Coyote recallZ-1042-2026 · Boston Scientific Corporation · initiated Dec 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.
FDA root cause category: Process change control
What the company is doing
Boston Scientific issued an Urgent Medical Device Removal notice to its consignees on 12/3/2025 by letter delivered via a courier service. The notice explained the issue, clinical impact to the patient, and requested the following actions: "Instructions: 1. Further distribution or use of any remaining devices (Attachment 1) affected by this removal should cease immediately. Do NOT use affected devices and remove those devices from your facility s inventory. Segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. If you are a distributor, this notice must be forwarded to your customers to ensure notification of this device removal is carried out to the end-user level. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page four. 5. Return affected devices." For questions regarding this communication, please contact your local Boston Scientific representative.
Recall details
- Recall number
- Z-1042-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 3, 2025
- Posted by the FDA
- Jan 9, 2026
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 21 units
- Product code
- LIT
- FDA submissions
- K111295
- Distribution
- US Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX, FL, IL.
Product description
Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2) 3.5mm x120mm, 150 cm, UPM/Model No. REF H74939186351210; Catheter, Angioplasty, Peripheral, Transluminal
Code information (models, lots, serial numbers)
1) GTIN 08714729796527, Batch Number 37370117, exp. 3-Sept-28; 2) GTIN 08714729796756, Batch Number 37416646, exp. 9-Sept-28
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Coyote™ H74939186251210Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall4 reports
- Coyote™ H74939186351210Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page