VITEK® 2 AST-N802Recalls & reported problems
Biomerieux , Inc. · VITEK® 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems to determine the susceptibility of clinically significant aerobic gram-negative bacilli · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K072038, K091899, K110106, K113200, K161217, K161227, K162737, K163563, K172731, K183415, N50510); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 14, 2024Open, Classified
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
Z-1464-2024 · Root cause: Employee error · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072038, K091899, K110106, K113200, K161217, K161227, K162737, K163563, K172731, K183415, N50510. One submission often covers many models, so check the product and lot details in each record.
- Z-1236-2026 · Jan 5, 2026 · Open, ClassifiedVITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebac
Other SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION devices
- BD Phoenix™ M50 Automated Microbiology System 443624Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION3 recalls417 reports
- BD Phoenix™ NMIC/ID-307 449289Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION529 reports
- BD Phoenix™ PMIC/ID-107 448607Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION2 recalls169 reports
- BD Phoenix™ NMIC-306 449292Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION1 recall143 reports
- VITEK® 2 AST-GP67Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls83 reports
- BD Phoenix™ AST Broth 246003Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION182 reports
- VITEK® 2 AST-GP75Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls52 reports
- VITEK® 2 AST-GP78Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls41 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.