AMSCO 7052HP (REF FH18043)Recalls & reported problems
Corporation Steris Canada · The AMSCO 7052HP Washer-Disinfector with Electric Heat delivers a validated and cleared process for cleaning reusable da Vinci EndoWrist X/Xi, S/Si, and Single-Site instruments and staplers. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00724995176594), received between Jul 26, 2023 and Sep 30, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 19, 2025Open, Classified
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Z-0909-2026 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Leak/Splash4
- Complete Loss of Power1
- Device Alarm System1
- Mechanical Problem1
- Sharp Edges1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Abrasion1
- Bruise/Contusion1
- Laceration(s)1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Leak/Splash
The user facility reported that their AMSCO 7052 HP Washer/Disinfector was leaking water.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946843 ↗InjuryAug 13, 2026Mechanical Problem
Patient problems: Abrasion; Bruise/Contusion
MAUDE report 50239193 ↗MalfunctionMar 26, 2026Use of Device Problem
WHILE ONSITE AT THE USER FACILITY, A STERIS INSTALLATION SPECIALIST OBSERVED THAT USER FACILITY PERSONNEL WERE USING A DETERGENT DURING CYCLES WHICH DID NOT ALIGN WITH THE DETERGENT PARAMETERS SET ON THEIR AMSCO 7052 HP WASHER/DISINFECTOR. IT IS UNKNOWN HOW MANY INSTRUMENTS WERE PROCESSED WITH THIS DETERGENT. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24697986 ↗MalfunctionApr 30, 2025Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR AMSCO 7052HP WASHER-DISINFECTOR ONTO THE FLOOR. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21936629 ↗MalfunctionOct 17, 2024Device Alarm System; Device Alarm System
THE USER FACILITY REPORTED THE RINSE RESERVOIR OVERTEMPERATURE ALARM TO THEIR AMSCO 7052HP WASHER-DISINFECTOR WAS "TRIPPED" AND NOT WORKING PROPERLY. THE UNIT WAS REMOVED FROM SERVICE TO AWAIT ONSITE INSPECTION; SUBSEQUENTLY ATTRIBUTING TO A PROCEDURE DELAY. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20475835 ↗MalfunctionOct 7, 2024Sharp Edges; Sharp Edges
THE USER FACILITY REPORTED THAT AN EMPLOYEE RECEIVED A CUT TO THEIR HAND WHILE CLEANING THE DOOR GASKET OF THEIR AMSCO 7052HP WASHER/DISINFECTOR. THE EMPLOYEE SOUGHT AND RECEIVED MEDICAL TREATMENT (STITCHES).
Patient problems: Laceration(s)
MAUDE report 20382795 ↗MalfunctionMar 28, 2024Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED A WATER LEAK FROM A AMSCO 7052 HP WASHER. THERE WAS NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18993055 ↗MalfunctionAug 9, 2023Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT THEY FOUND WATER UNDER THEIR AMSCO 7052HP WASHER/DISINFECTOR AND THE WATER WAS LEAKING ONTO THE FLOOR BELOW. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17488075 ↗MalfunctionJul 26, 2023Complete Loss of Power; Complete Loss of Power
THE USER FACILITY REPORTED THAT THE POWER SUPPLY OF THEIR AMSCO 7052HP WASHER/DISINFECTOR STOPPED WORKING RESULTING IN PROCEDURE CANCELLATIONS. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17396621 ↗
Other Disinfector, medical devices devices
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls6 reports
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls0 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls3 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls1 report
- GETINGE 86-SERIES S-8668Getinge Disinfection AB · Disinfector, Medical Devices1 recall37 reports
- Getinge Aquadis 56 56MGetinge Disinfection AB · Disinfector, Medical Devices1 recall11 reports
- AMSCO 7052HPCorporation Steris Canada · Disinfector, medical devices1 recall10 reports
- GETINGE 9100-SERIES 9128EGetinge Disinfection AB · Disinfector, Medical Devices1 recall10 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.