Product Name: Lumipulse -Amyloid Plasma Controls recallZ-1307-2026 · Fujirebio Diagnostics, Inc. · initiated Dec 11, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fujirebio Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
FDA root cause category: Under Investigation by firm
What the company is doing
On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.
Recall details
- Recall number
- Z-1307-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 11, 2025
- Posted by the FDA
- Feb 5, 2026
- Recalling firm
- Fujirebio Diagnostics, Inc.
- Quantity
- 179 units
- Product code
- SET
- FDA submissions
- K242706
- Distribution
- US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Product description
Product Name: Lumipulse -Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse -Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing -amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse -Amyloid Plasma Controls: Liquid (Frozen), 2 2 Concentrations L1 Level 1 (2 1.5 mL) L2 Level 2 (2 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Code information (models, lots, serial numbers)
Model/Catalog Number: 81559; UDI-DI: 05414470815593; Lot Number: T8B6116U;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Lumipulse B-Amyloid Plasma Controls 81559Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K242706 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio CL0073Fujirebio Diagnostics, Inc. · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G B-Amyloid 1-42-N Plasma 235447Fujirebio Inc. · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G B-Amyloid 1-42-N Plasma Calibrator 235454Fujirebio Inc. · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse G pTau 217 Plasma Calibrators 81557Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
- Lumipulse pTau 217 Plasma Controls 81558Fujirebio Europe · Immunoassay Blood Test For Amyloid Pathology Assessment1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98232).
- Z-1301-2026 · Open, ClassifiedProduct Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/-Amyloid 1-42 P…
- Z-1302-2026 · Open, ClassifiedProduct Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Im…
- Z-1303-2026 · Open, ClassifiedProduct Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The…
- Z-1304-2026 · Open, ClassifiedProduct Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description: Lumipulse pTau 217 Plasma Controls: This quality …
- Z-1305-2026 · Open, ClassifiedProduct Name: Lumipulse G -Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1…
- Z-1306-2026 · Open, ClassifiedProduct Name: Lumipulse G -Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1-42-N Plasma C…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page