Skip to content

AMSCO recallZ-0909-2026 · Steris Corporation · initiated Nov 19, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

FDA root cause category: Process control

What the company is doing

On November 19, 2025, firm began notifying customers via Urgent Medical Device Field Correction letters. All affected customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their units to perform the correction. If the affected unit has not been installed yet, the correction will be performed as part of the installation process. Users can continue to use their units prior to completion of the correction. Under normal operation, the system is designed to detect such electrical arcing specifically, indicating a Drying Over Temperature Switch Tripped alarm. Should your unit alarm, abort the cycle, turn off power to the unit, and contact STERIS.

Recall details

Recall number
Z-0909-2026
Status
Open, Classified
Initiated by the company
Nov 19, 2025
Posted by the FDA
Dec 11, 2025
Recalling firm
Steris Corporation
Quantity
10
Product code
MEC
Distribution
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.

Product description

AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

Code information (models, lots, serial numbers)

Lot Code: UDI:00724995176587, 00724995176594, 00724995176570 Serial number of distributed units: 3619625011, 3621625009, 3621625012, 3621925006, 3621925009, 3622425006, 3622725007, 3623425005, 3623825002, 3624525002

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97986).