Ingenia Ingenia Elition X (REF 782107)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Ingenia Elition X · Class 2

11 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
11

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838098336), received between Jun 19, 2025 and Feb 4, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).

11 FDA recall records name this device's identifier, 11 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 28, 2026Open, Classified

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

    Z-3311-2026 · Root cause: Software design · FDA record ↗

  2. Aug 21, 2026Open, Classified

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

    Z-3239-2026 · Root cause: Software design · FDA record ↗

  3. Apr 14, 2026Open, Classified

    The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

    Z-1963-2026 · Root cause: Under Investigation by firm · FDA record ↗

  4. Dec 3, 2025Open, Classified

    The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

    Z-1212-2026 · Root cause: Software design · FDA record ↗

  5. Jun 30, 2025Open, Classified

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

    Z-2211-2025 · Root cause: Device Design · FDA record ↗

  6. May 13, 2025Open, Classified

    Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

    Z-1803-2025 · Root cause: Software design · FDA record ↗

  7. Sep 3, 2024Open, Classified

    Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

    Z-3200-2024 · Root cause: Under Investigation by firm · FDA record ↗

  8. Mar 12, 2024Open, Classified

    Patient support table floor plate may be incorrectly installed.

    Z-1571-2024 · Root cause: Process design · FDA record ↗

  9. Jun 12, 2023Open, Classified

    The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.

    Z-2395-2023 · Root cause: Under Investigation by firm · FDA record ↗

  10. Jun 12, 2023Open, Classified

    During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.

    Z-2398-2023 · Root cause: Under Investigation by firm · FDA record ↗

  11. Nov 22, 2022Open, Classified

    Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.

    Z-0460-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information5
  • Improper or Incorrect Procedure or Method3
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Eye Injury1
  • Full thickness (Third Degree) Burn1
  • Partial thickness (Second Degree) Burn1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 4, 2026Improper or Incorrect Procedure or Method; Insufficient Device Problem Information

    PHILIPS RECEIVED A REPORT THAT AFTER THE MRI EXAMINATION WAS COMPLETED AND THE BED WAS PULLED OUT, SLEEPING PATIENT WOKE UP AND RAISED HIS HEAD WHILE THE IN-BORE MIRROR WAS STILL ATTACHED, CAUSING CONTACT WITH THE MIRROR AND INJURING HIS RIGHT EYE. THE PATIENT WAS SEEN BY A PROVIDER IN THE FACILITY AND WAS PRESCRIBED EYE DROPS TO TREAT CONJUNCTIVAL HEMORRHAGE IN THE RIGHT EYE. EYE DROPS PRESCRIBED BY A PHYSICIAN MEET THE CRITERIA FOR MEDICAL INTERVENTION TO PRECLUDE PERMANENT IMPAIRMENT. THE PATIENT WAS LATER OBSERVED AT OPHTHALMOLOGY DEPARTMENT. IT WAS REPORTED, THE PATIENT DID NOT HAVE VISION ABNORMALITIES. THE PATIENT CONDITION IS CURRENTLY UNDER OBSERVATION AFTER BEING EXAMINED BY AN OPHTHALMOLOGIST. THE REPORTED INJURY MEETS THE CRITERIA FOR A SERIOUS INJURY.

    Patient problems: Eye Injury

    MAUDE report 24255730 ↗
  • MalfunctionOct 28, 2025Improper or Incorrect Procedure or Method; Insufficient Device Problem Information; Improper or Incorrect Procedure or Method; Insufficient Device Problem Information

    PHILIPS RECEIVED A REPORT THAT THE PATIENT EXPERIENCED A HEATING SENSATION, SKIN REDDENING AND BLISTERING WAS NOTED ON THE RIGHT SHOULDER AREA. THE BLISTER AREA IS DESCRIBED AS TOP OF RIGHT SHOULDER AND SHOULDER BLADE/SCAPULA AREA. SIZE APPROXIMATELY 0.5-1 IN. MINOR WITH REDNESS AND POTENTIAL BLISTER. CUSTOMER QUESTIONED BUT UNKNOWN REGARDING BLISTER. THE PATIENT WAS SENT TO ONSITE CLINIC AND SENT HOME. NO FOLLOW-UP REQUIRED. THE BURN IS EXPECTED TO HEAL. THE INJURY REPORTED MEETS THE CRITERIA FOR A NON-SERIOUS MINOR INJURY.

    Patient problems: Partial thickness (Second Degree) Burn; Superficial (First Degree) Burn

    MAUDE report 23400852 ↗
  • MalfunctionSep 17, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    PHILIPS RECEIVED A REPORT ABOUT BURN MARKS ON AXIS Z OF GRADIENT COIL. DURING SCAN, AN USER MESSAGE DISPLAYED SMOKE DETECTOR ALARM. UPON OPENING THE REAR COVER, A VISUAL INSPECTION REVEALED BURN MARKS ON THE AXIS Z BUSBAR CONNECTED TO THE GRADIENT COIL. NO HARM WAS REPORTED RELATED TO THIS INCIDENT. BASED ON THIS EVIDENCE IT WAS DECIDED TO REPORT THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23071146 ↗
  • MalfunctionSep 16, 2025Sparking; Insufficient Device Problem Information; Sparking; Insufficient Device Problem Information

    PHILIPS RECEIVED A REPORT ABOUT BROKEN BUSBAR. DURING SCANNING A GRADIENT ERROR CAUSED AN INTERLOCK. BASED ON THIS EVIDENCE, THE INCIDENT WAS REPORTED. IT WAS FOUND THAT THE BUSBAR OF Z AXIS WAS BURNED AND BROKEN. NO HARM WAS REPORTED RELATED TO THIS INCIDENT. BASED ON THIS EVIDENCE IT WAS DECIDED TO REPORT THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23060857 ↗
  • InjuryJun 19, 2025Improper or Incorrect Procedure or Method; Insufficient Device Problem Information; Improper or Incorrect Procedure or Method; Insufficient Device Problem Information

    PHILIPS RECEIVED A REPORT THAT THE PATIENT SUSTAINED AN APPROXIMATE 21 CM BURN (THIRD DEGREE) TO HER LEFT HIP AREA DURING A LUMBAR EXAM IN THE HEADFIRST SUPINE POSITION. THE BURN IS DESCRIBED AS IRREGULAR CIRCULAR SHAPED OPEN AREA WOUND WITH AREAS THAT APPEAR TO BE BURNED THROUGH SEVERAL LAYERS OF SKIN. THE PATIENT HAS A HISTORY OF DIABETES, OBESITY, HYPERTENSION AND IS ON ANTICOAGULANT THERAPY. THE PATIENT IS REPORTED TO BE UNDERGOING TREATMENT OF THE BURN.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 22292180 ↗

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →