Medline Convenience Kits recallZ-1415-2026 · Medline Industries, LP · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Proxima Sterile Surgical Gowns, Packs, and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-003 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Medline Kits containing Proxima Sterile Surgical Gowns and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form,
Recall details
- Recall number
- Z-1415-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- Feb 17, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 12418 units
- Product code
- OHD
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Medline Convenience Kits: 1) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 2) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 3) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 4) ROBOTIC-LF, Model Number: CDS984543D; 5) ROBOTIC-LF, Model Number: CDS984543F; 6) ROBOTIC-LF, Model Number: CDS984543G; 7) ROBOTIC-LF, Model Number: CDS984543I; 8) GYN CDS, Model Number: CDS984860F; 9) GYN LAP PELVI, Model Number: CDS985423G; 10) GYN KIT, Model Number: DYKM1657A; 11) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 12) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 13) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 14) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 15) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 16) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 17) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 18) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 19) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 20) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91; 21) KIT GYN ONCOLOGY, Model Number: DYKMBNDL92; 22) KIT GYN ONCOLOGY, Model Number: DYKMBNDL92A; 23) KIT GYN ABDOMINAL, Model Number: DYKMBNDL92C; 24) KIT GYN LAP, Model Number: DYKMBNDL94; 25) KIT GYN LAP, Model Number: DYKMBNDL94A; 26) KIT GYN LAP, Model Number: DYKMBNDL94C; 27) KIT GYN LAP, Model Number: DYKMBNDL94D; 28) GYN LAPAROSCOPY PACK-LF, Model Number: DYNJ0281948N; 29) PROSTATE ROBOTIC PACK, Model Number: DYNJ0347372F; 30) PROSTATE ROBOTIC PACK, Model Number: DYNJ0347372G; 31) PROSTATE ROBOTIC PACK, Model Number: DYNJ0347372I; 32) HEALTH ALL GYN LAP PACK-LF, Model Number: DYNJ28252G; 33) GYN OPERATIVE LAP PACK-LF, Model Number: DYNJ30005G; 34) OB PACK, Model Number: DYNJ32706B; 35) ROBOTIC UROLOGY-RF, Model Number: DYNJ38844N; 36) ENSEMBLE GYNE LAPAROTOMIE-LF, Model Number: DYNJ47550C; 37) GYN LAP PACK, Model Number: DYNJ47714A; 38) TASC GYN LAP, Model Number: DYNJ49191B; 39) ROBOTIC PACK, Model Number: DYNJ49705K; 40) GYN LAPAROSCOPY PACK, Model Number: DYNJ50348F; 41) NWH GYN LAPAROSCOPY, Model Number: DYNJ50439F; 42) FH LAPAROSCOPIC GYN PK-LF, Model Number: DYNJ51795F; 43) GYN LAPAROSCOPY PACK, Model Number: DYNJ52569B; 44) LAVH PACK, Model Number: DYNJ55862; 45) PK, GEN-LAPAROSCOPY, Model Number: DYNJ57607A; 46) GYN LAPAROSCOPY MRNGSD, Model Number: DYNJ58321A; 47) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079B; 48) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079C; 49) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079D; 50) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079F; 51) GYN DAVINCI PACK, Model Number: DYNJ64266; 52) DAVINCI PACK, Model Number: DYNJ66298D; 53) GYN LAPAROSCOPY, Model Number: DYNJ67269; 54) GENERAL GYN PACK, Model Number: DYNJ68710; 55) LAP GYN PACK, Model Number: DYNJ69419A; 56) LAP GYN PACK, Model Number: DYNJ69419D; 57) GYN LAPAROSCOPY, Model Number: DYNJ81492A; 58) PK, ROBOTIC GYN & GU-ROSE MEDI, Model Number: DYNJ81761A; 59) GYN PACK, Model Number: DYNJ83475A; 60) ROBOTIC XI PACK, Model Number: DYNJ84783; 61) GYN PACK, Model Number: DYNJ86195A; 62) GYN LAPAROSCOPY, Model Number: DYNJ900262A; 63) D&C HYSTEROSCOPY, Model Number: DYNJ901133A; 64) GYN LAPAROSCOPY, Model Number: DYNJ901559J; 65) GYN ENDOSCOPY-LF, Model Number: DYNJ901812J; 66) GYN LITHOTOMY-LF, Model Number: DYNJ902711M; 67) GENERAL ROBOTIC, Model Number: DYNJ905066B; 68) TRINITY C-SEC W/LINEN CDS, Model Number: DYNJ905269G; 69) KIT GYN LAPAROSCOPY RFD, Model Number: DYNJ906322D; 70) GYN LAP LH, Model Number: DYNJ906452C; 71) GYN LAP LH, Model Number: DYNJ906452D; 72) GYN LAP LH, Model Number: DYNJ906452F; 73) FLOYD LITH GYN, Model Number: DYNJ906926B; 74) GYN LAPAROSCOPY-MRMC, Model Number: DYNJ907753F; 75) GYN LAPAROSCOPY-SFMC, Model Number: DYNJ907771C; 76) LAP/GYN ROBOTICS, Model Number: DYNJ908887B; 77) LAP/GYN ROBOTICS, Model Number: DYNJ908887C;
Code information (models, lots, serial numbers)
Show the full code information (30k characters)
1) CDS982893I, UDI-DI: 10889942208050(each), 40889942208051(case), Lot Number: 24IBO240; 2) CDS983300J, UDI-DI: 10193489425048(each), 40193489425049(case), Lot Number: 21CDB781; 3) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23CBX440; 4) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23CBN644; 5) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23ABC549; 6) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22OBM422; 7) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22KBD152; 8) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22JBC547; 9) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22HBV491; 10) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22GBP114; 11) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22FBV462; 12) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22BBL450; 13) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21VBA272; 14) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21IBM804; 15) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21HBN807; 16) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21GBI171; 17) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21EBA117; 18) CDS984543D, UDI-DI: 10195327072322(each), 40195327072323(case), Lot Number: 22CBC587; 19) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22OBL925; 20) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22KMD619; 21) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22JMA223; 22) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22HMG671; 23) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22HMB259; 24) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22GBD616; 25) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 24BMH483; 26) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23LMA864; 27) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23JMC376; 28) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23IMA629; 29) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23GMH282; 30) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23GMF425; 31) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23FMI171; 32) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23EMI460; 33) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23DMA508; 34) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23CMC582; 35) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23BMD680; 36) CDS984543I, UDI-DI: 10195327637071(each), 40195327637072(case), Lot Number: 24CMH050; 37) CDS984860F, UDI-DI: 10193489261974(each), 40193489261975(case), Lot Number: 21ADB205; 38) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21DBV243; 39) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21CBH547; 40) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21CBH282; 41) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25LBN225; 42) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25HBS386; 43) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25FBS108; 44) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25DBN805; 45) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21JBD206; 46) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21GBP406; 47) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21EBA097; 48) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21BBG514; 49) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21ABI962; 50) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABW347; 51) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABO060; 52) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABM086; 53) DYKMBNDL116D, UDI-DI: 10195327608866(each), 40195327608867(case), Lot Number: 24DBA349; 54) DYKMBNDL116D, UDI-DI: 10195327608866(each), 40195327608867(case), Lot Number: 24CBF889; 55) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24JBB328; 56) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24IBV184; 57) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24GBK948; 58) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24FBC411; 59) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24DBS388; 60) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22HBL832; 61) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22FBI695; 62) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22CBX517; 63) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22BBR618; 64) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21LBY301; 65) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21LBP646; 66) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21JBX803; 67) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21JBC127; 68) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21HBK337; 69) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21FBP762; 70) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21EBG218; 71) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21BBP181; 72) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21BBA024; 73) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21ABL615; 74) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21ABJ213; 75) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25HBB957; 76) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25GBX843; 77) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25FBR662; 78) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24KBI207; 79) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24KBF533; 80) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24JBD270; 81) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24DBV608; 82) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24CBJ465; 83) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24ABX678; 84) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 23LBE369; 85) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 23KBA328; 86) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 22JBC546; 87) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22HBP253; 88) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22CBX518; 89) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22ABL299; 90) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22ABL283; 91) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21KBQ606; 92) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21JBJ225; 93) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21IBL095; 94) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21GBU222; 95) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21GBP369; 96) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21EBK837; 97) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21BBG518; 98) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21ABI984; 99) DYKMBNDL168A, UDI-DI: 10195327240806(each), 40195327240807(case), Lot Number: 23FBJ346; 100) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25LBQ333; 101) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25LBG179; 102) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25HBC798; 103) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25FBJ143; 104) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25BBF014; 105) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25ABS671; 106) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24HBK278; 107) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24GBV008; 108) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24GBJ568; 109) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 22HBB985; 110) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 22GBZ017; 111) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21LBJ751; 112) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21IBO770; 113) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21HBK341; 114) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21FBP714; 115) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21BBG520; 116) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21ABJ086; 117) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22HLA664; 118) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ELB163; 119) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ELA227; 120) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22DLB036; 121) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22CLA119; 122) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22BLA918; 123) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ALA331; 124) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ALA125; 125) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21LLA284; 126) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21LLA163; 127) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21JLA896; 128) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21HLA598; 129) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA658; 130) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA352; 131) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA256; 132) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21DLA017; 133) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21BLA496; 134) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ABJ861; 135) DYKMBNDL92A, UDI-DI: 10195327241209(each), 40195327241200(case), Lot Number: 22KLA731; 136) DYKMBNDL92A, UDI-DI: 10195327241209(each), 40195327241200(case), Lot Number: 22JLA665; 137) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25ILA134; 138) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25HLA275; 139) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25GLA701; 140) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25ELA660; 141) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25DLB036; 142) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25BLA797; 143) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLB092; 144) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLA506; 145) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLA194; 146) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24JLB140; 147) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24JLA881; 148) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24HLB219; 149) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24HLB142; 150) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 22HBH516; 151) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 22FBK394; 152) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21IBV704; 153) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21HBO271; 154) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21GBB013; 155) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21CBS045; 156) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21CBB287; 157) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21ABM148; 158) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 20LBY985; 159) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23BBI491; 160) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABO813; 161) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABK219; 162) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABJ591; 163) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 22LBS837; 164) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25CBR621; 165) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25BBA805; 166) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25ABN371; 167) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25ABH279; 168) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24LBN994; 169) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24KBA746; 170) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24IBE086; 171) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24EBV430; 172) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24DBR903; 173) DYKMBNDL94D, UDI-DI: 10198459350641(each), 40198459350642(case), Lot Number: 25EBN175; 174) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24HBX130; 175) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24HBA113; 176) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24EBT021; 177) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24EBM495; 178) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24DBL074; 179) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24CBM999; 180) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24BBH633; 181) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24ABI719; 182) DYNJ0347372F, UDI-DI: 10195327213176(each), 40195327213177(case), Lot Number: 22GBR392; 183) DYNJ0347372G, UDI-DI: 10195327266028(each), 40195327266029(case), Lot Number: 23DBP818; 184) DYNJ0347372G, UDI-DI: 10195327266028(each), 40195327266029(case), Lot Number: 22KBU586; 185) DYNJ0347372I, UDI-DI: 10195327418335(each), 40195327418336(case), Lot Number: 23IBK088; 186) DYNJ0347372I, UDI-DI: 10195327418335(each), 40195327418336(case), Lot Number: 23GBO695; 187) DYNJ28252G, UDI-DI: 10198459088124(each), 40198459088125(case), Lot Number: 25CBC137; 188) DYNJ28252G, UDI-DI: 10198459088124(each), 40198459088125(case), Lot Number: 25BBP572; 189) DYNJ30005G, UDI-DI: 10193489601794(each), 40193489601795(case), Lot Number: 21DMA051; 190) DYNJ32706B, UDI-DI: 10195327364304(each), 40195327364305(case), Lot Number: 23EMB628; 191) DYNJ38844N, UDI-DI: 10195327243166(each), 40195327243167(case), Lot Number: 23HBI619; 192) DYNJ38844N, UDI-DI: 10195327243166(each), 40195327243167(case), Lot Number: 23DBI585; 193) DYNJ47550C, UDI-DI: 10195327209964(each), 40195327209965(case), Lot Number: 23DBD093; 194) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25KBF492; 195) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24FBG329; 196) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24CBQ847; 197) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23HBO785; 198) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23FBU806; 199) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23FBH830; 200) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 23CMB495; 201) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22KMH193; 202) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22HMC351; 203) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22HMA973; 204) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22EMA211; 205) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22BMH522; 206) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21KMG274; 207) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMH124; 208) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMF391; 209) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21GMD682; 210) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21AMA809; 211) DYNJ49705K, UDI-DI: 10195327020347(each), 40195327020348(case), Lot Number: 22IMI306; 212) DYNJ50348F, UDI-DI: 10195327406349(each), 40195327406340(case), Lot Number: 23IMD230; 213) DYNJ50439F, UDI-DI: 10193489704921(each), 40193489704922(case), Lot Number: 22EBC737; 214) DYNJ51795F, UDI-DI: 10193489475937(each), 40193489475938(case), Lot Number: 21ADA184; 215) DYNJ51795F, UDI-DI: 10193489475937(each), 40193489475938(case), Lot Number: 20LDC148; 216) DYNJ52569B, UDI-DI: 10193489685190(each), 40193489685191(case), Lot Number: 23EBS784; 217) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 23BBA462; 218) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22LBD784; 219) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22HBV228; 220) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22GBJ776; 221) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22ABC476; 222) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21JBR818; 223) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21IBG462; 224) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21CBU815; 225) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21GME150; 226) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21FMB889; 227) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21EMB589; 228) DYNJ58321A, UDI-DI: 10193489489316(each), 40193489489317(case), Lot Number: 22ABG197; 229) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 22AME373; 230) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21JMF224; 231) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21HMD002; 232) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21EMF383; 233) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21DMB839; 234) DYNJ63079C, UDI-DI: 10195327102180(each), 40195327102181(case), Lot Number: 23GMB716; 235) DYNJ63079C, UDI-DI: 10195327102180(each), 40195327102181(case), Lot Number: 23EMI624; 236) DYNJ63079D, UDI-DI: 10195327505677(each), 40195327505678(case), Lot Number: 24BME368; 237) DYNJ63079D, UDI-DI: 10195327505677(each), 40195327505678(case), Lot Number: 23JMB435; 238) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25JMK630; 239) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25JMH582; 240) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25AMH256; 241) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24JMD681; 242) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24HMH481; 243) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24HMC603; 244) DYNJ64266, UDI-DI: 10193489795516(each), 40193489795517(case), Lot Number: 21ADA447; 245) DYNJ66298D, UDI-DI: 10195327095895(each), 40195327095896(case), Lot Number: 23IBW233; 246) DYNJ67269, UDI-DI: 10193489394696(each), 40193489394697(case), Lot Number: 22KBG504; 247) DYNJ68710, UDI-DI: 10193489842906(each), 40193489842907(case), Lot Number: 22ABM641; 248) DYNJ69419A, UDI-DI: 10195327223588(each), 40195327223589(case), Lot Number: 23DBK434; 249) DYNJ69419A, UDI-DI: 10195327223588(each), 40195327223589(case), Lot Number: 23ABS144; 250) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24KMH362; 251) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24KME386; 252) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24GMG805; 253) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24GMC135; 254) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24EMD677; 255) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24EMC427; 256) DYNJ81492A, UDI-DI: 10195327348830(each), 40195327348831(case), Lot Number: 23BBT116; 257) DYNJ81761A, UDI-DI: 10195327574468(each), 40195327574469(case), Lot Number: 24FMF687; 258) DYNJ83475A, UDI-DI: 10195327460938(each), 40195327460939(case), Lot Number: 23HBY488; 259) DYNJ84783, UDI-DI: 10195327417949(each), 40195327417940(case), Lot Number: 23IBU517; 260) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 24EBD237; 261) DYNJ900262A, UDI-DI: 10195327477820(each), 40195327477821(case), Lot Number: 23KBU736; 262) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 22JBR884; 263) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21KBP568; 264) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21KBF079; 265) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21JBV891; 266) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21JBM245; 267) DYNJ901559J, UDI-DI: 10193489235890(each), 40193489235891(case), Lot Number: 21EBR740; 268) DYNJ901559J, UDI-DI: 10193489235890(each), 40193489235891(case), Lot Number: 21CBH289; 269) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21FDB236; 270) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21BDA744; 271) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21ADB190; 272) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25GBR954; 273) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25EBK653; 274) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25DBB182; 275) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25ABS755; 276) DYNJ905066B, UDI-DI: 10195327098674(each), 40195327098675(case), Lot Number: 23BMF722; 277) DYNJ905066B, UDI-DI: 10195327098674(each), 40195327098675(case), Lot Number: 23BMB931; 278) DYNJ905269G, UDI-DI: 10193489337112(each), 40193489337113(case), Lot Number: 21BBD394; 279) DYNJ905269G, UDI-DI: 10193489337112(each), 40193489337113(case), Lot Number: 21ABH751; 280) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23FBE986; 281) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23EBQ671; 282) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23DBT141; 283) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 22GBL545; 284) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23FBP128; 285) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23DBM357; 286) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23BBB047; 287) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23BBA689; 288) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25IBI139; 289) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25HBO792; 290) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25GBB683; 291) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25EBA889; 292) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25BBO090; 293) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25ABT788; 294) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24LBN196; 295) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24KBJ970; 296) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24IBN246; 297) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24HBM625; 298) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24FBJ836; 299) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24DBT181; 300) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24CBS248; 301) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24BBN302; 302) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24ABD306; 303) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23LBJ184; 304) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23JBT789; 305) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23IBO509; 306) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23HBO123; 307) DYNJ906452F, UDI-DI: 10198459575303(each), 40198459575304(case), Lot Number: 25KBG657; 308) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23EBN137; 309) DYNJ907753F, UDI-DI: 10195327194482(each), 40195327194483(case), Lot Number: 22LBN185; 310) DYNJ907771C, UDI-DI: 10195327120399(each), 40195327120390(case), Lot Number: 22LBP170; 311) DYNJ908887B, UDI-DI: 10195327620158(each), 40195327620159(case), Lot Number: 24DBD796; 312) DYNJ908887B, UDI-DI: 10195327620158(each), 40195327620159(case), Lot Number: 24CBP744; 313) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24LBP187; 314) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24JBP902; 315) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24HBI648; 316) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24GBK492; 317) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24EBM187; 318) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23KME175; 319) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23JMB528; 320) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23HMI269; 321) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24HMA202; 322) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24DMI073; 323) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24CMI706; 324) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24BMA251; 325) DYNJ910536, UDI-DI: 10198459029554(each), 40198459029555(case), Lot Number: 24GBA620; 326) DYNJ910536A, UDI-DI: 10198459054358(each), 40198459054359(case), Lot Number: 24IBE833; 327) DYNJ910589, UDI-DI: 10198459040894(each), 40198459040895(case), Lot Number: 24HBF839; 328) DYNJ910594, UDI-DI: 10198459041327(each), 40198459041328(case), Lot Number: 24HBF837; 329) DYNJ911567, UDI-DI: 10198459409004(each), 40198459409005(case), Lot Number: 25HBB593; 330) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21DMF113; 331) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21CKA812; 332) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21CKA647; 333) DYNJ9701096I, UDI-DI: 10193489939651(each), 40193489939652(case), Lot Number: 22AMC654; 334) DYNJT3381, UDI-DI: 10198459325205(each), 40198459325206(case), Lot Number: 25EBK724
Devices named in this recall
84 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ38844NMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. DYNJ909398AMedline Industries, Inc. · Gynecological laparoscopic kit4 recalls0 reports
- Medline Industries, Inc. DYNJ66298DMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls1 report
- Medline Industries, Inc. CDS984860FMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ49191BMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ64266Medline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ901812JMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ906926BMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ901133AMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ57607AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ901559JMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS985423GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ905269GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ67269Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS983300JMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ63079BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ51795FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL92Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL168Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL91Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL94Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL100Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL109Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL119Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ58321AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ30005GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ47714AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ52569BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ50439FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ9701096GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ68710Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS984243GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ9701096IMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906322DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ49705KMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS984543DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ905066BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ63079CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS984543FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907771CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907753FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ47550CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0347372FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ69419AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906452CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL168AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL119AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL92AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL94AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0347372GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS984543GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ81492AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ32706BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ50348FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ84783Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0347372IMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906452DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ83475AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ900262AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ63079DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ910009Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ81761AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0281948NMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL116DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ908887BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS984543IMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ86195AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ69419DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ908887CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL184DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL92CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL94CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL116FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ910536Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ910589Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ910594Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ63079FMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ910536AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ28252GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYKM1657AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJT3381Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYKMBNDL94DMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS982893IMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ55862Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98329).
- Z-1384-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A
- Z-1385-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55A; 3) TRACH TOTE, Model Number: DYNJ85691
- Z-1386-2026 · Open, ClassifiedMedline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number: DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number: DYNJ56436B; 3) CPNB SETUP TRAY, Model Number: DYNJRA1979; 4) CPNB SET…
- Z-1387-2026 · Open, ClassifiedMedline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C
- Z-1388-2026 · Open, ClassifiedMedline Convenience Kits: 1) ENDO KIT W/SYRINGE, Model Number: DYKE1743; 2) BGMC ANGIOGRAPHY PACK-LF, Model Number: DYNJ0800934D; 3) GENERAL ENDO PACK-LF, Model Number: DYNJ08428…
- Z-1389-2026 · Open, ClassifiedMedline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Numbe…
- Z-1390-2026 · Open, ClassifiedMedline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP…
- Z-1391-2026 · Open, ClassifiedMedline Convenience Kits: 1) PICC LINE TRAY, Model Number: 00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number: ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Numbe…
- Z-1392-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:…
- Z-1393-2026 · Open, ClassifiedMedline Convenience Kits: 1) DENTAL PACK, Model Number: DYNDA3003
- Z-1394-2026 · Open, ClassifiedMedline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number:…
- Z-1395-2026 · Open, ClassifiedMedline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W;…
- Z-1396-2026 · Open, ClassifiedMedline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721
- Z-1397-2026 · Open, ClassifiedMedline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA, Model …
- Z-1398-2026 · Open, ClassifiedMedline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A;
- Z-1399-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER #5, Model Number: ACC010475; 4…
- Z-1400-2026 · Open, ClassifiedMedline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE MOMENI SB CAPSULOTOMY REC, Model …
- Z-1401-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 2) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: …
- Z-1402-2026 · Open, ClassifiedMedline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43
- Z-1403-2026 · Open, ClassifiedMedline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L
- Z-1404-2026 · Open, ClassifiedMedline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
- Z-1405-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A
- Z-1406-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAPE PACK-CHOICE, Model Number: DYNJ63118A
- Z-1407-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
- Z-1409-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
- Z-1410-2026 · Open, ClassifiedMedline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A
- Z-1411-2026 · Open, ClassifiedMedline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
- Z-1413-2026 · Open, ClassifiedMedline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYK…
- Z-1414-2026 · Open, ClassifiedMedline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL…
- Z-1416-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL1…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page