Skip to content

Philips Incisive CT recallZ-1003-2026 · Philips Healthcare (Suzhou) Co., Ltd. · initiated Dec 3, 2025

Class II Open, Classified All recalls posted in January 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Healthcare (Suzhou) Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for incomplete scan due to unstable connection inside of floating sensor.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install a hardware solution that will change the method of connection for the rear handle floating sensor. Customers who need additional information or support concerning this issue should be directed to their local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center). Philips reference: 2025-PD-CTAMI-109 or FCO72800843.

Recall details

Recall number
Z-1003-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Dec 3, 2025
Posted by the FDA
Jan 8, 2026
Recalling firm
Philips Healthcare (Suzhou) Co., Ltd.
Quantity
35
Product code
JAK
FDA submissions
K180015, K212441
Distribution
U.S. and U.S. territories

Product description

Philips Incisive CT

Code information (models, lots, serial numbers)

N/A

Other devices under the same FDA submission

1 device cleared under K180015, K212441 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page