Ingenia recallZ-1212-2026 · Philips North America · initiated Dec 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
FDA root cause category: Software design
What the company is doing
Philips began notification of consignees on about 12/03/2025 via mailed Urgent Medical Device Correction letter, with a corrected notification sent on about 12/18/2025. Consignees were instructed to identify if any affected units were on hand, that the systems could continue to be use in accordance with their intended use. They were asked to provide the notification to all users of the device for awareness, maintain the Urgent Medical Device Correction letter with affected systems until the software upgrade is installed, and complete and return the provided response form. A Philips representative will contact each consignee to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issues beginning in January 2026 (reference FCO78100585, FCO78100620).
Recall details
- Recall number
- Z-1212-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 3, 2025
- Posted by the FDA
- Jan 29, 2026
- Recalling firm
- Philips North America
- Quantity
- 31 units
- Product code
- LNH
- FDA submissions
- K193215, K230972
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.
Product description
Ingenia Elition X. Product Code (REF): (1) 781358, (2) 782107, (3) 782136. MR systems with SW version R11.1 and R12.1.
Code information (models, lots, serial numbers)
MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781358; UDI-DI: 00884838088115; Serial Numbers: 45414, 45245, 45090, 45049, 45034, 45250, 45075, 45290, 45320, 45011, 45191, 45277, 45278, 45028, 45147. (2) Product Code (REF): 782107; UDI-DI: 00884838098336; Serial Numbers: 62031, 45533, 62051, 62182, 62024, 62047, 62048, 62052, 45519, 62181, 62081, 62153, 62088, 45589, 62073. (3) Product Code (REF): 782136; UDI-DI: 00884838108608; Serial Numbers: 28503.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls5 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging10 recalls1 report
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98111).
- Z-1204-2026 · Open, ClassifiedEvolution Upgrade 1.5T. Product Codes (REF): (1) 782148, (2) 782116. MR systems with SW version R11.1 and R12.1.
- Z-1205-2026 · Open, ClassifiedEvolution Upgrade 3.0T. Product Code (REF): 782143. MR systems with SW version R11.1 and R12.1.
- Z-1206-2026 · Open, ClassifiedIngenia 1.5T. Product Code (REF): (1) 781341, (2) 781396. MR systems with SW version R11.1 and R12.1.
- Z-1207-2026 · Open, ClassifiedIngenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R12.1.
- Z-1208-2026 · Open, ClassifiedIngenia 3.0T CX. Product Code (REF): 781271. MR systems with SW version R11.1 and R12.1.
- Z-1209-2026 · Open, ClassifiedIngenia Ambition S. Product Code (REF): 782108. MR systems with SW version R11.1 and R12.1.
- Z-1210-2026 · Open, ClassifiedIngenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.
- Z-1211-2026 · Open, ClassifiedIngenia Elition S. Product Code (REF): (1) 781357, (2) 782106, (3)782137. MR systems with SW version R11.1 and R12.1.
- Z-1213-2026 · Open, ClassifiedMR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.
- Z-1214-2026 · Open, ClassifiedUpgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.
- Z-1215-2026 · Open, ClassifiedSmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.
- Z-1216-2026 · Open, ClassifiedSmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page