Skip to content

Medline medical procedure convenience kits labeled as recallZ-1188-2026 · Medline Industries, LP · initiated Dec 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-1188-2026
Status
Open, Classified
Initiated by the company
Dec 12, 2025
Posted by the FDA
Jan 28, 2026
Recalling firm
Medline Industries, LP
Quantity
1,928 total
Product code
OEZ
Distribution
US Nationwide distribution.

Product description

Medline medical procedure convenience kits labeled as: 1) PACER / CATH LAB PACK-LF, Kit SKU DYNJ0972604A; 2) PACEMAKER/ICD PACK, Kit SKU DYNJ33822L; 3) VENOUS ACCESS PACK, Kit SKU DYNJ45640B; 4) PACEMAKER SET UP, Kit SKU DYNJ63377D; 5) PACK PACEMAKER CUSTOM,9909178, Kit SKU DYNJ65011D; 6) HEART BASIC OPEN PACK, Kit SKU DYNJ69143C; 7) SCC VEIN PACK, Kit SKU DYNJ69313; 8) CV PACK, Kit SKU DYNJ86007B.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Medline Kit SKU DYNJ0972604A: UDI/DI 10889942403905 (EA) 40889942403906 (CS), Lot Number 25KMD144; Medline Kit SKU DYNJ33822L: UDI/DI 10198459385582 (EA) 40198459385583 (CS), Lot Number 25KMD419; Medline Kit SKU DYNJ45640B: UDI/DI 10193489494983 (EA) 40193489494984 (CS), Lot Number 25KMD793; Medline Kit SKU DYNJ63377D: UDI/DI 10198459514579 (EA) 40198459514570 (CS), Lot Number 25KMI988; Medline Kit SKU DYNJ65011D: UDI/DI 10198459017018 (EA) 40198459017019 (CS), Lot Number 25KMH161; Medline Kit SKU DYNJ69143C: UDI/DI 10198459478505 (EA) 40198459478506 (CS), Lot Number 25LMB211; Medline Kit SKU DYNJ69313: UDI/DI 10193489931822 (EA) 40193489931823 (CS), Lot Number 25KMJ489; Medline Kit SKU DYNJ86007B: UDI/DI 10198459260452 (EA) 40198459260453 (CS), Lot Number 25KDB551.

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98226).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page