LeadCare® II Blood Lead Test KitRecalls & reported problems
Magellan Diagnostics, Inc. · LeadCare® II Blood Lead Test Kit · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Magellan LeadCare II kit measures lead in blood samples. Many of its 288 reports describe falsely high lead results, and some describe confirmation tests coming back low. The reports include no deaths and no injuries. All 4 recalls are open; they cover false positives from certain collection devices, mislabeled calibration parts, caps that may not seal, and temperature problems. Reports rose from 22 in the previous 12 months to 213 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 353
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 353 adverse event reports that list this exact device (DI 00850355006000), received between Apr 6, 2015 and Sep 30, 2026 — classified as 352 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (229 reports).
4 FDA recall records name this device's identifier, 4 still open.
8 other recalls cover devices cleared under the same FDA submission (K052549); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 13, 2026Open, Classified
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Z-1893-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 23, 2024Open, Classified
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
Z-0584-2026 · Root cause: Labeling mix-ups · Recall details · FDA record ↗
- Sep 19, 2023Open, Classified
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
Z-0219-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- May 25, 2022Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Z-1597-2022 · Root cause: Environmental control · Recall details · FDA record ↗
Most reported problems
Device problems
- False Positive Result229
- No Apparent Adverse Event56
- Nonstandard Device38
- Improper or Incorrect Procedure or Method10
- False Negative Result9
- Material Fragmentation6
- Device Handling Problem5
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions298
- Insufficient Information39
- No Known Impact Or Consequence To Patient9
- No Information6
- Laceration(s)1
Recent adverse event reports
The 25 most recent of 353 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949692 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949681 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949680 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949679 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949674 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949672 ↗MalfunctionSep 30, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949670 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934900 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934895 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934892 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934888 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934885 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934881 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934876 ↗MalfunctionSep 29, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934860 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917525 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917520 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917516 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917390 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917387 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917386 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917385 ↗MalfunctionSep 28, 2026False Positive Result
Customer reported a falsely elevated patient blood lead test result with LeadCare II.Customer reported an elevated LeadCare II pediatric patient blood lead test result of 39.1 ug/dL. Customer reported a secondary EDTA microtainer device (SARSTEDT) was used for blood collection. Customer reported a second sample was collected from the patient on the same day using a capillary tube from the LeadCare II Blood Lead Test Kit. Customer reported the second sample resulted in a LeadCare II test value of "LOW" (less than 3.3 ug/dL). Customer reported the corresponding confirmation test result was less than 1.0 ug/dL. Technical Services (TS) instructed the customer to only use the capillary tubes for blood collection provided in the LeadCare II Blood Lead Test Kit.During troubleshooting TS asked the customer to describe their method for collecting and preparing capillary blood. Customer reported they are not familiar with the procedure.TS instructed the customer to follow the CDC guidelines for capillary blood lead collection. TS provided the CDC Capillary Collection video to the customer.Customer reported Level 1 and Level 2 controls were within range according to the package insert instruc
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917384 ↗MalfunctionSep 28, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917381 ↗MalfunctionSep 28, 2026False Positive Result
Customer reported suspected falsely elevated patient blood lead test results with LeadCare II.On (b)(6)2026 the customer reported elevated patient blood lead test results and used sensor error messages with their LeadCare II Blood Lead Test Kit.Customer stated they are a manager and was unable to provide additional information at the time of contact.On (b)(6)2026 customer reported an elevated LeadCare II pediatric patient blood lead test result of 4.0 ug/dL. Customer stated the patient was not sent for venous confirmation testing.During troubleshooting Technical Services (TS) asked the customer to describe their method for collecting and preparing capillary blood. The customer stated they did not have time to describe the testing procedure in depth. TS briefly reviewed the CDC guidelines for capillary blood lead collection with the customer. TS provided the customer with the CDC Capillary Collection video.TS replaced the customer's LeadCare II Blood Lead Test Kit and analyzer due to the used sensor report.On (b)(6)2026 customer stated they would like a follow-up from TS in one month.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917379 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K052549. One submission often covers many models, so check the product and lot details in each record.
- Z-1895-2026 · Mar 13, 2026 · Open, ClassifiedLeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the following materials: " Sensors (8 containers of 24 each; 192 tests) " Tre
- Z-1952-2021 · May 7, 2021 · TerminatedLeadCare II Blood Lead Test Kit Catalog Number 70-6762
- Z-0285-2019 · Aug 8, 2018 · TerminatedMagellan LeadCare II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Bl
- Z-1295-2018 · Jan 9, 2018 · TerminatedLeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
- Z-2226-2017 · May 23, 2017 · TerminatedMagellan Diagnostics LeadCare Ultra and LeadCare Plus test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quan
- Z-3035-2017 · Jan 20, 2016 · TerminatedLeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagen
- Z-0043-2008 · Sep 20, 2007 · TerminatedESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-7142
- Z-0042-2008 · Sep 20, 2007 · TerminatedESA LeadCare II Blood Test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-6762
Other Lead, Atomic Absorption devices
- LeadCare® II Blood Lead Analyzer KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption137 reports
- LeadCare® Ultra Blood Lead Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption1 recall41 reports
- LeadCare® Plus Blood Lead Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption1 recall0 reports
- LeadCare® II 5X Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption1 recall0 reports
- LeadCare® II Blood Lead Testing System 9X Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption1 recall0 reports
- LeadCare® II Blood Lead Analyzer Kit, RefurbishedMagellan Diagnostics, Inc. · Lead, Atomic Absorption21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page