Vantage Titan 3t recallZ-1018-2026 · Canon Medical System, USA, INC. · initiated Dec 19, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
FDA root cause category: Under Investigation by firm
What the company is doing
On 12/19/2025, the firm sent via Adobe DocuSign an "Urgent: Medical Device Correction" letter to customers informing them that It has been discovered that in the superconducting magnet used in the applicable system, if there is slow leakage, over a long period of time, of refrigerant gas from the superconducting magnet, air ingress into the helium vessel via the leak point may result in ice forming. If this is left unchecked, the worst-case scenario is that the helium gas will not be able to escape through the designed exhaust path, and the helium tank could eventually rupture, releasing helium gas into the MR scan room. For safety management of the superconducting magnet, the status of the magnet is checked during maintenance inspections and continuously monitored by a remote monitoring system. However, in response to the above report from the manufacturer of the superconducting magnet, an inspection to check for slow gas leaks in the superconducting magnet used in systems installed at our customer sites will be provided as a repair service. Note that there have been no reports of health hazards related to this incident. Customer are instructed to: The system can continue to be used. However, please ensure there are adequate evacuation routes from the scan/imaging room and the control room. Additionally, do not press the quench button in cases other than emergencies. A representative of Canon Medical Systems USA will contact you to schedule an inspection. If you have any questions regarding this matter, please contact your service representative. Share the contents of letter with all users and reviewing radiologist as well as clinical engineering or biomedical group at their facility. In addition, posted the Recall Letter on or near potentially affected systems. For any questions regarding this letter please contact Orlando Tadeo, Director Regulatory Affairs, at 800-421-1968 or [email protected] or your Canon service representative at 800-521-1968.
Recall details
- Recall number
- Z-1018-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 19, 2025
- Posted by the FDA
- Jan 13, 2026
- Recalling firm
- Canon Medical System, USA, INC.
- Quantity
- 22 systems
- Product code
- LNH
- FDA submissions
- K191128
- Distribution
- U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Product description
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Number: MRT-3010/MEXL-3010 UDI-DI code: 04987670101649 CMSC Model Number/Model Number/Serial Numbers/SID/UDI code: MRT-3010/A5 MEXL-3010/A5 A5D14X2031 30000041 MRT-3010/A5 MEXL-3010/A5 A5D13Y2025 30001708 MRT-3010/A5 MEXL-3010/A5 A5D13Y2026 30005894 MRT-3010/A5 MEXL-3010/A5 A5G1692039 30007620 (01)04987670101649(21)A5G1692039 MRT-3010/A5 MEXL-3010/A5.007 A5B11Y2006 30019175 MRT-3010/A7 MEXL-3010/S7 S7A1642006 30024266 (01)04987670101649(21)A7A1642001 MRT-3010/A5 MEXL-3010/A5 A5G1692038 30027318 (01)04987670101649(21)A5G1692038 MRT-3010/A5 MEXL-3010/A5 A5G1642036 30029482 (01)04987670101649(21)A5G1642036 MRT-3010/A5 MEXL-3010/A5.007 A5F1562034 30050314 MRT-3010/A5 MEXL-3010/A5 A5B11Z2007 303354 MRT-3010/A5 MEXL-3010/A5 A5C1252012 336719 MRT-3010/A5 MEXL-3010/A5 A5D1312019 350553 MRT-3010/A5 MEXL-3010/A5 A5B1192003 350579 MRT-3010/A5 MEXL-3010/A5 A5C12X2016 350918 MRT-3010/A5 MEXL-3010/A5 A5C12Y2017 351122 MRT-3010/A5 MEXL-3010/A5 A5C1212009 351387 MRT-3010/A5 MEXL-3010/A5 A5B11Y2005 351452 MRT-3010/A5 MEXL-3010/A5 A5C1262013 356865 MRT-3010/A5 MEXL-3010/A5 A5C1292015 378015 MRT-3010/A5 MEXL-3010/A5 A5C1412028 383946 MRT-3010/A5 MEXL-3010/A5 A5E1522033 389630 MRT-3010/A5 MEXL-3010/A5 A5D1362021 390397
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vantage Titan 3T MRT-3010Canon Inc. · System, nuclear magnetic resonance imaging2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98112).
- Z-1017-2026 · Open, ClassifiedVANTAGE GALAN 3T Model MRT-3020/MEXL-3020
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page