The LeadCare II Blood Lead Test System relies on… recallZ-0584-2026 · Magellan Diagnostics, Inc. · initiated Feb 23, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Magellan Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.

FDA root cause category: Labeling mix-ups

What the company is doing

On March 3, 2024, MEDICAL DEVICE FIELD CORRECTION letters were sent to customers. Immediate Actions: 1. Identify existing inventory of impacted product and suspend its distribution. 2. Identify customers that received the impacted product and notify them about this field action. 3. Return remaining impacted product to Magellan Diagnostics, Inc. for destruction. To initiate the return process, ensure that you note the quantity of product to be returned on your Confirmation of Notification Form. Actions to Be Taken by Magellan: 1. Magellan Diagnostics will continue to investigate the cause of the labeling error and initiate appropriate preventive actions. 2. Magellan will replace unexpired impacted product upon request. Product will be replaced based on availability, and on a first-come first-served basis.

Recall details

Recall number
Z-0584-2026
Status
Open, Classified
Initiated by the company
Feb 23, 2024
Posted by the FDA
Nov 21, 2025
Recalling firm
Magellan Diagnostics, Inc.
Quantity
1,774 units
Product code
DOF
FDA submissions
K052549
Distribution
US Nationwide distribution.

Product description

The LeadCare II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)

Code information (models, lots, serial numbers)

Product Name: LeadCare¿ II Blood Lead Test Kit; Catalog Number: 70-6762; Impacted Lot/Sublot Numbers: 2333M, 2333M-01, 2333M-02, 2333M-03, 2333M-04, 2333M-05, 2333M-06, 2333M-07, 2333M-08, 2333M-09, 2333M-10, 2333M-11; Device Identifier: (01)00850355006000;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

4 devices cleared under K052549 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.