Convenience Kit, BREAST HERNIA recallZ-0447-2026 · Medline Industries, LP · initiated Oct 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Safety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-215-FG Recall Code: If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. An expanded "URGENT MEDICAL DEVICE RECALL" notification dated 10/24/2025 was provided to customers on 10/24/25, which included additional affected products. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0447-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 10, 2025
- Posted by the FDA
- Nov 13, 2025
- Recalling firm
- Medline Industries, LP
- Quantity
- 96,406 total units
- Product code
- FTN
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Product description
Convenience Kit, BREAST HERNIA DYNJ909125G
Code information (models, lots, serial numbers)
DYNJ909125G UDI-DI 40198459130664 (case), 10198459130663 (ea) Lot 25FBU543 25IBD459
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ909125GMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97882).
- Z-0442-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP…
- Z-0443-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550
- Z-0444-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
- Z-0445-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
- Z-0446-2026 · Open, ClassifiedDYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
- Z-0448-2026 · Open, ClassifiedConvenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
- Z-0449-2026 · Open, ClassifiedAnesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230
- Z-0450-2026 · Open, ClassifiedConvenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
- Z-0451-2026 · Open, ClassifiedConvenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page