PRISMAFLEX SETRecalls & reported problems across all models
Gambro Industries · 8 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
9 open recalls name at least one PRISMAFLEX SET model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 334
- Deaths reported
- 4
- Injuries reported
- 16
- Recalls
- 9
We found 8 models of PRISMAFLEX SET by Gambro Industries in the FDA's device registry (GUDID). Together they are named in 334 adverse event reports in the FDA's MAUDE database, received between Jan 3, 2018 and Sep 30, 2026 — classified as 4 deaths, 16 injuries, 314 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (221 reports).
9 FDA recall records name at least one of these models by its device identifier, 9 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
PRISMAFLEX SET recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jan 6, 2026Open, Classified
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Z-1230-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Jan 6, 2026Open, Classified
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Z-1229-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Jan 6, 2026Open, Classified
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Z-1228-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Jan 6, 2026Open, Classified
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Z-1227-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Jan 6, 2026Open, Classified
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Z-1225-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Jan 6, 2026Open, Classified
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Z-1224-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Sep 14, 2022Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Z-0083-2023 · Root cause: Error in labeling · names 2 of these models · Recall details · FDA record ↗
- Sep 14, 2022Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Z-0082-2023 · Root cause: Error in labeling · Recall details · FDA record ↗
- Sep 14, 2022Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Z-0081-2023 · Root cause: Error in labeling · names 2 of these models · Recall details · FDA record ↗
Most reported PRISMAFLEX SET problems
Device problems
- Fluid/Blood Leak221
- Crack36
- Leak/Splash30
- Break27
- Air/Gas in Device18
- Disconnection17
- Detachment of Device or Device Component13
- Adverse Event Without Identified Device or Use Problem7
Patient problems
- No Clinical Signs, Symptoms or Conditions234
- No Consequences Or Impact To Patient57
- No Patient Involvement12
- Blood Loss7
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- Air Embolism4
- Death4
- Hemorrhage/Blood Loss/Bleeding4
PRISMAFLEX SET models
8 models, most reported first. Each model has its own page with its reports, recalls and device details.
- PRISMAFLEX SET M100Dialyzer, high permeability with or without sealed dialysate system2 recalls1 death150 reports
- PRISMAFLEX SET M150Dialyzer, high permeability with or without sealed dialysate system2 recalls1 death66 reports
- PRISMAFLEX SET HF1000Dialyzer, high permeability with or without sealed dialysate system2 recalls41 reports
- PRISMAFLEX SET HF1400Dialyzer, high permeability with or without sealed dialysate system2 recalls2 deaths32 reports
- PRISMAFLEX SET 107640USDialyzer, high permeability with or without sealed dialysate system1 recall26 reports
- PRISMAFLEX SET 107636USDialyzer, high permeability with or without sealed dialysate system1 recall9 reports
- PRISMAFLEX SET M60Dialyzer, high permeability with or without sealed dialysate system1 recall9 reports
- PRISMAFLEX SET 107643USDialyzer, high permeability with or without sealed dialysate system1 report
Recent PRISMAFLEX SET adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
DeathSep 30, 2026Adverse Event Without Identified Device or Use Problem
It was reported that approximately fifteen (15) minutes after initiation of continuous renal replacement therapy (CRRT) using Prismaflex HF set, a patient experienced severe hypotension requiring up to three (3) vasopressors. Anallergic reaction was suspected. After switching to Prismaflex ST set, the hypotensive episodes reportedly did not recur. It is further reported that the patient passed away, however, the date and cause of death were not provided. No additional information is available.
Patient problems: Hypersensitivity/Allergic reaction; Low Blood Pressure/ Hypotension
MAUDE report 50951729 ↗MalfunctionAug 18, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50303544 ↗MalfunctionAug 13, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50237731 ↗MalfunctionAug 11, 2026Fluid/Blood Leak; Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50210942 ↗MalfunctionAug 6, 2026Obstruction of Flow
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 50160442 ↗MalfunctionAug 3, 2026Protective Measures Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50118813 ↗MalfunctionJun 29, 2026Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) WITH TWO (2) UNSPECIFIED TYPES OF PRISMAFLEX SETS, AN EXTERNAL FLUID LEAK WAS OBSERVED, DESCRIBED AS "THE EFFLUENT DRAINAGE LINE THAT HAD SEPARATED AT POD CAUSING THE FLUID LEAK." THE FIRST THERAPY WAS DISCONTINUED AND THE EXTRACORPOREAL BLOOD WAS RETURNED TO THE PATIENT. THE SET WAS REPLACED TO CONTINUE TREATMENT, AND THE SECOND SET LEAK WAS NOTED. CRRT WAS NOT RESTARTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25658073 ↗MalfunctionJun 22, 2026Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE EFFLUENT BAG OF A PRISMAFLEX M100 SET APPROXIMATELY THIRTY MINUTES AFTER THE START OF CONTINUOUS RENAL REPLACEMENT THERAPY. THE BAG WAS REPLACED TO CONTINUE TREATMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25580108 ↗MalfunctionJun 9, 2026Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE REPLACEMENT LINE OF A PRISMAFLEX M100 SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25447456 ↗MalfunctionMay 28, 2026Break; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED DURING CONTINUOUS RENAL REPLACEMENT THERAPY WITH A PRISMAFLEX ST150 SET. THE PRE BLOOD PUMP LUER LOCK BROKE DURING A BAG CHANGE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25316780 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT THE LUER CONNECTOR OF A PRISMAFLEX ST 150 SET WAS OBSERVED TO BE BROKEN DURING CONTINUOUS RENAL REPLACEMENT THERAPY. TREATMENT WAS DISCONTINUED WHEN "RISING FILTER PRESSURES" OCCURRED AND ON DISCONNECTION, THE ACCESS CONNECTOR WAS NOTED TO BE STUCK INSIDE THE PATIENT CATHETER. THE EXTRACORPOREAL BLOOD WAS NOT RETURNED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25316158 ↗MalfunctionMay 8, 2026Obstruction of Flow
IT WAS REPORTED THAT DURING CONTINUOUS RENAL REPLACEMENT THERAPY WITH TWO (2) PRISMAFLEX ST150 SETS AND A PRISMAFLEX MACHINE, TREATMENT WAS DISCONTINUED TWICE WHEN A "FLOW RATE ALARM PERSISTED". THE EXTRACORPOREAL BLOOD WAS NOT RETURNED TO THE PATIENT IN EITHER INSTANCE BECAUSE IT WAS "TOO COAGULATED". THERE WAS NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25126029 ↗MalfunctionMay 8, 2026Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM A PRISMAFLEX M100 SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25125119 ↗MalfunctionMay 5, 2026Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE REPLACEMENT LINE NEAR THE DEAERATION CHAMBER OF A PRISMAFLEX ST150 SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25083626 ↗MalfunctionMay 4, 2026Fluid/Blood Leak
IT WAS REPORTED THAT AN INTERNAL BLOOD LEAK WAS OBSERVED DURING CONTINUOUS RENAL REPLACEMENT THERAPY WITH FOUR (4) PRISMAFLEX HF 1400 SETS, WHICH "AFFECTED ONE PATIENT". THE SET WAS REPLACED TO CONTINUE TREATMENT WITH A SET FROM A DIFFERENT PRODUCT LOT. NO INFORMATION WAS PROVIDED TO CONFIRM WHETHER THE EXTRACORPOREAL BLOOD WAS RETURNED TO THE PATIENT. THERE WAS NO REPORT OF SERIOUS INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25072349 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.