Medline Convenience Kits recallZ-1403-2026 · Medline Industries, LP · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Proxima Sterile Surgical Gowns, Packs, and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-003 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Medline Kits containing Proxima Sterile Surgical Gowns and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form,
Recall details
- Recall number
- Z-1403-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- Feb 17, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 32 units
- Product code
- OIB
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Medline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L
Code information (models, lots, serial numbers)
1) DYNJ50806L, UDI-DI: 10195327344672(each), 40195327344673(case), Lot Number: 23FBK937
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ50806LMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98329).
- Z-1384-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A
- Z-1385-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55A; 3) TRACH TOTE, Model Number: DYNJ85691
- Z-1386-2026 · Open, ClassifiedMedline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number: DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number: DYNJ56436B; 3) CPNB SETUP TRAY, Model Number: DYNJRA1979; 4) CPNB SET…
- Z-1387-2026 · Open, ClassifiedMedline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C
- Z-1388-2026 · Open, ClassifiedMedline Convenience Kits: 1) ENDO KIT W/SYRINGE, Model Number: DYKE1743; 2) BGMC ANGIOGRAPHY PACK-LF, Model Number: DYNJ0800934D; 3) GENERAL ENDO PACK-LF, Model Number: DYNJ08428…
- Z-1389-2026 · Open, ClassifiedMedline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Numbe…
- Z-1390-2026 · Open, ClassifiedMedline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP…
- Z-1391-2026 · Open, ClassifiedMedline Convenience Kits: 1) PICC LINE TRAY, Model Number: 00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number: ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Numbe…
- Z-1392-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:…
- Z-1393-2026 · Open, ClassifiedMedline Convenience Kits: 1) DENTAL PACK, Model Number: DYNDA3003
- Z-1394-2026 · Open, ClassifiedMedline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number:…
- Z-1395-2026 · Open, ClassifiedMedline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W;…
- Z-1396-2026 · Open, ClassifiedMedline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721
- Z-1397-2026 · Open, ClassifiedMedline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA, Model …
- Z-1398-2026 · Open, ClassifiedMedline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A;
- Z-1399-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER #5, Model Number: ACC010475; 4…
- Z-1400-2026 · Open, ClassifiedMedline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE MOMENI SB CAPSULOTOMY REC, Model …
- Z-1401-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 2) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: …
- Z-1402-2026 · Open, ClassifiedMedline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43
- Z-1404-2026 · Open, ClassifiedMedline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
- Z-1405-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A
- Z-1406-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAPE PACK-CHOICE, Model Number: DYNJ63118A
- Z-1407-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
- Z-1409-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
- Z-1410-2026 · Open, ClassifiedMedline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A
- Z-1411-2026 · Open, ClassifiedMedline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
- Z-1413-2026 · Open, ClassifiedMedline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYK…
- Z-1414-2026 · Open, ClassifiedMedline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL…
- Z-1415-2026 · Open, ClassifiedMedline Convenience Kits: 1) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 2) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 3) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 4)…
- Z-1416-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL1…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page