TRUE METRIX recallZ-1545-2026 · Trividia Health, Inc. · initiated Feb 6, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Trividia Health, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
FDA root cause category: Software design
What the company is doing
On 2/6/2026, press was issued and correction notices were emailed to customers. E-5 Error Code in the Messages section of the Owner's Booklets/System Instructions for Use emphasizes that users must seek medical attention immediately if they receive an E-5 error code and are experiencing symptoms of high glucose. Customers: notify your customers immediately of the correction and to include the letter in their notification, and to ask that they notify their customers or facilities if they have further distributed the Products. Retail pharmacies: post the notice in areas where the Products are sold to notify patients. Multiple Patient Facilities: post the notice in areas where the Products are stored to notify personnel. People with Diabetes are asked to follow the updated E-5 error code instructions Visit www.trividiahealth.com for updated Owner's Booklets and help resources. Complete and return the acknowledgement form via email to [email protected] Firm Customer Care Department: 1-888-835-2723, [email protected] or visit www.trividiahealth.com/E-5productnotice , which has links to Owner's Booklets, correction notice, and press. On 5/5/2026, updated correction notices were mailed to customers informing them of the following: For people with diabetes, if possible, consider transitioning to an alternative method of testing your blood glucose. People with Diabetes who rely on intensive insulin therapy, sulfonylureas, or glucose monitoring due to frequent hypo- or hyperglycemia events should consider transitioning to an alternative testing method until the updated devices with updated software becomes available. Firm will notify users when the upgrade program becomes available. The firm asked customers to handle dissemination of the updated notice and completion/return of the acknowledgement form as stated above. Affected serial number search tool is located at www.trividiahealth.com/E-5productnotice
Recall details
- Recall number
- Z-1545-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 6, 2026
- Posted by the FDA
- Mar 12, 2026
- Recalling firm
- Trividia Health, Inc.
- Quantity
- 12,067,458
- Product code
- NBW
- FDA submissions
- K140100
- Distribution
- Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
Product description
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter Only/RE4095-40, HealthyAccents TRUE METRIX Kit/RE4097-01, Shopko TRUE METRIX Kit/RE4098-01, Henry Schein TRUE METRIX Kit/RE4099-40, Henry Schein TRUE METRIX Meter Only/RE4099-45, Millennium Pharmay Services TRUE METRIX Kit/RE4202-43, Kroger TRUE METRIX Kit/RE4203-01, TOPCARE TRUE METRIX Kit/RE4209-01, Foster and Thrive TRUE METRIX Kit/RE4211-01, TRUE METRIX Meter Only/RE4H01-00, TRUE METRIX Kit/RE4H01-01, TRUE METRIX Starter Kit/RE4H01-04, TRUE METRIX NFRS Meter Only/RE4H01-40, TRUE METRIX NFRS Starter Kit/RE4H01-43, TRUE METRIX Kit/RE4H01-51, TRUE METRIX Meter Only (mg/dL)/RE4i01-00, TRUE METRIX Kit (mg/dL)/RE4i01-01, TRUE METRIX Kit (mg/dL)/RE4i01-02, TRUE METRIX Starter Kit (mg/dL)/RE4i03-02, TRUE METRIX Kit (mmol/L)/RE4i03-11, Farmacia Benavides TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i05-00, Farmacia del Ahorro TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i07-00, TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i23-00, TRUE METRIX Meter Only (Jamaica - mmol/L)/RE4i29-11, TRUE METRIX Kit (mg/dL)/RE4i61-02, TRUE METRIX Kit (mg/dL)/RE4i61-02THI, TRUE METRIX Kit (United Kingdom - mmol/L)/RE4i82-11, TRUE METRIX Starter Kit (United Kingdom - mmol/L)/RE4i82-12.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE4011-00/87701426520, RE4011-01/87701426254, RE4019-01/713733069816, RE4023-01/21292007256, RE4025-01/93351018992, RE4027-01/21292007065, RE4031-01/41220001286, RE4051-43/20612479197231, RE4054-00/10939722447, RE4054-01/10939719447, RE4060-01/321130788669, RE4061-01/41520147745, RE4066-01/11822445610, RE4078-40/21292007614, RE4081-03/21292014797, RE4087-01/77890394113, RE4089-00/52569138557, RE4089-01/52569138526, RE4095-40/21292007041, RE4097-01/725439988933, RE4098-01/21292008901, RE4099-40/21292014902, RE4099-45/21292008697, RE4202-43/21292007454, RE4203-01/41260016028, RE4209-01/36800234529, RE4211-01/10939958693, RE4H01-00/21292006082, RE4H01-01/21292006051, RE4H01-04/21292008895, RE4H01-40/21292006068, RE4H01-43/21292006075, RE4H01-51/21292014940. International: RE4i01-00/N/A, RE4i01-01/N/A, RE4i01-02/N/A, RE4i03-02/N/A, RE4i03-11/N/A, RE4i05-00/N/A, RE4i07-00/N/A, RE4i23-00/21292010133, RE4i29-11/N/A, RE4i61-02/N/A, RE4i61-02THI/N/A, RE4i82-11/21292012205, RE4i82-12/21292009977. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX Owner's Booklet/ RE4TVH03 Rev 56 and prior, TRUE METRIX Owner's Booklet (English)/ RE4TVH35 Rev 51 and prior, TRUE METRIX Owner's Booklet (Spanish)/ RE4TVH35S Rev 51 and prior, McKesson Med Surg TRUE METRIX Self Monitoring Owners Booklet/ RE4SUN03 Rev 55 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior, TRUE METRIX Blood Glucose Monitoring System IFU Trividia House Brand/ RE4ITV16 Rev 53 and prior, TRUE METRIX System Instructions for Use (IFU) for Trividia Health UK Limited/RE4UKT16 Rev 55 and prior, Trividia Health LATAM - TRUE METRIX System IFU mg/dL/ RE4LAT16 Rev 50 and prior, TRUE METRIX System Instructions for Use (IFU) for Trividia Health Australia/ RE4AUT16 Rev 56 and prior. Device serial numbers: Prefix "T" followed by an eight-digit numeric sequence.
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- McKesson 06-RE4051-43Mckesson Medical-Surgical Inc. · System, Test, Blood Glucose, Over The Counter1 recall45 reports
- TRUE METRIX TRUE METRIX Self Monitoring Blood Glucose MeterTrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter1 recall6 reports
- TRUE METRIX TRUE METRIX self monitoring blood glucose meterTrividia Health, Inc. · Glucose Dehydrogenase, Glucose1 recall3 reports
- TRUE METRIX TRUE METRIX self monitoring blood glucose meterTrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter1 recall0 reports
- TRUE METRIX TRUE METRIX Self Monitoring Blood Glucose MeterTrividia Health, Inc. · Glucose Dehydrogenase, Glucose1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98317).
- Z-1546-2026 · Open, ClassifiedOwner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX …
- Z-1547-2026 · Open, ClassifiedOwner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE…
- Z-1548-2026 · Open, ClassifiedOwner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE ME…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.