Abre™ AB9U12060090Recalls & reported problems
Medtronic, Inc. · STENT AB9U12060090 ABRE V01 · Class 3 · Implant
All models of Abre (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169796225), received between Jul 9, 2021 and Jan 22, 2026 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (P200026); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 4, 2025Open, Classified
Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.
Z-1098-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Migration or Expulsion of Device2
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Chest Pain1
- Insufficient Information1
- Obstruction/Occlusion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 22, 2026Use of Device Problem
IT WAS REPORTED THAT AN ABRE STENT THAT EXPIRED ON 04-JAN-2026 WAS IMPLANTED IN A PATIENT ON (B)(6) 2026. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24143527 ↗InjuryNov 6, 2025Migration or Expulsion of Device; Device Dislodged or Dislocated; Migration or Expulsion of Device; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT SUPERIOR VENA CAVA STENOSIS DUE TO MALIGNANT OBSTRUCTION, THE STENT ABRE 12MM WAS DEPLOYED FROM INTERNAL JUGULAR ACCESS AFTER VESSEL MEASUREMENT WITH INTRAVASCULAR ULTRASOUND AND LOCATION CONFIRMATION WITH FLUOROSCOPY (IVUS MEASURED 9.5MM). THE VESSEL EXHIBITED 90% LESION STENOSIS, A LESION LENGTH OF 50MM, MODERATE CALCIFICATION, AND MINIMAL TORTUOSITY, WITH A REFERENCE VESSEL DIAMETER OF 9.6MM. UPON DEPLOYMENT, THE STENT IMMEDIATELY MIGRATED INTO THE RIGHT ATRIUM AND DID NOT COVER THE LESION. THE STENT WAS SUCCESSFULLY SNARED (GOSSENECK SNARE) AND REMOVED FROM THE BODY VIA FEMORAL VEIN ACCESS, AND INSPECTION CONFIRMED IT WAS FULLY INTACT. THE VESSEL WAS PRE-DILATED WITH A NON-MEDTRONIC (10MM CONQUEST) BALLOON AND POST-DILATED WITH A NON-MEDTRONIC (12MM CONQUEST) BALLOON. A NEW 14X60MM ABRE STENT WAS SUBSEQUENTLY PLACED FROM FEMORAL VEIN ACCESS WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23488068 ↗InjuryApr 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
TWO ABRE STENTS SELF-EXPANDING WERE IMPLANTED FOR THE TREATMENT OF A FIBROUS LESION IN THE LEFT COMMON ILIAC VEIN AND LEFT EXTERNAL ILIAC VEIN. THERE WAS MILD TORTUOSITY AND NO CALCIFICATION. THE DEVICES WERE PREPPED AS PER THE IFU WITH NO ISSUES IDENTIFIED. THE LESION WAS PRE-DILATED. THE PROCEDURE WAS COMPLETED WITH NO ISSUES MENTIONED. DURING A FOLLOW UP VISIT THE STENTS WERE NOTED TO BE OCCLUDED. REINTERVENTION ON THIS PATIENT WITH THE DOCTOR ON (B)(6) 2023. DURING THE CASE, THE STENTS WERE REOPENED THE BY REMOVING THE PRESENT CLOT WITH ASPIRATIONAL THROMBECTOMY AND USING A HIGH-PRESSURE BALLOON. THE STENTS WERE RE-EVALUATED THAT WERE ORIGINALLY PLACED, AND THE PHYSICIAN HAD THE FOLLOWING CONCLUSIONS: UNSURE ABOUT WHY THE STENTS WENT DOWN. STENTS APPEARED TO BE PROPERLY SIZED ACCORDING TO IVUS ¿ NO SPACE BETWEEN VESSEL WALL AND STENT, NO ¿STEP¿ INTO NATIVE VESSEL WHICH WOULD MEAN AN OVERSIZING. AFTER CLEARING STENTS OF CLOT AND REOPENING, NO COLLATERAL VESSELS APPEARED ON VENOGRAM. THIS LED THE DOCTOR TO BELIEVE THAT THERE ARE NO VESSELS THAT ARE STEALING FLOW FROM THE ILIAC VEINS AFTER THE CASE NO CLINICAL REASON COULD BE CONFIRMED WHY THE STENTS SHUT DOWN AT THAT TIME. PHYSIC
Patient problems: Obstruction/Occlusion
MAUDE report 16662501 ↗InjuryJul 9, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
PHYSICIAN USED AN ABRE SELF-EXPANDING STENT WITH A 9FR SHEATH DURING TREATMENT OF PROXIMAL LEFT COMMON ILIAC VEIN COMPRESSION, MAY-THURNER SYNDROME. MODERATE VESSEL TORTUOSTY REPORTED. HEALTH TISSUE WAS MEASURING 12 MM ON BOTH DISTAL AND PROXIMAL AREAS, WHERE AREA OF COMPRESSION HAD AVERAGE OF 8.5 MM. THERE WAS NO DAMAGE NOTED TO PACKAGING (I.E. SHELF CARTON, HOOP/TRAY). THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE AS PREPPED AS PER IFU WITHOUT ISSUE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE NOTED DURING ADVANCEMENT. THE DEVICE WAS IMPLANTED WITHOUT ANY ISSUES. PRE-DILATION WAS NOT PERFORMED BUT POST-DILATION WAS PERFORMED WITH 12 MM NON-MEDTRONIC BALLOON. NO DEVIATIONS NOTED DURING THE PROCEDURE. ONE DAY POST PROCEDURE THE PATIENT EXPERIENCED SEVERE CHEST PAINS AND HEART BURN. THE PATIENT IMMEDIATELY ATTENDED THE EMERGENCY ROOM AND ULTRASOUND REVEALED THE STENT HAD MIGRATED TO THE HEART. THE PATIENT WAS SENT TO IR TO HAVE STENT RETRIEVED. STENT RETRIEVAL WAS SUCCESSFUL. THE STENT WAS STUCK IN BETWEEN RIGHT ATRIUM AND RIGHT VENTRICLE IN THE TRICUSPID VALVE. PATIENT WILL RETURN FOR FOLLOW-UP TO SEE IF ANY DAMA
Patient problems: Chest Pain
MAUDE report 12141232 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200026. One submission often covers many models, so check the product and lot details in each record.
- Z-0464-2022 · Nov 19, 2021 · TerminatedInstructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.
- Z-1532-2021 · Mar 16, 2021 · TerminatedMedtronic Abre Venous Self-expanding Sent System, Catalog number REF AB9U16080090 - Product Usage: intended for use in the iliofemoral veins for the treatment of symptomatic venous
Other Stent, iliac vein devices
- Abre™ AB9U14080090Medtronic, Inc. · Stent, iliac vein2 recalls9 reports
- Venovo® VENUM16100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall55 reports
- Venovo® VENUM14080Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall36 reports
- Venovo® VENUM14120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall33 reports
- Venovo® VENUM14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall31 reports
- Venovo® VENUM14100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall30 reports
- Venovo® VENUM16120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall29 reports
- Venovo® VENUM14160Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page