Patient Information Center iX recallZ-1156-2026 · Philips North America · initiated Dec 22, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
FDA root cause category: Software Design Change
What the company is doing
On or about December 22, 2025 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: Review the Mobile Notification Event Filters settings and confirm all the events intended to be sent to mobile devices automatically are enabled for Mobile Notification Event Filters. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem " Philips representative will contact you to obtain PIC iX 4.5.1 SW patch for PIC iX systems running software version 4.5.0. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.
Recall details
- Recall number
- Z-1156-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 22, 2025
- Posted by the FDA
- Jan 22, 2026
- Recalling firm
- Philips North America
- Quantity
- 146 units (41 US, 105 OUS)
- Product code
- MHX
- FDA submissions
- K153702
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Utd.Arab.Emir.
Product description
Patient Information Center iX; Software Version Number: 4.5.0
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Model Numbers: (1) 866389, (2) 866390; UDI-DI: (1) 00884838127074, (2) N/A; Serial Numbers: 4Z24-5CKL-P, 0P6C-4TL3-D, 6J58-3LL6-0, 216G-7PGV-E, 0F2U-1BLC-H, 291E-7HJU-A, 1M6W-1UP5-1, 4620-0XNW-5, 4W5R-3FHA-L, 6B73-3ZKP-K, 3E79-49M4-3, 4F6Y-70HU-P, 0128-2WG5-0, 0830-3RLW-W, 097K-7LPC-5, 0D2V-1BLC-8, 120T-17M9-A, 173C-20JM-J, 190B-64MU-U, 1K2J-15LB-P, 1L3T-7KGB-E, 2K06-4ZPR-X, 3916-5DJ6-Z, 3V7H-06MT-E, 412A-2WG5-Y, 4271-2NL7-M, 4Z3L-5PP3-0, 5D0E-3EH3-3, 5E0B-3EH3-D, 5E25-14LG-9, 657X-31N5-L, 6G5U-3FHA-R, 781T-7TJA-B, 7N3U-51MJ-B, 7P31-7NPM-P, 283X-7ZPB-A, 7H77-09GU-Z, 4C59-26K6-P, 4N0L-0FKE-2, 5Y0A-02P6-L, 494H-1CJJ-3, 326G-7PGV-A, 292G-3YNZ-9, 0V2A-6DHP-0, 6354-58M8-C, 4W5B-6JK9-A, 5M7K-0VKT-T, 3401-1FHE-M, 6R30-5BNT-N, 6N2E-7YK1-1, 6R0H-6AMX-T, 107C-62PF-Y, 2X1T-7TJA-H, 1M6G-7PGV-K, 1C4Z-6KKD-H, 0B3Z-2RHR-L, 7H39-3JGK-U, 0M0V-6FKB-5, 793D-41PZ-X, 233G-46MR-V, 782Z-3RLW-D, 5N2B-5YJP-G, 0L5F-0ZNF-J, 4G4N-2LKY-K, 2M48-4XM7-Y, 133W-3UN7-B, 295N-0YG8-Y, 5D7W-00G0-1, 5R78-6CPV-0, 3C79-49M4-9, 3T24-5CKL-A, 6H74-27LY-T, 186W-1UP5-8, 0021-1DNN-P, 0247-7ELE-R, 0D5Y-32PN-A, 0J7W-31N5-2, 0R2W-1VNP-P, 0Y5Z-7FHX-N, 121L-1KG1-M, 1366-7XLV-X, 144W-7RPJ-5, 1472-5WKF-X, 1C19-1MHK-F, 1G0F-69LP-G, 1J5U-5HMV-N, 1P4F-34MB-W, 1P5C-6JK9-D, 212R-6HHJ-P, 221V-7TJA-4, 227N-7JG6-J, 232W-7DL9-G, 2A5X-2PJT-0, 2A75-27LY-3, 2F5J-0JLM-7, 2J19-1MHK-W, 2L16-59N8-M, 2M73-5WKF-U, 2N3V-6YM2-X, 2P6J-7PGV-E, 2Y5M-5AGH-E, 326W-21KC-B, 332M-47P7-Y, 3A29-0CKW-3, 3C0U-7AHH-6, 3C2U-05HF-W, 3E05-4ZPR-U, 3G6H-39G4-Y, 3H1T-22HW-5, 3H5R-5HMV-M, 3H79-09GU-5, 3J2K-2TP9-1, 3W1Y-3CH6-J, 4L7Y-00G0-F, 4M77-1HM0-G, 4N3D-20JM-K, 4R0G-60JV-1, 4V2Z-7NPM-M, 4W08-0UH7-U, 4X6C-1PJ5-J, 4Y08-0UH7-L, 500W-6WNX-A, 594D-0EG2-9, 5U0A-3WPK-C, 5W1G-50P2-N, 5Z6J-3HL1-B, 617W-31N5-F, 637V-7BG9-W, 673T-67ME-B, 686N-3NNV-6, 6D3P-4WMG-A, 6K70-3XNY-W, 6P42-35J4-V, 6U1T-75JN-Y, 771A-2JPY-0, 7F5L-73HZ-3, 7G3A-13JG-L, 7K0P-4AH2-6, 7Z2R-05HF-X, 1J1P-6PKN-T, 4R7H-7LPC-J, 282N-0WPA-X, 5G35-04K8-E, 366L-2XHV-A, 5L7D-2UGW-T, 4N0L-0FKE-2;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Patient Information Center iX 4.5Philips Medizin Systeme Böblingen GmbH · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 recall16 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.