Vantage Titan 3T MRT-3010Recalls & reported problems

Canon Inc. · Vantage Titan 3T · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K113093, K120487, K132160, K143008, K152371, K172878, K191128); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Dec 19, 2025Open, Classified

    There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

    Z-1018-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 10, 2024Open, Classified

    The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

    Z-2041-2024 · Root cause: Device Design · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113093, K120487, K132160, K143008, K152371, K172878, K191128. One submission often covers many models, so check the product and lot details in each record.

  • Z-2046-2021 · May 21, 2021 · Open, Classified
    Canon Vantage Titan 3T, Premium Open-Bore MRI System, REF: MRT-3010 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagitta
  • Z-2135-2016 · Apr 14, 2016 · Terminated
    MEXL-3010/A5- Vantage Titan 3T Magnetic Resonance Imaging System Product Usage: Vantage Titan 3T (MRT-3010, M-Power V3.5) is an open-bore 3T MRI system that permits outstanding im
  • Z-1101-2013 · Oct 15, 2012 · Terminated
    MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page