Skip to content

Brand Name: Spectrum IQ recallZ-1368-2026 · Baxter Healthcare Corporation · initiated Jan 13, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Pumps were released without full testing being performed, including occlusion alarm testing.

FDA root cause category: Employee error

What the company is doing

Firm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.

Recall details

Recall number
Z-1368-2026
Status
Open, Classified
Initiated by the company
Jan 13, 2026
Posted by the FDA
Feb 13, 2026
Recalling firm
Baxter Healthcare Corporation
Quantity
5
Product code
FRN
FDA submissions
K251636
Distribution
States: Ohio Countries: United States

Product description

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

Code information (models, lots, serial numbers)

Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K251636 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates