TRUE METRIX PRO recallZ-1546-2026 · Trividia Health, Inc. · initiated Feb 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Trividia Health, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

FDA root cause category: Software design

What the company is doing

On 2/6/2026, press was issued and correction notices were emailed to customers. E-5 Error Code in the Messages section of the Owner's Booklets/System Instructions for Use emphasizes that users must seek medical attention immediately if they receive an E-5 error code and are experiencing symptoms of high glucose. Customers: notify your customers immediately of the correction and to include the letter in their notification, and to ask that they notify their customers or facilities if they have further distributed the Products. Retail pharmacies: post the notice in areas where the Products are sold to notify patients. Multiple Patient Facilities: post the notice in areas where the Products are stored to notify personnel. People with Diabetes are asked to follow the updated E-5 error code instructions Visit www.trividiahealth.com for updated Owner's Booklets and help resources. Complete and return the acknowledgement form via email to [email protected] Firm Customer Care Department: 1-888-835-2723, [email protected] or visit www.trividiahealth.com/E-5productnotice , which has links to Owner's Booklets, correction notice, and press. On 5/5/2026, updated correction notices were mailed to customers informing them of the following: For people with diabetes, if possible, consider transitioning to an alternative method of testing your blood glucose. People with Diabetes who rely on intensive insulin therapy, sulfonylureas, or glucose monitoring due to frequent hypo- or hyperglycemia events should consider transitioning to an alternative testing method until the updated devices with updated software becomes available. Firm will notify users when the upgrade program becomes available. The firm asked customers to handle dissemination of the updated notice and completion/return of the acknowledgement form as stated above. Affected serial number search tool is located at www.trividiahealth.com/E-5productnotice

Recall details

Recall number
Z-1546-2026
Status
Open, Classified
Initiated by the company
Feb 6, 2026
Posted by the FDA
Mar 12, 2026
Recalling firm
Trividia Health, Inc.
Quantity
1,001,534
Product code
NBW
FDA submissions
K140100
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.

Product description

Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.

Code information (models, lots, serial numbers)

Device Item/UDI-DI: RE4H01P-40/21292006099, RE4H01P-43/21292006105, RE4212P-00/10840330705681, RE4099P-40/21292014919, RE4099P-43/21292008673, RE4051P-00/20612479197217, RE4068P-01/607415261097. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX PRO Owners Booklet/RE4TVHP03 Rev 56 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior, Device serial numbers: Prefix "TP" followed by a seven-digit numeric sequence.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98317).