Skip to content

Abre recallZ-1098-2026 · Medtronic Inc · initiated Dec 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.

FDA root cause category: Under Investigation by firm

What the company is doing

Beginning 04-Dec-2025, Medtronic distributed an Urgent Medical Device Recall notice to all U.S. consignees who have impacted products via mail courier. Consignees were instructed to complete and return a response form to the firm and to return any affected product they may have.

Recall details

Recall number
Z-1098-2026
Status
Open, Classified
Initiated by the company
Dec 4, 2025
Posted by the FDA
Jan 15, 2026
Recalling firm
Medtronic Inc
Quantity
442
Product code
QAN
FDA submissions
P200026
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.

Product description

Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 00763000547332 00763000547271 00763000547318 00763000547264 00643169796225 00643169796300 00643169796294 00643169796232 00643169796287 00763000547325 00643169796300 00763000547257 Serial/Lot numbers C112511 C113253 C113795 C117416 C117482 C121084 C123953 C133327 C133329 C134478 C138060 C150325

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

22 devices cleared under P200026 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.