PRISMAFLEX SET recallZ-1229-2026 · Vantive US Healthcare LLC · initiated Jan 6, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Vantive US Healthcare LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
FDA root cause category: Component design/selection
What the company is doing
Vantive issued an Urgent Medical Device Correction notice to its consignees on 01/06/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: 1. You may continue to perform therapy using the Prismaflex sets listed above. Please monitor the deaeration chamber during therapy to ensure it remains in an upright position within the holder. 2. Vantive is aware of customers who have experienced this issue and attempted to further secure the deaeration chamber in an upright position. If attempting to secure the deaeration chamber in an upright position, please consider the following: . Ensure there are no kinks in the tubing and that the deaeration chamber remains visible. . If the Prismaflex Control Unit issues an Air in Blood alarm, check if air is present. If there is no air present, check if the deaeration chamber is dislodged from the holder and ensure there are no clots present. 1. If no dislodgement, please follow the operator s manual instructions for Air in Blood alarm. 2. In case of dislodgement with clotting, follow the associated instructions within the Prismaflex Control Unit operator s manual to stop the therapy and replace the set. 3. Only in case of dislodgement without clotting or any other alarms, if the chamber is secured in an upright position, follow the associated instructions within the Prismaflex Control Unit operator s manual to proceed with treatment. Facilities who further distributed the products were directed to notify those departments or facilities. Dealers, wholesalers, distributor/resellers, or original equipment manufacturers (OEM) who further distributed the products were directed to notify their customers. For general questions regarding this communication, please contact Vantive Customer Care at 833-542-CRRT (2778). Vantive Customer Care is available 24 hours a day, seven days a week.
Recall details
- Recall number
- Z-1229-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 6, 2026
- Posted by the FDA
- Feb 2, 2026
- Recalling firm
- Vantive US Healthcare LLC
- Quantity
- 93,576 units
- Product code
- KDI
- FDA submissions
- K042938
- Distribution
- US Nationwide distribution.
Product description
PRISMAFLEX HF1000 SET, Product Code 107140; Dialyzer, High Permeability With Or Without Sealed Dialysate System
Code information (models, lots, serial numbers)
UDI/DI 07332414069254, All lots including and manufactured after 24G0034CA
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- PRISMAFLEX SET HF1000Gambro Industries · Dialyzer, high permeability with or without sealed dialysate system2 recalls41 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K042938 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- PRISMAFLEX SET HF1400Gambro Industries · Dialyzer, high permeability with or without sealed dialysate system2 recalls2 deaths32 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98180).
- Z-1224-2026 · Open, ClassifiedPRISMAFLEX ST100 SET US, Product Code 107636US; Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Z-1225-2026 · Open, ClassifiedPRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Z-1227-2026 · Open, ClassifiedPRIMSAFLEX M100 SET, Product Code 106697; Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Z-1228-2026 · Open, ClassifiedPRISMAFLEX M150 SET, Product Code 109990; Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Z-1230-2026 · Open, ClassifiedPRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Z-1231-2026 · Open, ClassifiedPRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page