Why it was recalled

Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.

FDA root cause category: Under Investigation by firm

What the company is doing

Customers with clinically significant isolates of Enterobacterales species and Pseudomonas aeruginosa with resistant results to Ceftazidime/Avibactam (cza02n), Imipenem (ipm05n), Imipenem/Relebactam (ipr01n), Meropenem/Vaborbactam (mev01n), or Polymyxin B (pb02n) should consider using an alternative method to confirm results. The determination of clinical significance should be made by each laboratory based on the antimicrobial susceptibility profile and the local availability of alternative antibiotics. Customers can set up a custom bioART rule to trigger when resistant results are obtained for the impacted bug drug combinations above. Based on individual customer policy, customers may wish to perform retrospective review of resistant results. ***Update*** On 4/2/26, the firm added additional products to the scope of the recall. The additional product was not manufactured at the time of the original notification, and the additional products are distributed OUS only.

Recall details

Recall number
Z-1236-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jan 5, 2026
Posted by the FDA
Feb 2, 2026
Recalling firm
Biomerieux Inc
Quantity
2,668,954 total
Product code
LTW
FDA submissions
K172944, K183360, K183415, K193567, K211136
Distribution
Worldwide - US Nationwide distribution.

Product description

VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 ***Update 4/2/26 -- Additional product references not manufactured at time of initial notification. These products will not be distributed within the United States.*** AST-N484 Ref. 425181 AST-N486 Ref. 425183 AST-N491 Ref. 425206 AST-N494 Ref. 425224 AST-N485 Ref. 425182 AST-XN41 Ref. 425204 AST-XN42 Ref. 425309 AST-N504 Ref. 425491

Code information (models, lots, serial numbers)

All lot codes of the identified Gram-negative AST test kits.

Other devices under the same FDA submission

5 devices cleared under K172944, K183360, K183415, K193567, K211136 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page