505616 (case), Lot Numbers: 23LMG702; REF DYNJ63473C: UDI/DI 10195327510923 (each) 40195327510924 (case), Lot Numbers: 23IMF358; REF DYNJ63473C: UDI/DI 10195327510923 (each) 40195327510924 (case), Lot Numbers: 24BMA722; REF DYNJ63559B: UDI/DI 10193489900422 (each) 40193489900423 (case), Lot Numbers: 24ABV064; REF DYNJ64233C: UDI/DI 10195327452957 (each) 40195327452958 (case), Lot Numbers: 23HME045; REF DYNJ64233C: UDI/DI 10195327452957 (each) 40195327452958 (case), Lot Numbers: 23IMI013; REF DYNJ64240A: UDI/DI 10193489382631 (each) 40193489382632 (case), Lot Numbers: 23DMF071; REF DYNJ64276C: UDI/DI 10195327091095 (each) 40195327091096 (case), Lot Numbers: 23HBN381; REF DYNJ64276C: UDI/DI 10195327091095 (each) 40195327091096 (case), Lot Numbers: 23LBB663; REF DYNJ64276C: UDI/DI 10195327091095 (each) 40195327091096 (case), Lot Numbers: 24ABT845; REF DYNJ64276C: UDI/DI 10195327091095 (each) 40195327091096 (case), Lot Numbers: 24DBB304; REF DYNJ64672B: UDI/DI 10195327377496 (each) 40195327377497 (case), Lot Numbers: 23KMJ289; REF DYNJ64672B: UDI/DI 10195327377496 (each) 40195327377497 (case), Lot Numbers: 23LMD300; REF DYNJ64981F: UDI/DI 10195327528133 (each) 40195327528134 (case), Lot Numbers: 23KMA711; REF DYNJ64981F: UDI/DI 10195327528133 (each) 40195327528134 (case), Lot Numbers: 24AMI362; REF DYNJ64981F: UDI/DI 10195327528133 (each) 40195327528134 (case), Lot Numbers: 24DMH548; REF DYNJ65036C: UDI/DI 10195327528225 (each) 40195327528226 (case), Lot Numbers: 23KMJ862; REF DYNJ65036C: UDI/DI 10195327528225 (each) 40195327528226 (case), Lot Numbers: 24AMI363; REF DYNJ65036C: UDI/DI 10195327528225 (each) 40195327528226 (case), Lot Numbers: 24CMF696; REF DYNJ65037B: UDI/DI 10195327528232 (each) 40195327528233 (case), Lot Numbers: 23JMF193; REF DYNJ65037B: UDI/DI 10195327528232 (each) 40195327528233 (case), Lot Numbers: 24AMA504; REF DYNJ65037B: UDI/DI 10195327528232 (each) 40195327528233 (case), Lot Numbers: 24BMD982; REF DYNJ65178G: UDI/DI 10195327528270 (each) 40195327528271 (case), Lot Numbers: 23KMG308; REF DYNJ65178G: UDI/DI 10195327528270 (each) 40195327528271 (case), Lot Numbers: 23LMG257; REF DYNJ65178G: UDI/DI 10195327528270 (each) 40195327528271 (case), Lot Numbers: 24CMC043; REF DYNJ65178G: UDI/DI 10195327528270 (each) 40195327528271 (case), Lot Numbers: 24DMD461; REF DYNJ65240B: UDI/DI 10195327163778 (each) 40195327163779 (case), Lot Numbers: 24DBE167; REF DYNJ65358B: UDI/DI 10195327415037 (each) 40195327415038 (case), Lot Numbers: 23HMC318; REF DYNJ65359A: UDI/DI 10195327415020 (each) 40195327415021 (case), Lot Numbers: 23HMC429; REF DYNJ65378A: UDI/DI 10195327418076 (each) 40195327418077 (case), Lot Numbers: 23HMA809; REF DYNJ65381A: UDI/DI 10195327418090 (each) 40195327418091 (case), Lot Numbers: 23KMC819; REF DYNJ65384A: UDI/DI 10195327410070 (each) 40195327410071 (case), Lot Numbers: 23EMC378; REF DYNJ65386C: UDI/DI 10195327410087 (each) 40195327410088 (case), Lot Numbers: 23FMC511; REF DYNJ65386C: UDI/DI 10195327410087 (each) 40195327410088 (case), Lot Numbers: 23HMF637; REF DYNJ65386C: UDI/DI 10195327410087 (each) 40195327410088 (case), Lot Numbers: 24AMB920; REF DYNJ65388B: UDI/DI 10195327418106 (each) 40195327418107 (case), Lot Numbers: 23FMB854; REF DYNJ65388B: UDI/DI 10195327418106 (each) 40195327418107 (case), Lot Numbers: 23GMA062; REF DYNJ65388B: UDI/DI 10195327418106 (each) 40195327418107 (case), Lot Numbers: 23GMI628; REF DYNJ65726D: UDI/DI 10195327528263 (each) 40195327528264 (case), Lot Numbers: 23LBU922; REF DYNJ65726D: UDI/DI 10195327528263 (each) 40195327528264 (case), Lot Numbers: 24ABH394; REF DYNJ65726D: UDI/DI 10195327528263 (each) 40195327528264 (case), Lot Numbers: 24FMH618; REF DYNJ65855B: UDI/DI 10195327232146 (each) 40195327232147 (case), Lot Numbers: 23EBO883; REF DYNJ65855B: UDI/DI 10195327232146 (each) 40195327232147 (case), Lot Numbers: 23GBM189; REF DYNJ65855B: UDI/DI 10195327232146 (each) 40195327232147 (case), Lot Numbers: 23KBB910; REF DYNJ66357C: UDI/DI 10195327227975 (each) 40195327227976 (case), Lot Numbers: 23DBI896; REF DYNJ66357C: UDI/DI 10195327227975 (each) 40195327227976 (case), Lot Numbers: 23EBI969; REF DYNJ67195C: UDI/DI 10195327299705 (each) 40195327299706 (case), Lot Numbers: 23EBQ378; REF DYNJ67195C: UDI/DI 10195327299705 (each) 40195327299706 (case), Lot Numbers: 23IBR151; REF DYNJ67196B: UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23DBJ727; REF DYNJ67196B: UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23EBL268; REF DYNJ67196B: UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23GBN228; REF DYNJ67196B: UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23HBR581; REF DYNJ67196B: UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23IBF522; REF DYNJ67196B: UDI/DI 10195327299712 (each) 40195327299713 (case), Lot Numbers: 23KBC812; REF DYNJ67197B: UDI/DI 10195327299729 (each) 40195327299720 (case), Lot Numbers: 23DBJ791; REF DYNJ67197B: UDI/DI 10195327299729 (each) 40195327299720 (case), Lot Numbers: 23GBA237; REF DYNJ67197B: UDI/DI 10195327299729 (each) 40195327299720 (case), Lot Numbers: 23HBR553; REF DYNJ67801B: UDI/DI 10195327228026 (each) 40195327228027 (case), Lot Numbers: 23EBR971; REF DYNJ67801B: UDI/DI 10195327228026 (each) 40195327228027 (case), Lot Numbers: 23HBS083; REF DYNJ67802B: UDI/DI 10195327227951 (each) 40195327227952 (case), Lot Numbers: 23EBN502; REF DYNJ67802B: UDI/DI 10195327227951 (each) 40195327227952 (case), Lot Numbers: 23GBI489; REF DYNJ67802B: UDI/DI 10195327227951 (each) 40195327227952 (case), Lot Numbers: 23HBO052; REF DYNJ67886D: UDI/DI 10195327418694 (each) 40195327418695 (case), Lot Numbers: 23GLA310; REF DYNJ67886D: UDI/DI 10195327418694 (each) 40195327418695 (case), Lot Numbers: 23GLB065; REF DYNJ67886D: UDI/DI 10195327418694 (each) 40195327418695 (case), Lot Numbers: 23HLB132; REF DYNJ67886D: UDI/DI 10195327418694 (each) 40195327418695 (case), Lot Numbers: 24ALA167; REF DYNJ67887D: UDI/DI 10195327418656 (each) 40195327418657 (case), Lot Numbers: 23IMG368; REF DYNJ67983A: UDI/DI 10195327104429 (each) 40195327104420 (case), Lot Numbers: 23LBF077; REF DYNJ68275B: UDI/DI 10195327406400 (each) 40195327406401 (case), Lot Numbers: 23EBD505; REF DYNJ68275B: UDI/DI 10195327406400 (each) 40195327406401 (case), Lot Numbers: 23HBA768; REF DYNJ68275B: UDI/DI 10195327406400 (each) 40195327406401 (case), Lot Numbers: 23KBL224; REF DYNJ80424A: UDI/DI 10195327234119 (each) 40195327234110 (case), Lot Numbers: 23FBR713; REF DYNJ80424A: UDI/DI 10195327234119 (each) 40195327234110 (case), Lot Numbers: 23IBO925; REF DYNJ80424A: UDI/DI 10195327234119 (each) 40195327234110 (case), Lot Numbers: 24ABL758; REF DYNJ80424A: UDI/DI 10195327234119 (each) 40195327234110 (case), Lot Numbers: 24ABR356; REF DYNJ80435A: UDI/DI 10195327234102 (each) 40195327234103 (case), Lot Numbers: 23DBH684; REF DYNJ80435A: UDI/DI 10195327234102 (each) 40195327234103 (case), Lot Numbers: 23KBH606; REF DYNJ80450B: UDI/DI 10195327328542 (each) 40195327328543 (case), Lot Numbers: 23ILA031; REF DYNJ80450B: UDI/DI 10195327328542 (each) 40195327328543 (case), Lot Numbers: 23KLA256; REF DYNJ81508A: UDI/DI 10195327433789 (each) 40195327433780 (case), Lot Numbers: 23HMD599; REF DYNJ81598A: UDI/DI 10195327314934 (each) 40195327314935 (case), Lot Numbers: 23EBK599; REF DYNJ81598A: UDI/DI 10195327314934 (each) 40195327314935 (case), Lot Numbers: 23HBR320; REF DYNJ81598A: UDI/DI 10195327314934 (each) 40195327314935 (case), Lot Numbers: 23IBV886; REF DYNJ81598A: UDI/DI 10195327314934 (each) 40195327314935 (case), Lot Numbers: 23KBI451; REF DYNJ81608A: UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers: 23EBD853; REF DYNJ81608A: UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers: 23GBK092; REF DYNJ81608A: UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers: 23HBR322; REF DYNJ81608A: UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers: 23IBP510; REF DYNJ81608A: UDI/DI 10195327314927 (each) 40195327314928 (case), Lot Numbers: 23JBK149; REF DYNJ81612A: UDI/DI 10195327315061 (each) 40195327315062 (case), Lot Numbers: 23DBM719; REF DYNJ81612B: UDI/DI 10195327441838 (each) 40195327441839 (case), Lot Numbers: 23FBR976; REF DYNJ81612B: UDI/DI 10195327441838 (each) 40195327441839 (case), Lot Numbers: 23HBV686; REF DYNJ82950: UDI/DI 10195327218966 (each) 40195327218967 (case), Lot Numbers: 23EBS135; REF DYNJ82950: UDI/DI 10195327218966 (each) 40195327218967 (case), Lot Numbers: 23HBS279; REF DYNJ83107A: UDI/DI 10195327381950 (each) 40195327381951 (case), Lot Numbers: 23KMB670; REF DYNJ83107A: UDI/DI 10195327381950 (each) 40195327381951 (case), Lot Numbers: 24AMC085; REF DYNJ83660: UDI/DI 10195327292430 (each) 40195327292431 (case), Lot Numbers: 23LBF868; REF DYNJ83664: UDI/DI 10195327292553 (each) 40195327292554 (case), Lot Numbers: 24BBT467; REF DYNJ83760: UDI/DI 10195327301743 (each) 40195327301744 (case), Lot Numbers: 24BBQ407; REF DYNJ83760: UDI/DI 10195327301743 (each) 40195327301744 (case), Lot Numbers: 24DBA190; REF DYNJ83761: UDI/DI 10195327301750 (each) 40195327301751 (case), Lot Numbers: 23IBV052; REF DYNJ84326: UDI/DI 10195327352059 (each) 40195327352050 (case), Lot Numbers: 24ABE390; REF DYNJ84326: UDI/DI 10195327352059 (each) 40195327352050 (case), Lot Numbers: 24ABS255; REF DYNJ84327: UDI/DI 10195327352066 (each) 40195327352067 (case), Lot Numbers: 23LBQ086; REF DYNJ84327: UDI/DI 10195327352066 (each) 40195327352067 (case), Lot Numbers: 24ABS253; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 23GMD380; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 23IMF877; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 23LMF620; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 24AMG111; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 24BMJ724; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 24DMF761; REF DYNJ84674: UDI/DI 10195327398453 (each) 40195327398454 (case), Lot Numbers: 24FMB302; REF DYNJ84905: UDI/DI 10195327435080 (each) 40195327435081 (case), Lot Numbers: 23IBD776; REF DYNJ84905: UDI/DI 10195327435080 (each) 40195327435081 (case), Lot Numbers: 24DBK140; REF DYNJ85116: UDI/DI 10195327460723 (each) 40195327460724 (case), Lot Numbers: 23HBR797; REF DYNJ85116: UDI/DI 10195327460723 (each) 40195327460724 (case), Lot Numbers: 23IBS584; REF DYNJ85120: UDI/DI 10195327460709 (each) 40195327460700 (case), Lot Numbers: 23JBC563; REF DYNJ85272: UDI/DI 10195327484439 (each) 40195327484430 (case), Lot Numbers: 23LBO077; REF DYNJ85272: UDI/DI 10195327484439 (each) 40195327484430 (case), Lot Numbers: 24EBU397; REF DYNJ85273: UDI/DI 10195327484491 (each) 40195327484492 (case), Lot Numbers: 23LBP123; REF DYNJ86112: UDI/DI 10195327594794 (each) 40195327594795 (case), Lot Numbers: 24CMD842; REF DYNJ86318: UDI/DI 10195327621452 (each) 40195327621453 (case), Lot Numbers: 24CBK983; REF DYNJ86318: UDI/DI 10195327621452 (each) 40195327621453 (case), Lot Numbers: 24DBG610; REF DYNJ86318: UDI/DI 10195327621452 (each) 40195327621453 (case), Lot Numbers: 24FMD763; REF DYNJ86321: UDI/DI 10195327621681 (each) 40195327621682 (case), Lot Numbers: 24DMF764; REF DYNJ86321: UDI/DI 10195327621681 (each) 40195327621682 (case), Lot Numbers: 24FME942; REF DYNJ900211I: UDI/DI 10195327539795 (each) 40195327539796 (case), Lot Numbers: 23KBS503; REF DYNJ900267I: UDI/DI 10195327348175 (each) 40195327348176 (case), Lot Numbers: 23HBJ206; REF DYNJ900267I: UDI/DI 10195327348175 (each) 40195327348176 (case), Lot Numbers: 23KBW284; REF DYNJ900267I: UDI/DI 10195327348175 (each) 40195327348176 (case), Lot Numbers: 24ABI228; REF DYNJ900267I: UDI/DI 10195327348175 (each) 40195327348176 (case), Lot Numbers: 24ABW982; REF DYNJ900268G: UDI/DI 10195327348168 (each) 40195327348169 (case), Lot Numbers: 24BBE017; REF DYNJ900268G: UDI/DI 10195327348168 (each) 40195327348169 (case), Lot Numbers: 24CBC790; REF DYNJ900268G: UDI/DI 10195327348168 (each) 40195327348169 (case), Lot Numbers: 24DBQ318; REF DYNJ900283L: UDI/DI 10195327553975 (each) 40195327553976 (case), Lot Numbers: 24ABP008; REF DYNJ900283L: UDI/DI 10195327553975 (each) 40195327553976 (case), Lot Numbers: 24BBR835; REF DYNJ900283L: UDI/DI 10195327553975 (each) 40195327553976 (case), Lot Numbers: 24CBI681; REF DYNJ900283L: UDI/DI 10195327553975 (each) 40195327553976 (case), Lot Numbers: 24CBU940; REF DYNJ900283L: UDI/DI 10195327553975 (each) 40195327553976 (case), Lot Numbers: 24EBP449; REF DYNJ900288I: UDI/DI 10195327554026 (each) 40195327554027 (case), Lot Numbers: 23LBS777; REF DYNJ900288I: UDI/DI 10195327554026 (each) 40195327554027 (case), Lot Numbers: 24ABI414; REF DYNJ900288I: UDI/DI 10195327554026 (each) 40195327554027 (case), Lot Numbers: 24CBI950; REF DYNJ900288I: UDI/DI 10195327554026 (each) 40195327554027 (case), Lot Numbers: 24DBD787; REF DYNJ900288I: UDI/DI 10195327554026 (each) 40195327554027 (case), Lot Numbers: 24DBK841; REF DYNJ900288I: UDI/DI 10195327554026 (each) 40195327554027 (case), Lot Numbers: 24EBP454; REF DYNJ900302I: UDI/DI 10195327554132 (each) 40195327554133 (case), Lot Numbers: 23LBQ325; REF DYNJ900302I: UDI/DI 10195327554132 (each) 40195327554133 (case), Lot Numbers: 24BBA390; REF DYNJ900302I: UDI/DI 10195327554132 (each) 40195327554133 (case), Lot Numbers: 24EBN323; REF DYNJ900303M: UDI/DI 10195327175269 (each) 40195327175260 (case), Lot Numbers: 23JBL007; REF DYNJ900303N: UDI/DI 10195327554149 (each) 40195327554140 (case), Lot Numbers: 24ABV097; REF DYNJ900303N: UDI/DI 10195327554149 (each) 40195327554140 (case), Lot Numbers: 24BBF072; REF DYNJ900303N: UDI/DI 10195327554149 (each) 40195327554140 (case), Lot Numbers: 24EBA039; REF DYNJ900304O: UDI/DI 10195327554156 (each) 40195327554157 (case), Lot Numbers: 24ABB086; REF DYNJ900304O: UDI/DI 10195327554156 (each) 40195327554157 (case), Lot Numbers: 24CBG296; REF DYNJ900304O: UDI/DI 10195327554156 (each) 40195327554157 (case), Lot Numbers: 24CBJ698; REF DYNJ900304O: UDI/DI 10195327554156 (each) 40195327554157 (case), Lot Numbers: 24DBD775; REF DYNJ900304O: UDI/DI 10195327554156 (each) 40195327554157 (case), Lot Numbers: 24EBK195; REF DYNJ900606G: UDI/DI 10195327224097 (each) 40195327224098 (case), Lot Numbers: 23KDC266; REF DYNJ900624G: UDI/DI 10195327224134 (each) 40195327224135 (case), Lot Numbers: 23KDA658; REF DYNJ900651D: UDI/DI 10195327421762 (each) 40195327421763 (case), Lot Numbers: 23LBO181; REF DYNJ900651D: UDI/DI 10195327421762 (each) 40195327421763 (case), Lot Numbers: 24ABC722; REF DYNJ900651D: UDI/DI 10195327421762 (each) 40195327421763 (case), Lot Numbers: 24ABT069; REF DYNJ900651D: UDI/DI 10195327421762 (each) 40195327421763 (case), Lot Numbers: 24DBI169; REF DYNJ900651D: UDI/DI 10195327421762 (each) 40195327421763 (case), Lot Numbers: 24DBR618; REF DYNJ900651D: UDI/DI 10195327421762 (each) 40195327421763 (case), Lot Numbers: 24EBE001; REF DYNJ900658C: UDI/DI 10195327421717 (each) 40195327421718 (case), Lot Numbers: 24BBN608; REF DYNJ900713D: UDI/DI 10193489951318 (each) 40193489951319 (case), Lot Numbers: 23EBT513; REF DYNJ900713D: UDI/DI 10193489951318 (each) 40193489951319 (case), Lot Numbers: 23FBE839; REF DYNJ900713D: UDI/DI 10193489951318 (each) 40193489951319 (case), Lot Numbers: 23HBE024; REF DYNJ900713D: UDI/DI 10193489951318 (each) 40193489951319 (case), Lot Numbers: 23HBT798; REF DYNJ900718C: UDI/DI 10193489951578 (each) 40193489951579 (case), Lot Numbers: 23FBA297; REF DYNJ900718C: UDI/DI 10193489951578 (each) 40193489951579 (case), Lot Numbers: 23HBR477; REF DYNJ900720D: UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers: 23EBJ422; REF DYNJ900720D: UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers: 23FBF316; REF DYNJ900720D: UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers: 23FBV989; REF DYNJ900720D: UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers: 23GBW066; REF DYNJ900720D: UDI/DI 10193489951592 (each) 40193489951593 (case), Lot Numbers: 23JBL130; REF DYNJ900722C: UDI/DI 10193489951615 (each) 40193489951616 (case), Lot Numbers: 23FBC702; REF DYNJ900722C: UDI/DI 10193489951615 (each) 40193489951616 (case), Lot Numbers: 23GBS042; REF DYNJ900722C: UDI/DI 10193489951615 (each) 40193489951616 (case), Lot Numbers: 23HBT411; REF DYNJ900722C: UDI/DI 10193489951615 (each) 40193489951616 (case), Lot Numbers: 23IBU127; REF DYNJ901069L: UDI/DI 10195327471279 (each) 40195327471270 (case), Lot Numbers: 23JMD456; REF DYNJ901648G: UDI/DI 10195327487874 (each) 40195327487875 (case), Lot Numbers: 24CBC570; REF DYNJ901648G: UDI/DI 10195327487874 (each) 40195327487875 (case), Lot Numbers: 24CBC758; REF DYNJ901648G: UDI/DI 10195327487874 (each) 40195327487875 (case), Lot Numbers: 24DBJ348; REF DYNJ901653I: UDI/DI 10195327487867 (each) 40195327487868 (case), Lot Numbers: 24DBB739; REF DYNJ901653I: UDI/DI 10195327487867 (each) 40195327487868 (case), Lot Numbers: 24DBL920; REF DYNJ901653I: UDI/DI 10195327487867 (each) 40195327487868 (case), Lot Numbers: 24DBV191; REF DYNJ901653I: UDI/DI 10195327487867 (each) 40195327487868 (case), Lot Numbers: 24FBF057; REF DYNJ901658K: UDI/DI 10195327487973 (each) 40195327487974 (case), Lot Numbers: 24BBK152; REF DYNJ901658K: UDI/DI 10195327487973 (each) 40195327487974 (case), Lot Numbers: 24CBJ491; REF DYNJ901658K: UDI/DI 10195327487973 (each) 40195327487974 (case), Lot Numbers: 24DBD087; REF DYNJ901658K: UDI/DI 10195327487973 (each) 40195327487974 (case), Lot Numbers: 24EBH026; REF DYNJ901659K: UDI/DI 10195327487980 (each) 40195327487981 (case), Lot Numbers: 24ABQ886; REF DYNJ901659K: UDI/DI 10195327487980 (each) 40195327487981 (case), Lot Numbers: 24BBD989; REF DYNJ901659K: UDI/DI 10195327487980 (each) 40195327487981 (case), Lot Numbers: 24CBA350; REF DYNJ901659K: UDI/DI 10195327487980 (each) 40195327487981 (case), Lot Numbers: 24EBI419; REF DYNJ901659K: UDI/DI 10195327487980 (each) 40195327487981 (case), Lot Numbers: 24FBP189; REF DYNJ902974O: UDI/DI 10195327333997 (each) 40195327333998 (case), Lot Numbers: 23GDA383; REF DYNJ902978L: UDI/DI 10195327147525 (each) 40195327147526 (case), Lot Numbers: 23LDB471; REF DYNJ902978L: UDI/DI 10195327147525 (each) 40195327147526 (case), Lot Numbers: 24ADB459; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 23HBT784; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 23IBJ229; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 23JBB030; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 23LBP019; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 23LBQ443; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 24BBR761; REF DYNJ903043K: UDI/DI 10195327378974 (each) 40195327378975 (case), Lot Numbers: 24EBH011; REF DYNJ903561C: UDI/DI 10195327348212 (each) 40195327348213 (case), Lot Numbers: 24DBI297; REF DYNJ903561C: UDI/DI 10195327348212 (each) 40195327348213 (case), Lot Numbers: 24DBT204; REF DYNJ903561C: UDI/DI 10195327348212 (each) 40195327348213 (case), Lot Numbers: 24JBA560; REF DYNJ903734C: UDI/DI 10195327592530 (each) 40195327592531 (case), Lot Numbers: 24CBE953; REF DYNJ903734C: UDI/DI 10195327592530 (each) 40195327592531 (case), Lot Numbers: 24CBJ644; REF DYNJ903734C: UDI/DI 10195327592530 (each) 40195327592531 (case), Lot Numbers: 24CBT593; REF DYNJ903803C: UDI/DI 10193489274981 (each) 40193489274982 (case), Lot Numbers: 24ABB471; REF DYNJ903803C: UDI/DI 10193489274981 (each) 40193489274982 (case), Lot Numbers: 24BBM812; REF DYNJ903803D: UDI/DI 10195327640613 (each) 40195327640614 (case), Lot Numbers: 24CBT595; REF DYNJ903863D: UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 23JMI474; REF DYNJ903863D: UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 23LMC709; REF DYNJ903863D: UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 24AMB271; REF DYNJ903863D: UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 24CMJ920; REF DYNJ903863D: UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 24DME141; REF DYNJ903863D: UDI/DI 10195327236632 (each) 40195327236633 (case), Lot Numbers: 24EMD044; REF DYNJ903863F: UDI/DI 10198459086175 (each) 40198459086176 (case), Lot Numbers: 24LMH419; REF DYNJ904024D: UDI/DI 10195327541231 (each) 40195327541232 (case), Lot Numbers: 23KBP053; REF DYNJ904024D: UDI/DI 10195327541231 (each) 40195327541232 (case), Lot Numbers: 23KBP705; REF DYNJ904024D: UDI/DI 10195327541231 (each) 40195327541232 (case), Lot Numbers: 23LBQ455; REF DYNJ904078J: UDI/DI 10195327427450 (each) 40195327427451 (case), Lot Numbers: 23GBS725; REF DYNJ904078J: UDI/DI 10195327427450 (each) 40195327427451 (case), Lot Numbers: 23JBH152; REF DYNJ904078K: UDI/DI 10195327516741 (each) 40195327516742 (case), Lot Numbers: 24CBG810; REF DYNJ904698F: UDI/DI 10195327169282 (each) 40195327169283 (case), Lot Numbers: 23FDA007; REF DYNJ904698F: UDI/DI 10195327169282 (each) 40195327169283 (case), Lot Numbers: 23FDB074; REF DYNJ904698F: UDI/DI 10195327169282 (each) 40195327169283 (case), Lot Numbers: 23KDB010; REF DYNJ904702G: UDI/DI 10195327169312 (each) 40195327169313 (case), Lot Numbers: 23EDB534; REF DYNJ904702G: UDI/DI 10195327169312 (each) 40195327169313 (case), Lot Numbers: 23GDB294; REF DYNJ904703F: UDI/DI 10195327169268 (each) 40195327169269 (case), Lot Numbers: 23FDB070; REF DYNJ904703F: UDI/DI 10195327169268 (each) 40195327169269 (case), Lot Numbers: 23IDA580; REF DYNJ904704D: UDI/DI 10195327126247 (each) 40195327126248 (case), Lot Numbers: 23FDA288; REF DYNJ904704D: UDI/DI 10195327126247 (each) 40195327126248 (case), Lot Numbers: 23GDB551; REF DYNJ904967D: UDI/DI 10195327453572 (each) 40195327453573 (case), Lot Numbers: 24ABT016; REF DYNJ904967D: UDI/DI 10195327453572 (each) 40195327453573 (case), Lot Numbers: 24CBH283; REF DYNJ905078F: UDI/DI 10193489437768 (each) 40193489437769 (case), Lot Numbers: 23IBJ925; REF DYNJ905078F: UDI/DI 10193489437768 (each) 40193489437769 (case), Lot Numbers: 24ABB910; REF DYNJ905078F: UDI/DI 10193489437768 (each) 40193489437769 (case), Lot Numbers: 24CBO467; REF DYNJ905080G: UDI/DI 10195327046552 (each) 40195327046553 (case), Lot Numbers: 23HBX447; REF DYNJ905080G: UDI/DI 10195327046552 (each) 40195327046553 (case), Lot Numbers: 23IBI067; REF DYNJ905196J: UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 23EMG331; REF DYNJ905196J: UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 23FMC662; REF DYNJ905196J: UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 23GMI565; REF DYNJ905196J: UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 23IMD620; REF DYNJ905196J: UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 23JMF477; REF DYNJ905196J: UDI/DI 10195327230821 (each) 40195327230822 (case), Lot Numbers: 24BMG170; REF DYNJ905443C: UDI/DI 10193489883688 (each) 40193489883689 (case), Lot Numbers: 24BBB918; REF DYNJ905443C: UDI/DI 10193489883688 (each) 40193489883689 (case), Lot Numbers: 24BBC021; REF DYNJ905443C: UDI/DI 10193489883688 (each) 40193489883689 (case), Lot Numbers: 24CBJ604; REF DYNJ905443C: UDI/DI 10193489883688 (each) 40193489883689 (case), Lot Numbers: 24EBM286; REF DYNJ905604J: UDI/DI 10195327465827 (each) 40195327465828 (case), Lot Numbers: 23GBU275; REF DYNJ905604K: UDI/DI 10195327505332 (each) 40195327505333 (case), Lot Numbers: 23IBL740; REF DYNJ905604K: UDI/DI 10195327505332 (each) 40195327505333 (case), Lot Numbers: 24ABI706; REF DYNJ905604K: UDI/DI 10195327505332 (each) 40195327505333 (case), Lot Numbers: 24BBC924; REF DYNJ905604K: UDI/DI 10195327505332 (each) 40195327505333 (case), Lot Numbers: 24CBU965; REF DYNJ905604K: UDI/DI 10195327505332 (each) 40195327505333 (case), Lot Numbers: 24EBA600; REF DYNJ905605K: UDI/DI 10195327505349 (each) 40195327505340 (case), Lot Numbers: 23LBU989; REF DYNJ905605K: UDI/DI 10195327505349 (each) 40195327505340 (case), Lot Numbers: 24ABX876; REF DYNJ905605K: UDI/DI 10195327505349 (each) 40195327505340 (case), Lot Numbers: 24BBH029; REF DYNJ905605K: UDI/DI 10195327505349 (each) 40195327505340 (case), Lot Numbers: 24CBO685; REF DYNJ905867D: UDI/DI 10195327482299 (each) 40195327482290 (case), Lot Numbers: 24BBT678; REF DYNJ905867D: UDI/DI 10195327482299 (each) 40195327482290 (case), Lot Numbers: 24CBS067; REF DYNJ905867D: UDI/DI 10195327482299 (each) 40195327482290 (case), Lot Numbers: 24EBP498; REF DYNJ905873D: UDI/DI 10195327482442 (each) 40195327482443 (case), Lot Numbers: 24ABP634; REF DYNJ905873D: UDI/DI 10195327482442 (each) 40195327482443 (case), Lot Numbers: 24BBL945; REF DYNJ905873D: UDI/DI 10195327482442 (each) 40195327482443 (case), Lot Numbers: 24CBO397; REF DYNJ905873D: UDI/DI 10195327482442 (each) 40195327482443 (case), Lot Numbers: 24DBS088; REF DYNJ905874D: UDI/DI 10195327482459 (each) 40195327482450 (case), Lot Numbers: 24CBI849; REF DYNJ906223B: UDI/DI 10195327344405 (each) 40195327344406 (case), Lot Numbers: 23FBN451; REF DYNJ906223B: UDI/DI 10195327344405 (each) 40195327344406 (case), Lot Numbers: 23FBN526; REF DYNJ906223B: UDI/DI 10195327344405 (each) 40195327344406 (case), Lot Numbers: 23HBL719; REF DYNJ906223B: UDI/DI 10195327344405 (each) 40195327344406 (case), Lot Numbers: 23JBA212; REF DYNJ906259B: UDI/DI 10195327344429 (each) 40195327344420 (case), Lot Numbers: 23FBL777; REF DYNJ906259B: UDI/DI 10195327344429 (each) 40195327344420 (case), Lot Numbers: 23HBH067; REF DYNJ906259B: UDI/DI 10195327344429 (each) 40195327344420 (case), Lot Numbers: 23JBI947; REF DYNJ906261C: UDI/DI 10195327344436 (each) 40195327344437 (case), Lot Numbers: 23HBF248; REF DYNJ906261D: UDI/DI 10195327530082 (each) 40195327530083 (case), Lot Numbers: 23KBU806; REF DYNJ906261D: UDI/DI 10195327530082 (each) 40195327530083 (case), Lot Numbers: 24BBG450; REF DYNJ906262C: UDI/DI 10195327344368 (each) 40195327344369 (case), Lot Numbers: 23IBK874; REF DYNJ906262C: UDI/DI 10195327344368 (each) 40195327344369 (case), Lot Numbers: 23JBP288; REF DYNJ906539F: UDI/DI 10195327427412 (each) 40195327427413 (case), Lot Numbers: 23GBE724; REF DYNJ906539F: UDI/DI 10195327427412 (each) 40195327427413 (case), Lot Numbers: 23HBY972; REF DYNJ906539I: UDI/DI 10195327516628 (each) 40195327516629 (case), Lot Numbers: 23LBV716; REF DYNJ906539I: UDI/DI 10195327516628 (each) 40195327516629 (case), Lot Numbers: 24BBQ098; REF DYNJ906539I: UDI/DI 10195327516628 (each) 40195327516629 (case), Lot Numbers: 24CBC842; REF DYNJ906539I: UDI/DI 10195327516628 (each) 40195327516629 (case), Lot Numbers: 24DBP613; REF DYNJ906602D: UDI/DI 10195327230814 (each) 40195327230815 (case), Lot Numbers: 23IMA714; REF DYNJ906602D: UDI/DI 10195327230814 (each) 40195327230815 (case), Lot Numbers: 24BMA977; REF DYNJ906658F: UDI/DI 10195327000899 (each) 40195327000890 (case), Lot Numbers: 24ADB515; REF DYNJ906665G: UDI/DI 10195327092481 (each) 40195327092482 (case), Lot Numbers: 23LDB928; REF DYNJ906665G: UDI/DI 10195327092481 (each) 40195327092482 (case), Lot Numbers: 24ADB791; REF DYNJ906666J: UDI/DI 10195327129590 (each) 40195327129591 (case), Lot Numbers: 24BBC879; REF DYNJ906666J: UDI/DI 10195327129590 (each) 40195327129591 (case), Lot Numbers: 24CBB594; REF DYNJ906666J: UDI/DI 10195327129590 (each) 40195327129591 (case), Lot Numbers: 24EBM259; REF DYNJ906666J: UDI/DI 10195327129590 (each) 40195327129591 (case), Lot Numbers: 24FBT066; REF DYNJ906938A: UDI/DI 10195327453343 (each) 40195327453344 (case), Lot Numbers: 24ABF414; REF DYNJ906938A: UDI/DI 10195327453343 (each) 40195327453344 (case), Lot Numbers: 24BBG466; REF DYNJ907022C: UDI/DI 10193489846577 (each) 40193489846578 (case), Lot Numbers: 24ABB162; REF DYNJ907022C: UDI/DI 10193489846577 (each) 40193489846578 (case), Lot Numbers: 24BBO676; REF DYNJ907038F: UDI/DI 10195327372682 (each) 40195327372683 (case), Lot Numbers: 23EBJ311; REF DYNJ907038F: UDI/DI 10195327372682 (each) 40195327372683 (case), Lot Numbers: 23FBO835; REF DYNJ907038F: UDI/DI 10195327372682 (each) 40195327372683 (case), Lot Numbers: 23FBP603; REF DYNJ907581A: UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 23GBJ494; REF DYNJ907581A: UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 23IBD950; REF DYNJ907581A: UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 23LBD400; REF DYNJ907581A: UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 24ABA308; REF DYNJ907581A: UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 24ABC683; REF DYNJ907581A: UDI/DI 10195327045296 (each) 40195327045297 (case), Lot Numbers: 24DBR230; REF DYNJ907790D: UDI/DI 10195327431655 (each) 40195327431656 (case), Lot Numbers: 23HBQ547; REF DYNJ907790G: UDI/DI 10195327627041 (each) 40195327627042 (case), Lot Numbers: 24BBG059; REF DYNJ908063A: UDI/DI 10195327125554 (each) 40195327125555 (case), Lot Numbers: 23IBT942; REF DYNJ908063A: UDI/DI 10195327125554 (each) 40195327125555 (case), Lot Numbers: 23LBJ390; REF DYNJ908063A: UDI/DI 10195327125554 (each) 40195327125555 (case), Lot Numbers: 23LBT934; REF DYNJ908063A: UDI/DI 10195327125554 (each) 40195327125555 (case), Lot Numbers: 24FBD370; REF DYNJ908067A: UDI/DI 10195327125721 (each) 40195327125722 (case), Lot Numbers: 23HBI562; REF DYNJ908067A: UDI/DI 10195327125721 (each) 40195327125722 (case), Lot Numbers: 24CBG151; REF DYNJ908070A: UDI/DI 10195327125752 (each) 40195327125753 (case), Lot Numbers: 23HBE198; REF DYNJ908070A: UDI/DI 10195327125752 (each) 40195327125753 (case), Lot Numbers: 24ABH359; REF DYNJ908070A: UDI/DI 10195327125752 (each) 40195327125753 (case), Lot Numbers: 24DBR226; REF DYNJ908070A: UDI/DI 10195327125752 (each) 40195327125753 (case), Lot Numbers: 24EBL488; REF DYNJ908071A: UDI/DI 10195327125769 (each) 40195327125760 (case), Lot Numbers: 23KLA905; REF DYNJ908071A: UDI/DI 10195327125769 (each) 40195327125760 (case), Lot Numbers: 23LLB053; REF DYNJ908073: UDI/DI 10193489967869 (each) 40193489967860 (case), Lot Numbers: 23LBU985; REF DYNJ908073: UDI/DI 10193489967869 (each) 40193489967860 (case), Lot Numbers: 24FBB806; REF DYNJ908081A: UDI/DI 10195327479145 (each) 40195327479146 (case), Lot Numbers: 23IBE003; REF DYNJ908081A: UDI/DI 10195327479145 (each) 40195327479146 (case), Lot Numbers: 23LBO168; REF DYNJ908081A: UDI/DI 10195327479145 (each) 40195327479146 (case), Lot Numbers: 24ABT613; REF DYNJ908081A: UDI/DI 10195327479145 (each) 40195327479146 (case), Lot Numbers: 24BBU754; REF DYNJ908081A: UDI/DI 10195327479145 (each) 40195327479146 (case), Lot Numbers: 24CBS060; REF DYNJ908331C: UDI/DI 10195327474423 (each) 40195327474424 (case), Lot Numbers: 24CBE962; REF DYNJ908331C: UDI/DI 10195327474423 (each) 40195327474424 (case), Lot Numbers: 24DBD832; REF DYNJ908332B: UDI/DI 10195327474416 (each) 40195327474417 (case), Lot Numbers: 23JBM281; REF DYNJ908638: UDI/DI 10195327122287 (each) 40195327122288 (case), Lot Numbers: 23LBN391; REF DYNJ908638: UDI/DI 10195327122287 (each) 40195327122288 (case), Lot Numbers: 24BBM097; REF DYNJ908638: UDI/DI 10195327122287 (each) 40195327122288 (case), Lot Numbers: 24CBQ357; REF DYNJ908638: UDI/DI 10195327122287 (each) 40195327122288 (case), Lot Numbers: 24EBK198; REF DYNJ908642D: UDI/DI 10195327404291 (each) 40195327404292 (case), Lot Numbers: 23EBT798; REF DYNJ908642D: UDI/DI 10195327404291 (each) 40195327404292 (case), Lot Numbers: 23GBI824; REF DYNJ908642D: UDI/DI 10195327404291 (each) 40195327404292 (case), Lot Numbers: 23HBO429; REF DYNJ908642D: UDI/DI 10195327404291 (each) 40195327404292 (case), Lot Numbers: 23IBF496; REF DYNJ908642D: UDI/DI 10195327404291 (each) 40195327404292 (case), Lot Numbers: 23JBY557; REF DYNJ908644C: UDI/DI 10195327315016 (each) 40195327315017 (case), Lot Numbers: 23EBK932; REF DYNJ908644D: UDI/DI 10195327404307 (each) 40195327404308 (case), Lot Numbers: 23GBG730; REF DYNJ908644D: UDI/DI 10195327404307 (each) 40195327404308 (case), Lot Numbers: 23HBF890; REF DYNJ908647A: UDI/DI 10195327168438 (each) 40195327168439 (case), Lot Numbers: 23GBT172; REF DYNJ908647A: UDI/DI 10195327168438 (each) 40195327168439 (case), Lot Numbers: 23JBS907; REF DYNJ908998A: UDI/DI 10195327439910 (each) 40195327439911 (case), Lot Numbers: 23IDA279; REF DYNJ909004A: UDI/DI 10195327444129 (each) 40195327444120 (case), Lot Numbers: 23GDB166; REF DYNJ909162: UDI/DI 10195327281861 (each) 40195327281862 (case), Lot Numbers: 23EBV296; REF DYNJ909162: UDI/DI 10195327281861 (each) 40195327281862 (case), Lot Numbers: 23FBF595; REF DYNJ909162: UDI/DI 10195327281861 (each) 40195327281862 (case), Lot Numbers: 23GBM593; REF DYNJ909405A: UDI/DI 10195327402006 (each) 40195327402007 (case), Lot Numbers: 23EMF292; REF DYNJ909405A: UDI/DI 10195327402006 (each) 40195327402007 (case), Lot Numbers: 23EMG875; REF DYNJ909405A: UDI/DI 10195327402006 (each) 40195327402007 (case), Lot Numbers: 23JMH421; REF DYNJ909462: UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers: 23GBV029; REF DYNJ909462: UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers: 23HBC140; REF DYNJ909462: UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers: 23JBD213; REF DYNJ909462: UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers: 23LBR507; REF DYNJ909462: UDI/DI 10195327361563 (each) 40195327361564 (case), Lot Numbers: 24BBO009; REF DYNJ909462A: UDI/DI 10195327637163 (each) 40195327637164 (case), Lot Numbers: 24DBM791; REF DYNJ909462A: UDI/DI 10195327637163 (each) 40195327637164 (case), Lot Numbers: 24EBI621; REF DYNJ909463: UDI/DI 10195327361570 (each) 40195327361571 (case), Lot Numbers: 23LBM681; REF DYNJ909463: UDI/DI 10195327361570 (each) 40195327361571 (case), Lot Numbers: 24BBI537; REF DYNJ909463A: UDI/DI 10195327637002 (each) 40195327637003 (case), Lot Numbers: 24CBP241; REF DYNJ909463A: UDI/DI 10195327637002 (each) 40195327637003 (case), Lot Numbers: 24CBU623; REF DYNJ909463A: UDI/DI 10195327637002 (each) 40195327637003 (case), Lot Numbers: 24CBU918; REF DYNJ909463A: UDI/DI 10195327637002 (each) 40195327637003 (case), Lot Numbers: 24EBH688; REF DYNJ909464: UDI/DI 10195327361594 (each) 40195327361595 (case), Lot Numbers: 23LBG642; REF DYNJ909464: UDI/DI 10195327361594 (each) 40195327361595 (case), Lot Numbers: 24ABJ463; REF DYNJ909464: UDI/DI 10195327361594 (each) 40195327361595 (case), Lot Numbers: 24BBT247; REF DYNJ909464: UDI/DI 10195327361594 (each) 40195327361595 (case), Lot Numbers: 24EBI231; REF DYNJ909505A: UDI/DI 10195327448448 (each) 40195327448449 (case), Lot Numbers: 23JME303; REF DYNJ909505A: UDI/DI 10195327448448 (each) 40195327448449 (case), Lot Numbers: 23LMA802; REF DYNJ909505A: UDI/DI 10195327448448 (each) 40195327448449 (case), Lot Numbers: 24AMA487; REF DYNJ909505A: UDI/DI 10195327448448 (each) 40195327448449 (case), Lot Numbers: 24BMD766; REF DYNJ909640A: UDI/DI 10195327469047 (each) 40195327469048 (case), Lot Numbers: 24ABF489; REF DYNJ909640B: UDI/DI 10195327609917 (each) 40195327609918 (case), Lot Numbers: 24BBP338; REF DYNJ909640B: UDI/DI 10195327609917 (each) 40195327609918 (case), Lot Numbers: 24DBC804; REF DYNJ909640B: UDI/DI 10195327609917 (each) 40195327609918 (case), Lot Numbers: 24EBD304; REF DYNJ909642A: UDI/DI 10195327469054 (each) 40195327469055 (case), Lot Numbers: 23KME157; REF DYNJ909643A: UDI/DI 10195327469061 (each) 40195327469062 (case), Lot Numbers: 23LMC700; REF DYNJ909643A: UDI/DI 10195327469061 (each) 40195327469062 (case), Lot Numbers: 24AME316; REF DYNJ909643B: UDI/DI 10195327610050 (each) 40195327610051 (case), Lot Numbers: 24BBR850; REF DYNJ910165: UDI/DI 10195327604134 (each) 40195327604135 (case), Lot Numbers: 24BBO011; REF DYNJ910165: UDI/DI 10195327604134 (each) 40195327604135 (case), Lot Numbers: 24GBC474; REF DYNJ910169: UDI/DI 10195327604189 (each) 40195327604180 (case), Lot Numbers: 24CBL711; REF DYNJ910307: UDI/DI 10195327622770 (each) 40195327622771 (case), Lot Numbers: 24CBP577; REF DYNJ910308: UDI/DI 10195327622787 (each) 40195327622788 (case), Lot Numbers: 24CBN438; REF DYNJ910381: UDI/DI 10195327659844 (each) 40195327659845 (case), Lot Numbers: 24DBG976; REF DYNJ910381: UDI/DI 10195327659844 (each) 40195327659845 (case), Lot Numbers: 24EBE949; REF DYNJ910549: UDI/DI 10198459032950 (each) 40198459032951 (case), Lot Numbers: 24GMD218; REF DYNJV0277K: UDI/DI 10195327441074 (each) 40195327441075 (case), Lot Numbers: 23IMH164; REF DYNJV0277L: UDI/DI 10195327550905 (each) 40195327550906 (case), Lot Numbers: 23KMH790; REF DYNJV0277L: UDI/DI 10195327550905 (each) 40195327550906 (case), Lot Numbers: 24AMF551; REF DYNJV0277L: UDI/DI 10195327550905 (each) 40195327550906 (case), Lot Numbers: 24AMJ156; REF DYNJV0277L: UDI/DI 10195327550905 (each) 40195327550906 (case), Lot Numbers: 24FME444; REF DYNJV0279K: UDI/DI 10195327440961 (each) 40195327440962 (case), Lot Numbers: 23LMB944; REF DYNJV0279K: UDI/DI 10195327440961 (each) 40195327440962 (case), Lot Numbers: 24BME402; REF DYNJV0279K: UDI/DI 10195327440961 (each) 40195327440962 (case), Lot Numbers: 24BME565; REF DYNJV0279K: UDI/DI 10195327440961 (each) 40195327440962 (case), Lot Numbers: 24DMF496; REF DYNJV0279K: UDI/DI 10195327440961 (each) 40195327440962 (case), Lot Numbers: 24EMB250; REF DYNJV0279K: UDI/DI 10195327440961 (each) 40195327440962 (case), Lot Numbers: 24FME073; REF DYNJVB91069A: UDI/DI 10195327466404 (each) 40195327466405 (case), Lot Numbers: 23HLA812; REF DYNJVB91069A: UDI/DI 10195327466404 (each) 40195327466405 (case), Lot Numbers: 23ILA198; REF DYNJVB91069A: UDI/DI 10195327466404 (each) 40195327466405 (case), Lot Numbers: 24ALA409; REF PHS396895006F: UDI/DI 10195327454203 (each) 40195327454204 (case), Lot Numbers: 23HDB162.
Devices named in this recall
285 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- Medline Industries, Inc. DYNJ900718CMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ21826RMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. CDS780119VMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ909505AMedline Industries, Inc. · Orthopedic tray4 recalls1 report
- Medline Industries, Inc. DYNJ61663AMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ900720DMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ900722CMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ907581AMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ62650BMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ66357CMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ0160354CMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ0162022GMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ35835GMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ903803CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ42793DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ905078FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ53006CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ60049BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ43208GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ58557CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ50621KMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ51843FMedline Industries, Inc. · Cardiovascular procedure kit3 recalls0 reports
- Medline Industries, Inc. DYNJ57648DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ64276CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ906665GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ906666JMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ0536892UMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS983610LMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ46698GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67801BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ905196JMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ52331FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ52332GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS983907LMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS983915MMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS983909PMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ83761Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ81612AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ38234GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ908642DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ908644DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS984254MMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ43218GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS981923XMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ48091IMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ63082CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS780208IMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ58665FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ910307Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ910308Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ40995BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ65240BMedline Industries, Inc. · Orthopedic tray2 recalls1 report
- Medline Industries, Inc. CDS984253LMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ64240AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS941065LMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0536903ALMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0530895AOMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ907022CMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS983031QMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS940823ABMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS940963YMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ905443CMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ46676FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63559BMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900713DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43226IMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43215IMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43220GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43225GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908073Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ58563DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ58564DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906658FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ905080GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0519273NMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ67983AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908638Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908063AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908067AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908070AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908071AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904704DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44860IMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44688IMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44685LMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44687LMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44849FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44878DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44681JMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902978LMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS940843AAMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908647AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904703FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904698FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904702GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900303MMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ82950Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0885263FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ41416GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900606GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97846).
- Z-0511-2026 · Open, ClassifiedMEDLINE convenience kits labeled as LOCAL BASIC, REF DYNJ34418M
- Z-0512-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: CATH LAB PACEMAKER PACK, REF DYNJ59818A
- Z-0513-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GENERAL AAA #11-RF, REF CDS840261AB; 2) CAROTID ENDARTERECTOMY CDS-LF, REF CDS983637I; 3) PERIPHERAL VASCULAR CDS, REF CDS983723J; 4) A…
- Z-0514-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) CENTRAL LINE PACK, REF DYNJ00281P; 2) FREE FLAP PACK, REF DYNJ68213C.
- Z-0515-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) BREAST BIOPSY PACK-LF, REF DYNJ0160598C; 2) BIOPSY BREAST PACK, REF DYNJ65361A; 3) BREAST BIOPSY/MASTECTOMY, REF DYNJ908839A.
- Z-0516-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) HEAD AND NECK, REF CDS984259O; 2) HEAD AND NECK, REF CDS984259P; 3) MINOR ENT HARPER PACK-LF, REF DYNJ0160700F; 4) MAJOR ENT HARPER PAC…
- Z-0517-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MAJOR LAPAROTOMY CDS, REF CDS860015W; 2) MAJOR LAPAROTOMY CDS, REF CDS860015X; 3) MINOR LAPAROTOMY CDS, REF CDS860016R; 4) MINOR LAPARO…
- Z-0518-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597
- Z-0519-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) ACH PLASTICS BREAST-LF, REF CDS983639K; 2) BREAST ABD, REF CDS984267J; 3) PLASTIC PACK, REF DYNJ00264M; 4) BREAST ABDOMINOPLASTY MZ PK-…
- Z-0520-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MINOR SINGLE BASIN PACK, REF DYNJ37698L; 2) BASIN PACK, REF DYNJ50123D; 3) MAJOR BASIN PACK, REF DYNJ59816A; 4) OSC BASIN PACK, REF DYN…
- Z-0521-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GENERAL LAPAROSCOPY, REF CDS930027W; 2) GENERAL MAJOR CDS, REF CDS980245X; 3) BASIC GENERAL CDS, REF CDS980566F; 4) TOTAL KNEE CDS, REF…
- Z-0522-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, RE…
- Z-0523-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MAJOR VAGINAL HARPER PACK-LF, REF DYNJ0161768G; 2) ACH MAJOR VAGINAL PACK, REF DYNJ67757B.
- Z-0524-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GYN LAPAROSCOPY CDS #34-RF, REF CDS920099S; 2) GYN LAPAROSCOPY CDS-LF, REF CDS983764J; 3) ROBOTIC, REF CDS984262N; 4) ROBOTIC, REF CDS9…
- Z-0525-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: CIRCUMCISION PACK, REF DYNJ16826O
- Z-0526-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) C-SECTION OR 30 CDS, REF CDS820075W; 2) C-SECTION, REF CDS984261J; 3) C-SECTION CDS, REF CDS985282F; 4) C-SECTION PACK, REF DYNJ02577D;…
- Z-0527-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F; 2) MAJOR LITHOTOMY PACK-LF, REF DYNJ0425778O; 3) MINOR VAGINAL #76-RF, REF DYNJ27434R; …
- Z-0528-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: GP-LEEP PACK-LF, REF PHS396969004F
- Z-0529-2026 · Open, ClassifiedMEDLINE convenience kits labeled as: 1) MINOR EYE TRAY #81, REF DYNJ27466S; 2) MINOR EYE TRAY #81, REF DYNJ27466T.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page