Medline medical procedure convenience kits labeled as recallZ-1193-2026 · Medline Industries, LP · initiated Dec 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1193-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2025
- Posted by the FDA
- Jan 28, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 857 units
- Product code
- LRO
- Distribution
- US Nationwide distribution.
Product description
Medline medical procedure convenience kits labeled as: 1) DRAWER 1 AIRWAY, Kit SKU ACC010270A; 2) DRAWER 2, Kit SKU ACC010690; 3) DRAWER 3 SYR/DRS/MON/NG 1 OF 3, Kit SKU ACC011073; 4) HYPOSPADIUS PACK, Kit SKU DYNJ00282Q; 5) SJ NEURO PACK RFID-LF, Kit SKU DYNJ0101360K; 6) CENTRAL LINE PACK-LF, Kit SKU DYNJ0220136S; 7) SPINE PACK-LF, Kit SKU DYNJ0376635Q; 8) ORTHO SPINAL PACK-LF, Kit SKU DYNJ0460768T; 9) W.C. MINOR LAPAROTOMY PK-LF, Kit SKU DYNJ0509186J; 10) NEURO SHUNT PACK-LF, Kit SKU DYNJ0578815W; 11) ANCILLARY SPINE PACK-LF, Kit SKU DYNJ0610290T; 12) MINOR LAPAROTOMY PACK-LF, Kit SKU DYNJ0753436P; 13) KNEE SCOPE PACK-LF, Kit SKU DYNJ0890502R; 14) PEDIATRIC PACK-LF, Kit SKU DYNJ28212L; 15) ANTERIOR CERVICAL PACK, Kit SKU DYNJ31341P; 16) PEDIATRIC MINOR PROCEDURE PACK, Kit SKU DYNJ37615J; 17) LAPAROSCOPY PACK, Kit SKU DYNJ39215N; 18) MAJOR LAP VCH, Kit SKU DYNJ39217P; 19) NEURO PACK, Kit SKU DYNJ39220N; 20) CAS HAND PK, Kit SKU DYNJ39468G; 21) RFT ISC NOBLES MINOR PACK, Kit SKU DYNJ43954B; 22) PED OPEN HEART PACK, Kit SKU DYNJ45373G; 23) MAJOR GENERAL PACK-LF, Kit SKU DYNJ51436F; 24) MINOR PK-LF, Kit SKU DYNJ51437C; 25) VEIN ABLATION & EXCISION ASC, Kit SKU DYNJ52586C; 26) SHOULDER PACK, Kit SKU DYNJ56889B; 27) LAP CHOLE PACK, Kit SKU DYNJ57206F; 28) TRANSGENDER MASTECTOMY 2, Kit SKU DYNJ58257I; 29) SHOULDER ARTH PACK, Kit SKU DYNJ58410D; 30) MAJOR PACK, Kit SKU DYNJ58528J; 31) PACK BREAST, Kit SKU DYNJ60157F; 32) CH TRANSPHENOIDAL PACK, Kit SKU DYNJ61696I; 33) RR-SPINE PACK, Kit SKU DYNJ62643B; 34) MINOR PACK, Kit SKU DYNJ64238D; 35) PBMS NECK PACK-LF, Kit SKU DYNJ64961C; 36) BREAST PACK, Kit SKU DYNJ64988C; 37) MINOR PACK 180110, Kit SKU DYNJ65447C; 38) PACK,V-P SHUNT, Kit SKU DYNJ65472C; 39) BASIC PACK, Kit SKU DYNJ65996D; 40) MINOR SS, Kit SKU DYNJ66094A; 41) CRMC ROBOTIC GENERAL PACK, Kit SKU DYNJ66613D; 42) PEDIATRIC GENERAL PACK, Kit SKU DYNJ66644F; 43) URO/GYN PACK, Kit SKU DYNJ66675F; 44) SHOULDER SCOPE PACK, Kit SKU DYNJ68707B; 45) SPECIALTY FACIAL PLASTICS PACK, Kit SKU DYNJ68844B; 46) SPECIALTY FACIAL PLASTICS PACK, Kit SKU DYNJ68844C; 47) ACF PACK, Kit SKU DYNJ69112C; 48) SM OR-DR. YEH PACK-LF, Kit SKU DYNJ69705C; 49) SHOULDER ARTHROSCOPY PACK, Kit SKU DYNJ81752C; 50) C BIRTH SET UP PACK, Kit SKU DYNJ81994C; 51) TRIAL PACK, Kit SKU DYNJ82770A; 52) PROSTATE ROBOTIC PACK, Kit SKU DYNJ84120C; 53) HAND PACK, Kit SKU DYNJ85033A; 54) ARTHROSCOPY PACK, Kit SKU DYNJ85197A; 55) GU-GYN-LITHO PACK, Kit SKU DYNJ86013A; 56) DBD-ARTHROSCOPY PACK, Kit SKU DYNJ86963A; 57) BREAST PACK, Kit SKU DYNJ87587A; 58) PACK,PEDIATRIC MINOR, Kit SKU DYNJ88718; 59) UNIVERSAL-ASC, Kit SKU DYNJ900071C; 60) BASIC, Kit SKU DYNJ905869G; 61) PACK,PEDIATRIC MAJOR, Kit SKU DYNJ906888F; 62) HAND/FOOT PACK, Kit SKU DYNJT3131A; 63) SHOULDER PACK, Kit SKU DYNJT3266; 64) LUMBAR PACK, Kit SKU DYNJT4178; 65) CVS ICU PACK, Kit SKU PHS41728D; 66) SINGLE TAP BLOCK TRAY, Kit SKU SPEC0156A.
Code information (models, lots, serial numbers)
Show the full code information (7k characters)
Medline Kit SKU ACC010270A: UDI/DI 10889942219827 (EA) 40889942219828 (CS), Lot Number 25LDA153; Medline Kit SKU ACC010690: UDI/DI 10195327433307 (EA) 40195327433308 (CS), Lot Number 25LDA088; Medline Kit SKU ACC011073: UDI/DI 10198459569333 (EA) 40198459569334 (CS), Lot Number 25LDA072; Medline Kit SKU DYNJ00282Q: UDI/DI 10195327412029 (EA) 40195327412020 (CS), Lot Number 25KMC904; Medline Kit SKU DYNJ00282Q: UDI/DI 10195327412029 (EA) 40195327412020 (CS), Lot Number 25LMA071; Medline Kit SKU DYNJ0101360K: UDI/DI 10195327662066 (EA) 40195327662067 (CS), Lot Number 25KMC709; Medline Kit SKU DYNJ0220136S: UDI/DI 10198459566073 (EA) 40198459566074 (CS), Lot Number 25KMB403; Medline Kit SKU DYNJ0220136S: UDI/DI 10198459566073 (EA) 40198459566074 (CS), Lot Number 25JMF503; Medline Kit SKU DYNJ0376635Q: UDI/DI 10198459422591 (EA) 40198459422592 (CS), Lot Number 25KMH121; Medline Kit SKU DYNJ0376635Q: UDI/DI 10198459422591 (EA) 40198459422592 (CS), Lot Number 25KMF780; Medline Kit SKU DYNJ0460768T: UDI/DI 10198459466809 (EA) 40198459466800 (CS), Lot Number 25KMH789; Medline Kit SKU DYNJ0509186J: UDI/DI 10198459117626 (EA) 40198459117627 (CS), Lot Number 25KME895; Medline Kit SKU DYNJ0578815W: UDI/DI 10198459466892 (EA) 40198459466893 (CS), Lot Number 25JMJ695; Medline Kit SKU DYNJ0610290T: UDI/DI 10198459021558 (EA) 40198459021559 (CS), Lot Number 25KMC723; Medline Kit SKU DYNJ0753436P: UDI/DI 10198459466854 (EA) 40198459466855 (CS), Lot Number 25KMJ740; Medline Kit SKU DYNJ0890502R: UDI/DI 10198459431098 (EA) 40198459431099 (CS), Lot Number 25KME522; Medline Kit SKU DYNJ28212L: UDI/DI 10193489322118 (EA) 40193489322119 (CS), Lot Number 25KBE916; Medline Kit SKU DYNJ31341P: UDI/DI 10198459069352 (EA) 40198459069353 (CS), Lot Number 25KMA795; Medline Kit SKU DYNJ37615J: UDI/DI 10195327560881 (EA) 40195327560882 (CS), Lot Number 25KDB706; Medline Kit SKU DYNJ39215N: UDI/DI 10198459102240 (EA) 40198459102241 (CS), Lot Number 25KMH027; Medline Kit SKU DYNJ39217P: UDI/DI 10198459102356 (EA) 40198459102357 (CS), Lot Number 25LMA409; Medline Kit SKU DYNJ39220N: UDI/DI 10198459102387 (EA) 40198459102388 (CS), Lot Number 25KMJ608; Medline Kit SKU DYNJ39468G: UDI/DI 10198459408595 (EA) 40198459408596 (CS), Lot Number 25KMD121; Medline Kit SKU DYNJ43954B: UDI/DI 10193489509168 (EA) 40193489509169 (CS), Lot Number 25KMD726; Medline Kit SKU DYNJ45373G: UDI/DI 10198459254635 (EA) 40198459254636 (CS), Lot Number 25JMH775; Medline Kit SKU DYNJ51436F: UDI/DI 10198459004315 (EA) 40198459004316 (CS), Lot Number 25KMG978; Medline Kit SKU DYNJ51437C: UDI/DI 10198459005114 (EA) 40198459005115 (CS), Lot Number 25JMJ673; Medline Kit SKU DYNJ52586C: UDI/DI 10195327596521 (EA) 40195327596522 (CS), Lot Number 25KMH637; Medline Kit SKU DYNJ56889B: UDI/DI 10198459490088 (EA) 40198459490089 (CS), Lot Number 25KBE707; Medline Kit SKU DYNJ57206F: UDI/DI 10195327682958 (EA) 40195327682959 (CS), Lot Number 25KMG080; Medline Kit SKU DYNJ58257I: UDI/DI 10198459565526 (EA) 40198459565527 (CS), Lot Number 25LMA492; Medline Kit SKU DYNJ58410D: UDI/DI 10198459366468 (EA) 40198459366469 (CS), Lot Number 25KBF113; Medline Kit SKU DYNJ58528J: UDI/DI 10198459322815 (EA) 40198459322816 (CS), Lot Number 25JBW494; Medline Kit SKU DYNJ60157F: UDI/DI 10198459343445 (EA) 40198459343446 (CS), Lot Number 25KMI064; Medline Kit SKU DYNJ61696I: UDI/DI 10198459208713 (EA) 40198459208714 (CS), Lot Number 25KMF819; Medline Kit SKU DYNJ62643B: UDI/DI 10198459164446 (EA) 40198459164447 (CS), Lot Number 25KMB841; Medline Kit SKU DYNJ64238D: UDI/DI 10198459068447 (EA) 40198459068448 (CS), Lot Number 25KME422; Medline Kit SKU DYNJ64961C: UDI/DI 10198459569777 (EA) 40198459569778 (CS), Lot Number 25KMD570; Medline Kit SKU DYNJ64988C: UDI/DI 10198459373008 (EA) 40198459373009 (CS), Lot Number 25LMB275; Medline Kit SKU DYNJ65447C: UDI/DI 10198459605222 (EA) 40198459605223 (CS), Lot Number 25LMA822; Medline Kit SKU DYNJ65472C: UDI/DI 10198459168727 (EA) 40198459168728 (CS), Lot Number 25KMD067; Medline Kit SKU DYNJ65996D: UDI/DI 10198459376726 (EA) 40198459376727 (CS), Lot Number 25KMJ550; Medline Kit SKU DYNJ66094A: UDI/DI 10198459462689 (EA) 40198459462680 (CS), Lot Number 25LMB276; Medline Kit SKU DYNJ66613D: UDI/DI 10198459504488 (EA) 40198459504489 (CS), Lot Number 25KMI271; Medline Kit SKU DYNJ66644F: UDI/DI 10198459574160 (EA) 40198459574161 (CS), Lot Number 25KMD640; Medline Kit SKU DYNJ66675F: UDI/DI 10198459574108 (EA) 40198459574109 (CS), Lot Number 25LMA853; Medline Kit SKU DYNJ68707B: UDI/DI 10198459001734 (EA) 40198459001735 (CS), Lot Number 25KBE843; Medline Kit SKU DYNJ68844B: UDI/DI 10198459232428 (EA) 40198459232429 (CS), Lot Number 25KMF829; Medline Kit SKU DYNJ68844C: UDI/DI 10198459608360 (EA) 40198459608361 (CS), Lot Number 25LMA922; Medline Kit SKU DYNJ69112C: UDI/DI 10198459176128 (EA) 40198459176129 (CS), Lot Number 25KME071; Medline Kit SKU DYNJ69705C: UDI/DI 10198459164460 (EA) 40198459164461 (CS), Lot Number 25LMB062; Medline Kit SKU DYNJ81752C: UDI/DI 10198459272721 (EA) 40198459272722 (CS), Lot Number 25KDB800; Medline Kit SKU DYNJ81994C: UDI/DI 10198459105203 (EA) 40198459105204 (CS), Lot Number 25KMH388; Medline Kit SKU DYNJ82770A: UDI/DI 10198459411755 (EA) 40198459411756 (CS), Lot Number 25JME725; Medline Kit SKU DYNJ84120C: UDI/DI 10198459574177 (EA) 40198459574178 (CS), Lot Number 25KMB382; Medline Kit SKU DYNJ85033A: UDI/DI 10195327605629 (EA) 40195327605620 (CS), Lot Number 25KMI735; Medline Kit SKU DYNJ85197A: UDI/DI 10198459217395 (EA) 40198459217396 (CS), Lot Number 25JMD376; Medline Kit SKU DYNJ86013A: UDI/DI 10198459133527 (EA) 40198459133528 (CS), Lot Number 25LDA111; Medline Kit SKU DYNJ86963A: UDI/DI 10198459573330 (EA) 40198459573331 (CS), Lot Number 25KMD001; Medline Kit SKU DYNJ87587A: UDI/DI 10198459114199 (EA) 40198459114190 (CS), Lot Number 25KBE931; Medline Kit SKU DYNJ88718: UDI/DI 10198459195099 (EA) 40198459195090 (CS), Lot Number 25KMF429; Medline Kit SKU DYNJ900071C: UDI/DI 10198459121722 (EA) 40198459121723 (CS), Lot Number 25KMH899; Medline Kit SKU DYNJ905869G: UDI/DI 10198459394577 (EA) 40198459394578 (CS), Lot Number 25JBX559; Medline Kit SKU DYNJ906888F: UDI/DI 10198459194764 (EA) 40198459194765 (CS), Lot Number 25LMA693; Medline Kit SKU DYNJT3131A: UDI/DI 10198459519123 (EA) 40198459519124 (CS), Lot Number 25LMA999; Medline Kit SKU DYNJT3266: UDI/DI 10198459320965 (EA) 40198459320966 (CS), Lot Number 25KMI561; Medline Kit SKU DYNJT4178: UDI/DI 10198459421020 (EA) 40198459421021 (CS), Lot Number 25KMB763; Medline Kit SKU PHS41728D: UDI/DI 10198459438691 (EA) 40198459438692 (CS), Lot Number 25JBW576; Medline Kit SKU SPEC0156A: UDI/DI 10888277239012 (EA) 40888277239013 (CS), Lot Number 25JBW895.
Devices named in this recall
48 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. ACC010690Medline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ62643BMedline Industries, Inc. · General surgery tray3 recalls0 reports
- Medline Industries, Inc. DYNJ85197AMedline Industries, Inc. · General surgery tray2 recalls1 report
- Medline Industries, Inc. DYNJ68844BMedline Industries, Inc. · General surgery tray2 recalls1 report
- Medline Industries, Inc. DYNJ28212LMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43954BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ00282QMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ37615JMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ31341PMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ39217PMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ87587AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0509186JMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900071CMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69112CMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906888FMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61696IMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ45373GMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJT3266Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ58528JMedline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ905869GMedline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJT4178Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ0376635QMedline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ0753436PMedline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ0578815WMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ51437CMedline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. DYNJ69705CMedline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. DYNJ52586CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ85033AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ0101360KMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57206FMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ68707BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ51436FMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ0610290TMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64238DMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ39215NMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ39220NMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ81994CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ86013AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ65472CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ88718Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ81752CMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ64988CMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ82770AMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ0890502RMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. PHS41728DMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ66094AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline SPEC0156AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010270AMedline Industries, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98226).
- Z-1187-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) CATH LAB PACEMAKER PACK-LF, Kit SKU DYNJ64837C; 2) LOOP RECOVERY PACK, Kit SKU DYNJT7034;
- Z-1188-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) PACER / CATH LAB PACK-LF, Kit SKU DYNJ0972604A; 2) PACEMAKER/ICD PACK, Kit SKU DYNJ33822L; 3) VENOUS ACCESS PACK, Kit …
- Z-1189-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) BREAST-HERNIA-PORT CDS-LF, Kit SKU CDS984853L; 2) BREAST BIOPSY PACK, Kit SKU DYNJ45641B; 3) DERMATOLOGY PACK, Kit SKU…
- Z-1190-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) MIDDLE EAR CDS, Kit SKU CDS984347O; 2) HEAD AND NECK CDS-LF, Kit SKU CDS984857N; 3) NECK PACK-LF, Kit SKU DYNJ0966130K…
- Z-1191-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) SMJ LAP CHOLE PACK-LF, Kit SKU DYNJ19974P; 2) DEPAUL ABDOMINAL ENDOSCOPY-LF, Kit SKU DYNJ20027K; 3) LAPAROSCOPY PACK-L…
- Z-1192-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) MEM MAXILLOFACIAL PACK-LF, Kit SKU DYNJ0660040R; 2) PLASTICS VCH, Kit SKU DYNJ39225K; 3) WT BREAST PACK, Kit SKU DYNJ4…
- Z-1194-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: HEAD SHAVE-LATEX SAFE KIT, Kit SKU P903933B
- Z-1195-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.
- Z-1196-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF, Kit SKU DYNJ20028L; 2) NEURO MINOR PACK, Kit SKU DYNJ46546M; 3) NEURO SHUNT PACK, Kit SKU DYNJ5…
- Z-1197-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) RR-GYN LAPAROSCOPY PACK-LF, Kit SKU DYNJ0395022O; 2) PELVISCOPY PACK-LF, Kit SKU DYNJ0415776Q; 3) DIVA PACK, Kit SKU D…
- Z-1198-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: CIRCUMCISION PACK, Kit SKU DYNJ16826O
- Z-1199-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) C-SECTION PACK, Kit SKU DYNJ35494C; 2) SMJ C-SECTION PACK-LF, Kit SKU DYNJ38582O; 3) RR-L&D C-SECTION PACK-LF, Kit SKU…
- Z-1200-2026 · Open, ClassifiedMedline medical procedure convenience kits labeled as: 1) WATKINS SPINE PACK-LF, Kit SKU DYNJ0275635M; 2) SM-SPINE PACK-LF, Kit SKU DYNJ0395037X; 3) TOTAL HIP PACK, Kit SKU DYN…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page