Skip to content

Flexor G32233Recalls & reported problems

Cook Incorporated · Flexor, Check-Flo, Introducer Set · Class 2

All models of Flexor (134)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002322332), received between Apr 13, 2016 and Nov 11, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 26, 2025Class IIOpen, Classified

    Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.

    Z-0993-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Unraveled Material2
  • Break1
  • Component Missing1
  • Device Contaminated During Manufacture or Shipping1
  • Material Separation1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    AS REPORTED, DURING AN UNKNOWN PROCEDURE, A HAIR WAS DISCOVERED INSIDE THE FLEXOR GUIDING SHEATH'S SEALED PACKAGE. THE PROCEDURE WAS NOT DELAYED AS ANOTHER OF THE SAME-TYPE SHEATH WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20654073 ↗
  • MalfunctionDec 18, 2023Break; Peeled/Delaminated; Unraveled Material; Break; Peeled/Delaminated; Unraveled Material

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18350072 ↗
  • MalfunctionMar 29, 2022Unraveled Material; Unraveled Material

    AS REPORTED, AT THE END OF A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) PROCEDURE, A FLEXOR GUIDING SHEATH UNRAVELED UPON REMOVAL OF THE DEVICE. THE ANATOMY WAS NOT TORTUOUS, CALCIFIED, OR SCARRED. RESISTANCE WAS ENCOUNTERED UPON REMOVAL OF THE DEVICE FROM THE PATIENT, AND THE SHEATH'S INNER COIL UNRAVELED. THE DEVICE, WHICH DID NOT FULLY SEPARATE, WAS REMOVED FROM THE PATIENT INTACT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13940744 ↗
  • MalfunctionJun 1, 2021Fluid/Blood Leak; Fluid/Blood Leak

    NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11913722 ↗
  • MalfunctionFeb 14, 2020Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED, DURING A LIVER BIOPSY VIA TRANSJUGULAR ACCESS, THE HUB OF A FLEXOR GUIDING SHEATH LEAKED. AN UNKNOWN WIRE WAS IN THE LUMEN OF THE DEVICE AT THE TIME THE LEAK OCCURRED. ANOTHER SHEATH WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9711114 ↗
  • MalfunctionJan 31, 2017Material Separation; Material Separation

    CUSTOMER REPORTED THAT A MALE PATIENT UNDERWENT A TRANSJUGULAR LIVER BIOPSY. AFTER INITIAL PLACEMENT OF THE DEVICE, THE ATTENDING PHYSICIAN RETRACTED THE DEVICE AND ATTEMPTED TO CURVE THE TIP. THE DILATOR WAS IN THE SHEATH WHILE CURVING THE DEVICE. THE DISTAL TIP OF THE DILATOR SEPARATED WHILE IT WAS OUTSIDE OF THE PATIENTS BODY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6291275 ↗
  • MalfunctionApr 13, 2016Component Missing; Component Missing

    A PATIENT OF UNKNOWN GENDER AND AGE WAS UNDERGOING A TIPS PROCEDURE. UTILIZING INTERMITTENT FLUOROSCOPIC GUIDANCE, THE SHEATH WAS ADVANCED INTO THE RIGHT ATRIUM. AT THIS POINT IT WAS NOTED THAT THE SHEATH DID NOT HAVE A RADIOPAQUE BAND TO MARK WHERE THE SHEATH ENDED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5573157 ↗

FDA clearance and classification

The FDA regulates product code DYB (introducer, catheter) under 21 CFR 870.1340 as a Class II device. 21 CFR 870.1340 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Flexor G32233

Has the Flexor G32233 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0993-2026 (Class II), began on Nov 26, 2025: Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Flexor G32233?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Apr 13, 2016 and Nov 11, 2024. In those reports, the most common specific device problems are “Fluid/Blood Leak” (2), “Unraveled Material” (2) and “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Flexor G32233?

None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Flexor G32233, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Flexor, model G32233). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page