Product Name: Lumipulse G pTau 217 Plasma Immunoreaction… recallZ-1302-2026 · Fujirebio Diagnostics, Inc. · initiated Dec 11, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fujirebio Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
FDA root cause category: Under Investigation by firm
What the company is doing
On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.
Recall details
- Recall number
- Z-1302-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 11, 2025
- Posted by the FDA
- Feb 5, 2026
- Recalling firm
- Fujirebio Diagnostics, Inc.
- Quantity
- 1569 units
- Product code
- SET
- FDA submissions
- K242706
- Distribution
- US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Product description
Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Code information (models, lots, serial numbers)
Model/Catalog Number: 81550; UDI-DI: 05414470815500; Lot Numbers: D4C6028U, D4C6029U, D4C6051U, D4C6052U;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K242706 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio CL0073Fujirebio Diagnostics, Inc. · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G B-Amyloid 1-42-N Plasma 235447Fujirebio Inc. · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G B-Amyloid 1-42-N Plasma Calibrator 235454Fujirebio Inc. · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G pTau 217 Plasma Calibrators 81557Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse pTau 217 Plasma Controls 81558Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse B-Amyloid Plasma Controls 81559Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98232).
- Z-1301-2026 · Open, ClassifiedProduct Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/-Amyloid 1-42 P…
- Z-1303-2026 · Open, ClassifiedProduct Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The…
- Z-1304-2026 · Open, ClassifiedProduct Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description: Lumipulse pTau 217 Plasma Controls: This quality …
- Z-1305-2026 · Open, ClassifiedProduct Name: Lumipulse G -Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1…
- Z-1306-2026 · Open, ClassifiedProduct Name: Lumipulse G -Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1-42-N Plasma C…
- Z-1307-2026 · Open, ClassifiedProduct Name: Lumipulse -Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse -Amyloid Plasma Controls: This quality …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page