Flexor recallZ-0993-2026 · Cook Incorporated · initiated Nov 26, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
FDA root cause category: Process control
What the company is doing
Cook Medical issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 11/26/2025 by email and/or letter delivered via a courier. The notice explained the issue, risk to health, and requested the identification and return of the affected product. The notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. For questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Recall details
- Recall number
- Z-0993-2026
- Status
- Open, Classified
- Initiated by the company
- Nov 26, 2025
- Posted by the FDA
- Dec 23, 2025
- Recalling firm
- Cook Incorporated
- Quantity
- 7952 units
- Product code
- DYB
- FDA submissions
- K142829
- Distribution
- Worldwide distribution.
Product description
Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-ANL1-HC; 3) KCFW-10.0-35-45-RB-HFANL0-HC; 4) KCFW-10.0-35-45-RB-HFANL1-HC; 5) KCFW-12.0-35-45-RB-HFANL0-HC; 6) KCFW-12.0-35-45-RB-HFANL1-HC; 7) KCFW-9.0-35-38.5-RB-RTPS-100; 8) KCFW-10.0-35-RB-RTPS-100; 9) KCFW-9.0-38-55-RB-RAABE; 10) KCFW-9.0-38-70-RB-RAABE; 11) KCFW-9.0-38-80-RB; 12) KCFW-10.0-38-30-RB; 13) KCFW-10.0-38-40-RB; 14) KCFW-10.0-38-80-RB; 15) KCFW-10.0-38-RB; 16) KCFW-12.0-38-40-RB; 17) KCFW-12.0-38-80-RB; Flexor Introducers and Guiding Sheaths are designed to perform as a guiding sheath and/or introducer sheath. Flexor Introducers and Guiding Sheaths incorporate a Flexor shaft with a hemostasis valve and are provided with one or more dilators. These devices are available in various sizes, lengths and configurations. Configurations include differences in shaft construction, such as varying shaft stiffness and distal tip material and shape, dilator material, hydrophilic coating and/or distal radiopaque markers.
Code information (models, lots, serial numbers)
Show the full code information (16k characters)
Reference Part Number (RPN)KCFW-9.0-18/38-45-RB-ANL0-HC: Order Number G49101, UDI (01)00827002491014(17)280708(10)16683401, Lot Number 16683401; Reference Part Number (RPN)KCFW-9.0-18/38-45-RB-ANL1-HC: Order Number G49102, UDI (01)00827002491021(17)280709(10)16685978, Lot Number 16685978; Reference Part Number (RPN)KCFW-9.0-18/38-45-RB-ANL1-HC: Order Number G49102, UDI (01)00827002491021(17)280804(10)NS16724225, Lot Number NS16724225; Reference Part Number (RPN)KCFW-9.0-18/38-45-RB-ANL1-HC: Order Number G49102, UDI (01)00827002491021(17)280808(10)NS16733075, Lot Number NS16733075; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280619(10)16652188, Lot Number 16652188; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280620(10)16653824, Lot Number 16653824; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280626(10)16662713, Lot Number 16662713; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280626(10)16662715, Lot Number 16662715; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280701(10)16675034, Lot Number 16675034; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280716(10)16696277, Lot Number 16696277; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280721(10)16703198, Lot Number 16703198; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280722(10)16704975, Lot Number 16704975; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280724(10)16709796, Lot Number 16709796; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280731(10)16720389, Lot Number 16720389; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL0-HC: Order Number G56278, UDI (01)00827002562783(17)280617(10)16647121, Lot Number 16647121; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL1-HC: Order Number G47700, UDI (01)00827002477001(17)280723(10)16707723, Lot Number 16707723; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL1-HC: Order Number G47700, UDI (01)00827002477001(17)280806(10)16727474, Lot Number 16727474; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL1-HC: Order Number G47700, UDI (01)00827002477001(17)280604(10)16629368, Lot Number 16629368; Reference Part Number (RPN)KCFW-10.0-35-45-RB-HFANL1-HC: Order Number G47700, UDI (01)00827002477001(17)280627(10)16664784, Lot Number 16664784; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL0-HC: Order Number G56279, UDI (01)00827002562790(17)280714(10)NS16692276, Lot Number NS16692276; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL0-HC: Order Number G56279, UDI (01)00827002562790(17)280610(10)16637317, Lot Number 16637317; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL0-HC: Order Number G56279, UDI (01)00827002562790(17)280618(10)16649121, Lot Number 16649121; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL0-HC: Order Number G56279, UDI (01)00827002562790(17)280729(10)16716240, Lot Number 16716240; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL0-HC: Order Number G56279, UDI (01)00827002562790(17)280523(10)16614648, Lot Number 16614648; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL0-HC: Order Number G56279, UDI (01)00827002562790(17)280611(10)16638978, Lot Number 16638978; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280717(10)16699013, Lot Number 16699013; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280527(10)16616270, Lot Number 16616270; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280609(10)16635801, Lot Number 16635801; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280620(10)16653825, Lot Number 16653825; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280627(10)16664786, Lot Number 16664786; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280702(10)16677072, Lot Number 16677072; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280728(10)16712983, Lot Number 16712983; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280728(10)16712985, Lot Number 16712985; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280731(10)16720393, Lot Number 16720393; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280813(10)16739603, Lot Number 16739603; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280520(10)16607958, Lot Number 16607958; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280605(10)16632678, Lot Number 16632678; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280702(10)16677073, Lot Number 16677073; Reference Part Number (RPN)KCFW-12.0-35-45-RB-HFANL1-HC: Order Number G47701, UDI (01)00827002477018(17)280711(10)16690408, Lot Number 16690408; Reference Part Number (RPN)KCFW-9.0-35-38.5-RB-RTPS-100: Order Number G13160, UDI (01)00827002131606(17)280721(10)NS16703184, Lot Number NS16703184; Reference Part Number (RPN)KCFW-9.0-35-38.5-RB-RTPS-101: Order Number G13160, UDI (01)00827002131606(17)280723(10)16707656, Lot Number 16707656; Reference Part Number (RPN)KCFW-9.0-35-38.5-RB-RTPS-102: Order Number G13160, UDI (01)00827002131606(17)280609(10)16635746, Lot Number 16635746; Reference Part Number (RPN)KCFW-10.0-35-RB-RTPS-100: Order Number G13081, UDI (01)00827002130814(17)280612(10)16641423, Lot Number 16641423; Reference Part Number (RPN)KCFW-10.0-35-RB-RTPS-100: Order Number G13081, UDI (01)00827002130814(17)280618(10)16649110, Lot Number 16649110; Reference Part Number (RPN)KCFW-9.0-38-55-RB-RAABE: Order Number G36393, UDI (01)00827002363939(17)280520(10)16607912, Lot Number 16607912; Reference Part Number (RPN)KCFW-9.0-38-70-RB-RAABE: Order Number G12606, UDI (01)00827002126060(17)280711(10)16690398, Lot Number 16690398; Reference Part Number (RPN)KCFW-9.0-38-70-RB-RAABE: Order Number G12606, UDI (01)00827002126060(17)280721(10)16703171, Lot Number 16703171; Reference Part Number (RPN)KCFW-9.0-38-70-RB-RAABE: Order Number G12606, UDI (01)00827002126060(17)280529(10)16621284, Lot Number 16621284; Reference Part Number (RPN)KCFW-9.0-38-70-RB-RAABE: Order Number G12606, UDI (01)00827002126060(17)280620(10)16653803, Lot Number 16653803; Reference Part Number (RPN)KCFW-9.0-38-70-RB-RAABE: Order Number G12606, UDI (01)00827002126060(17)280523(10)16614610, Lot Number 16614610; Reference Part Number (RPN)KCFW-9.0-38-80-RB: Order Number G09442, UDI (01)00827002094420(17)280723(10)NS16707657, Lot Number NS16707657; Reference Part Number (RPN)KCFW-9.0-38-80-RB: Order Number G09442, UDI (01)00827002094420(17)280515(10)NS16602824, Lot Number NS16602824; Reference Part Number (RPN)KCFW-9.0-38-80-RB: Order Number G09442, UDI (01)00827002094420(17)280609(10)NS16635753, Lot Number NS16635753; Reference Part Number (RPN)KCFW-9.0-38-80-RB: Order Number G09442, UDI (01)00827002094420(17)280611(10)NS16638955, Lot Number NS16638955; Reference Part Number (RPN)KCFW-10.0-38-30-RB: Order Number G13032, UDI (01)00827002130326(17)280729(10)16716200, Lot Number 16716200; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280520(10)16607936, Lot Number 16607936; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280611(10)16638956, Lot Number 16638956; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280610(10)16637278, Lot Number 16637278; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280610(10)16637282, Lot Number 16637282; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280616(10)16644768, Lot Number 16644768; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280617(10)16647072, Lot Number 16647072; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280617(10)16647073, Lot Number 16647073; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280624(10)16657904, Lot Number 16657904; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280624(10)16657905, Lot Number 16657905; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280630(10)16673098, Lot Number 16673098; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280630(10)16673112, Lot Number 16673112; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280630(10)16673113, Lot Number 16673113; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280710(10)16688777, Lot Number 16688777; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280710(10)16688778, Lot Number 16688778; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280711(10)16690378, Lot Number 16690378; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280715(10)16693991, Lot Number 16693991; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280722(10)16704894, Lot Number 16704894; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280722(10)16704906, Lot Number 16704906; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280724(10)16709786, Lot Number 16709786; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280724(10)16709787, Lot Number 16709787; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280729(10)16716212, Lot Number 16716212; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280729(10)16716213, Lot Number 16716213; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280801(10)16721859, Lot Number 16721859; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280801(10)16721860, Lot Number 16721860; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280801(10)16721868, Lot Number 16721868; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280801(10)16721875, Lot Number 16721875; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280801(10)16721876, Lot Number 16721876; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280515(10)16602829, Lot Number 16602829; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280520(10)16607937, Lot Number 16607937; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280523(10)16614632, Lot Number 16614632; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280530(10)16623383, Lot Number 16623383; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280605(10)16632466, Lot Number 16632466; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280610(10)16637284, Lot Number 16637284; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280611(10)16638969, Lot Number 16638969; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280630(10)16673114, Lot Number 16673114; Reference Part Number (RPN)KCFW-10.0-38-40-RB: Order Number G32233, UDI (01)00827002322332(17)280613(10)16643080, Lot Number 16643080; Reference Part Number (RPN)KCFW-10.0-38-80-RB: Order Number G13033, UDI (01)00827002130333(17)280529(10)16621305, Lot Number 16621305; Reference Part Number (RPN)KCFW-10.0-38-80-RB: Order Number G13033, UDI (01)00827002130333(17)280523(10)16614616, Lot Number 16614616; Reference Part Number (RPN)KCFW-10.0-38-80-RB: Order Number G13033, UDI (01)00827002130333(17)280610(10)NS16637285, Lot Number NS16637285; Reference Part Number (RPN)KCFW-10.0-38-80-RB: Order Number G13033, UDI (01)00827002130333(17)280617(10)NS16647074, Lot Number NS16647074; Reference Part Number (RPN)KCFW-10.0-38-80-RB: Order Number G13033, UDI (01)00827002130333(17)280729(10)NS16716215, Lot Number NS16716215; Reference Part Number (RPN)KCFW-10.0-38-RB: Order Number G13031, UDI (01)00827002130319(17)280613(10)16643081, Lot Number 16643081; Reference Part Number (RPN)KCFW-10.0-38-RB: Order Number G13031, UDI (01)00827002130319(17)280715(10)16693981, Lot Number 16693981; Reference Part Number (RPN)KCFW-10.0-38-RB: Order Number G13031, UDI (01)00827002130319(17)280523(10)16614624, Lot Number 16614624; Reference Part Number (RPN)KCFW-10.0-38-RB: Order Number G13031, UDI (01)00827002130319(17)280611(10)16638970, Lot Number 16638970; Reference Part Number (RPN)KCFW-12.0-38-40-RB: Order Number G32373, UDI (01)00827002323735(17)280605(10)16632469, Lot Number 16632469; Reference Part Number (RPN)KCFW-12.0-38-40-RB: Order Number G32373, UDI (01)00827002323735(17)280708(10)16683354, Lot Number 16683354; Reference Part Number (RPN)KCFW-12.0-38-40-RB: Order Number G32373, UDI (01)00827002323735(17)280619(10)16652152, Lot Number 16652152; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280806(10)16727445, Lot Number 16727445; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280529(10)16621277, Lot Number 16621277; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280603(10)16627324, Lot Number 16627324; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280529(10)16621278, Lot Number 16621278; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280731(10)16720362, Lot Number 16720362; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280522(10)16613273, Lot Number 16613273; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280523(10)16614607, Lot Number 16614607; Reference Part Number (RPN)KCFW-12.0-38-80-RB: Order Number G49487, UDI (01)00827002494879(17)280528(10)16618070, Lot Number 16618070.
Devices named in this recall
17 devices in the FDA device registry whose identifier appears in the recall's code information.
- Flexor G47701Cook Incorporated · INTRODUCER, CATHETER1 recall37 reports
- Flexor G56279Cook Incorporated · INTRODUCER, CATHETER1 recall12 reports
- Flexor G13081Cook Incorporated · INTRODUCER, CATHETER1 recall9 reports
- Flexor G32233Cook Incorporated · INTRODUCER, CATHETER1 recall7 reports
- Flexor G36393Cook Incorporated · INTRODUCER, CATHETER1 recall4 reports
- Flexor G12606Cook Incorporated · INTRODUCER, CATHETER1 recall3 reports
- Flexor G13033Cook Incorporated · INTRODUCER, CATHETER1 recall3 reports
- Flexor G32373Cook Incorporated · INTRODUCER, CATHETER1 recall3 reports
- Flexor G49101Cook Incorporated · INTRODUCER, CATHETER1 recall3 reports
- Flexor G13160Cook Incorporated · INTRODUCER, CATHETER1 recall2 reports
- Flexor G09442Cook Incorporated · INTRODUCER, CATHETER1 recall1 report
- Flexor G13032Cook Incorporated · INTRODUCER, CATHETER1 recall1 report
- Flexor G47700Cook Incorporated · INTRODUCER, CATHETER1 recall1 report
- Flexor G49487Cook Incorporated · INTRODUCER, CATHETER1 recall1 report
- Flexor G56278Cook Incorporated · INTRODUCER, CATHETER1 recall1 report
- Flexor G13031Cook Incorporated · INTRODUCER, CATHETER1 recall0 reports
- Flexor G49102Cook Incorporated · INTRODUCER, CATHETER1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98027).
- Z-0994-2026 · Open, ClassifiedRing Transjugular Intrahepatic Access Sets, Reference Part Numbers: 1) RTPS-100; 2) RTPS-100-10.0; The Ring Transjugular Intrahepatic Access Set is intended for transjugular li…
- Z-0995-2026 · Open, ClassifiedRsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnosti…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗